Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K053154

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-12-16

    (32 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Esaote's MyLab15/MyLab20 is a compact console ultrasound system intended to be used by a physician to perform general diagnostic ultrasound studies including Fetal, A bdominal, Pediatric, Small organ, Neonatal Cephalic, Cardiac, Transrectal, Transvaginal, Peripheral Vascular, Musculoskeletal (Conventional and Superficial).

    Device Description

    The MyLab15/20 is a compact console ultrasound system used to perform general The WYLao is a coulies. Its primary modes of operation are: B-Mode, M-Mode, PW Glaghoone and Color Flow Mapping and Tissue Enhancement Imaging (TEI). Dopper and Color 110 mages and time 2D images and 3D images (in manual mode) with all probes. The system, in combination with the probe BC432P, offers the possibility to also produce automatic 3D and real time 4D images

    AI/ML Overview

    The provided 510(k) summary for the MyLab15/20 Just3D/4D Ultrasound System focuses on demonstrating substantial equivalence to predicate devices, rather than establishing acceptance criteria and proving performance through a standalone study with specific metrics.

    Therefore, the document does not contain the acceptance criteria and the results of a study designed to prove the device meets those criteria.

    Here's why the requested information cannot be extracted from the provided text:

    • No Acceptance Criteria Defined: The document does not explicitly state any numerical or qualitative "acceptance criteria" against which the device's performance is measured. Instead, it relies on comparison to predicate devices to establish substantial equivalence.
    • No Performance Study Reported: There is no description of a dedicated study (clinical or analytical) designed to evaluate the device's performance against specific metrics. The tables provided are for "Comparison Chart for Substantial Equivalence" and "Indications for Use," which list features and approved applications, not performance results.
    • Focus on Substantial Equivalence: The entire submission is built around demonstrating that the new device is "substantially equivalent" to existing, legally marketed predicate devices. This typically involves showing that the new device has the same intended use, similar technological characteristics, and that any differences do not raise new questions of safety or effectiveness. It does not ordinarily require a separate, hypothesis-driven performance study with pre-defined acceptance criteria.

    Therefore, the bulk of the requested information (points 1-9) cannot be answered from the given text.

    I can provide the information that is present in the document which touches upon aspects of equivalence:


    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated. The implicit acceptance criterion is "substantial equivalence" to the listed predicate devices in terms of intended use and technological characteristics, as determined by the FDA.
    • Reported Device Performance: No specific numerical performance metrics (e.g., sensitivity, specificity, accuracy, resolution) are reported in the document. The performance is implied to be equivalent to the predicate devices.
    Feature / CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (MyLab15/20 Just3D/4D)
    General CharacteristicsSubstantially equivalent to predicate devices (K014168, K043588, K043455, K040060)
    Fetal ApplicationYesYes
    Abdominal ApplicationYesYes
    Pediatric ApplicationYesYes
    Small Organ ApplicationYesYes
    Neonatal Cephalic ApplicationYesYes
    Cardiac ApplicationYesYes
    Transrectal ApplicationYesYes
    Transvaginal ApplicationYesYes
    Peripheral Vascular ApplicationYesYes
    Musculo-skeletal ApplicationYesYes
    Transducer Type
    LinearYesYes
    ConvexYesYes
    Phased arrayNoNo (Unlike some predicates which have it)
    2D Freq MHzComparable range2.7 - 15 (Predicates range 1-20 MHz)
    MultifrequencyYesYes
    Special ProbesSimilar functionalityEndocavity, Mechanically Driven 3D Convex Array
    Biopsy attachments: ConvexYesYes
    Biopsy attachments: LinearYesYes
    Imaging Modes
    Real Time 2DYesYes
    M-modeYesYes
    PW DopplerYesYes
    CW DopplerNoNo (Unlike some predicates which have it)
    CFM DopplerYesYes
    Amplitude DopplerYesYes
    TriplexYesYes
    3D/4DYesYes (includes Just3D and Just4D)
    Monitor size (inches)Comparable15" CRT / 15" LCD
    ECGOptionalOptional
    Digital archival capabilitiesYesYes
    VCR & Video printersYesYes
    M&A capabilitiesYesYes
    Safety
    Electrical safetyEN60601-1EN60601-1

    Regarding the study aspects:

    1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): Not mentioned. This document does not describe a performance study with a test set.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not mentioned.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not mentioned.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not mentioned. This device is an ultrasound system with 3D/4D imaging, not specifically an AI-powered diagnostic tool in the sense of current AI submissions (which would typically involve MRMC studies).
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable, as this is an imaging system, not a standalone algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not mentioned.
    7. The sample size for the training set: Not mentioned.
    8. How the ground truth for the training set was established: Not mentioned.

    Additional relevant information from the document:

    • Device Name: MyLab15/20 - Just3D and Just 4D Ultrasound System
    • Manufacturer: Pie Medical (represented by Carri Graham, Official Correspondent)
    • Date of Submission: November 10, 2005 (Received: November 15, 2005)
    • FDA Clearance Date: December 16, 2005
    • Predicate Devices:
      • K014168 - Technos Esaote, S.p.A.
      • K043588 - MyLab15/20 Ultrasound System Pie Medical (This appears to be an earlier version of the same system, indicating an upgrade or new features)
      • K043455 - 8000Live Medison
      • K040060 - G50 Siemens
    • Intended Uses (MyLab15/MyLab20): Fetal, Abdominal, Pediatric, Small organ, Neonatal Cephalic, Cardiac, Transrectal, Transvaginal, Peripheral Vascular, Musculoskeletal (Conventional and Superficial).
    • New Indications (N): The submission introduces new indications for various transducers and applications. For example, the BC432P transducer supports new indications (N) for Fetal, Abdominal, and Pediatric applications across B, M, PW, Color Doppler, Amplitude Doppler, and combined modes, including N[2] for Tissue Enhancement Imaging (TEI), N[3] for 3D Imaging, and N[4] for 4D Imaging. This is how the 3D/4D capability is being introduced and cleared.

    Acoustic Output Post-Clearance: The FDA letter (page 5) requires a post-clearance special report detailing acoustic output measurements based on production units. This is a common requirement for ultrasonic systems to ensure safety, but it's not a pre-market performance study for diagnostic accuracy.

    In summary, the provided document is a 510(k) summary demonstrating substantial equivalence for an ultrasound system, not a report on a performance study against specific acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1