Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K052675

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-11-30

    (64 days)

    Product Code
    Regulation Number
    886.5916
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SynergEyes™ KC (paflufocon D hem-iberfilcon A) Hybrid Contact Lenses for keratoconus are indicated for use in the correction of eyes with refractive errors that include hyperopia and myopia that manifest irregular astigmatism, in aphakic and not aphakic, and otherwise non-diseased eyes. The lenses are indicated for daily wear for the correction of up to +20.00 and -20.00 D in eyes with irregular astigmatism up to 6.00 D. The lenses may be disinfected using only a chemical disinfecting system compatible with both hydrogel and rigid gas permeable lenses.

    Device Description

    The SynergEyes™ KC (paflufocon D hem-iberfilcon A) Hybrid Daily Wear Contact Lens is a corneal optic portion surrounded by a soft hydrophilic skirt that straddles the limbus of the eye:

    • in the power range of -20.00 to +20.00 diopters for sphere, -0.50 to -6.00D cylinder.
    • with center thickness from 0.18mm to 0.30mm.
    • with base curves of 5.50mm to 7.70mm.
    • with diameter of 14.50mm.
      The lens material (paflufocon D hem-iberfilcon A) is identical to the predicate current material. There are no differences to the chemical composition, formulation, or sterilization as described in the referenced 510(k).
      This lens material for the rigid portion is paflufocon D lathe cut, surrounded by soft hydrophilic Copolymer (hem-iberfilcon A), by means of e-beam sterilization. When placed on the eye, the Contact Lens acts as a refracting medium to focus light rays onto the retina. The device is available as a lathe cut contact lens in the following design for Keratoconus: In A and M Hybrid Contact Lens is material design and composition, and the visibility tinted material. This device is equivalent to the SynergEyes - A and M Hybrid Contact Lens for Keratoconus and manufactured by Ciba Vision Corporation.
      The SynergEyes™ KC Hybrid Daily Wear Contact Lens is a rigid gas permeable material of (paflulofon-D) rigid gas permeable polymer. The soft skirt is comprised of HEMA (hydroxyethylmethacrylate) of 27% water and 73% polymer.
      The junction between the rigid material is bound by a proprietary chemical bonding method.
    AI/ML Overview

    The provided text describes the SynergEyes™ KC Hybrid Contact Lens for Keratoconus and its clinical study. Here's a breakdown of the acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    The documents do not explicitly state pre-defined acceptance criteria with specific numerical targets for efficacy or safety endpoints (e.g., "visual acuity must improve by X lines for Y% of subjects"). Instead, the conclusion states that the device "provided satisfactory performance as expected" and "demonstrated safe and effective use of the device for its intended use" based on the observed outcomes.

    Therefore, the "acceptance criteria" can be inferred from the study's primary outcomes for safety and effectiveness, and the "reported device performance" are the results achieved in these areas.

    Acceptance Criteria (Inferred)Reported Device Performance
    Safety:
    - Adverse Events are within acceptable limits.Four (4) adverse events reported among 62 subjects (2 completed, 2 discontinued). (Details: mechanical abrasion, transient elevated IOP, edema/infiltrates, red eye/infiltrates). All resolved.
    - Loss of Visual Acuity (VA) is minimal/acceptable.5 completed eyes and 3 discontinued eyes had VA decreases of > 2 lines of Snellen VA (comparing contact lens VA to best corrected VA). This was considered "expected with this population."
    - Complications, symptoms, and complaints are acceptable.Slit lamp exams showed expected characteristics for keratoconus patients (high rates of staining, injection, edema, neovascularization). Most common complaints for completed eyes: discomfort/awareness (26.0%), dryness/scratchiness (12.5%), blurred vision (11.7%), itching/burning (10.6%). Higher rates for discontinued patients.
    Effectiveness:
    - Contact lens visual acuity is improved/satisfactory.Final visual acuity for completed subjects: 20/20 or better (26.3%), 20/25 or better (55.1%), 20/30 (71.4%), 20/40 or better (83.9%). Rates lower for discontinued subjects.
    - High percentage of subjects complete the study.44 of 62 subjects (71%) completed the study. Discontinuation rate was 29% (18 subjects). The document states higher than estimated discontinuation rate was "anticipated due to the nature of the subject population."
    - Acceptable lens wearing time.Average daily wearing time for completed patients was 11.4 hours per day.
    - Overall performance is satisfactory for intended use.Conclusion: "The SynergEyes™ KC Hybrid Contact Lens for Keratoconus provided satisfactory performance as expected."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: A total of 62 subjects were dispensed lenses and evaluated. 44 subjects (71%) completed the study, and 18 subjects (29%) were discontinued.
    • Data Provenance: The study was a "prospective, multi-site, open label study" conducted in the United States, enrolling subjects at 12 sites. The data is prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of external experts to establish "ground truth" in the way it might for an AI diagnostic device. For this contact lens study, the assessment of safety and effectiveness relied on:

    • Visual acuity measurements.
    • Slit lamp (biomicroscopic) examinations performed by the clinical investigators at each site.
    • Patient-reported symptoms, problems, and complaints.
    • Adverse event reporting.

    The qualifications of the clinicians involved (e.g., ophthalmologists, optometrists) are not explicitly stated, beyond the fact that the contact person is an Optometrist (Richard Lippman, OD FAAO VP Ophthalmic Regulatory Products).

    4. Adjudication Method for the Test Set

    There is no mention of an adjudication method (like 2+1 or 3+1 consensus) for the clinical data. Clinical assessments (visual acuity, slit lamp, etc.) and adverse events were likely recorded by the site investigators and then aggregated and analyzed by the study sponsor.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This was not an MRMC comparative effectiveness study involving AI. It was a clinical study evaluating the safety and effectiveness of a contact lens in human subjects.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    This was a clinical study of a medical device (contact lens) and not an AI algorithm. Therefore, no standalone algorithm performance study was done.

    7. The Type of Ground Truth Used

    The "ground truth" in this context refers to the direct clinical observations and measurements of patient health, visual performance, and comfort. This includes:

    • Clinical Measurements: Visual acuity, slit lamp examination findings (staining, injection, edema, neovascularization).
    • Patient-Reported Outcomes: Discomfort/awareness, dryness/scratchiness, blurred vision, itching/burning, and reasons for discontinuation (e.g., poor comfort, poor vision, loss of interest).
    • Adverse Event Reporting: Documented adverse events and their resolution.

    8. The Sample Size for the Training Set

    There is no mention of a "training set" as this was a clinical study for a physical medical device, not an AI model requiring machine learning training data.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as this was not an AI/machine learning study.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1