Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K051935

    Validate with FDA (Live)

    Date Cleared
    2005-08-02

    (18 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    powder free vinyl patient examination gloves, white(non-colored) that meets all of the requirements of ASTM standard D 5250-00-4

    AI/ML Overview

    The provided K051935 Premarket Notification [510(k)] Summary pertains to "Powder Free Vinyl Patient Examination Gloves, White (non-colored)". This device is a medical glove, and the acceptance criteria and study detailed are for demonstrating its compliance with relevant standards and safety.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
    DimensionASTM standard D 5250-00c4Meets
    Physical PropertiesASTM standard D 5250-00c4Meets
    Freedom from pinholes21 CFR 800.20Meets
    Powder ResidualASTM standard D 5250-00c4 and D6124-01<2mg/glove
    BiocompatibilityPrimary Skin Irritation in rabbitsPasses (Not a Primary Skin Irritation)
    Dermal sensitization in the guinea pigPasses (Not a Dermal Sensitization)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the specific sample sizes used for each test (Dimension, Physical Properties, Freedom from pinholes, Powder Residual, Biocompatibility). It refers to meeting the requirements of the standards (ASTM D 5250-00c4, 21 CFR 800.20, D6124-01, and ISO10993-10). These standards would define the required sample sizes for such testing.

    The data provenance is not explicitly stated as retrospective or prospective, nor does it mention a specific country of origin for the data (beyond the submitter's address being in China). However, generally, performance testing for medical devices like gloves is conducted prospectively on manufactured samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable in the context of this device and its testing. Performance criteria for medical gloves are based on established engineering and safety standards (e.g., ASTM, FDA regulations). There is no "ground truth" established by human experts in the same way it would be for an AI diagnostic device. The "ground truth" is defined by the objective measurements and pass/fail criteria of the specified standards.

    4. Adjudication Method for the Test Set:

    This information is not applicable. As mentioned above, the tests are objective measurements against established standards, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. This type of study is specifically relevant for diagnostic devices where human readers interpret data (e.g., medical images) with and without AI assistance to assess the AI's impact on human performance. For a patient examination glove, such a study is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    No, a standalone study was not done. This concept is also primarily applicable to diagnostic algorithms. The "device" in this case is a physical product (gloves) whose performance is assessed through laboratory testing against physical, chemical, and biological standards, not through an algorithm.

    7. The Type of Ground Truth Used:

    The "ground truth" used for this device is based on established objective criteria defined within recognized standards and regulations. These include:

    • ASTM standard D 5250-00c4: Specifies requirements for dimensions, physical properties, and powder residue.
    • 21 CFR 800.20: Relates to freedom from pinholes (waterleak test AQL).
    • ASTM D6124-01: Also concerns powder residue.
    • ISO10993-10: Addresses biocompatibility (primary skin irritation and dermal sensitization).

    Essentially, the ground truth is the compliance with these pre-defined, measurable specifications.

    8. The Sample Size for the Training Set:

    This device does not have a "training set" in the typical sense of machine learning or AI. Performance testing is done on manufactured samples according to the specifications of the referenced standards. There is no algorithm being "trained."

    9. How the Ground Truth for the Training Set Was Established:

    As there is no training set, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1