Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K051089

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-10-26

    (181 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GII Spinal Fixation System When used as a pedicle screw fixation system in the non-cervical posterior spine in skeletally mature patients, the GII@ Spinal System is indicated for one or more of the following: (1) degenerative spondylolisthesis with objective evidence of neurologic impairment, (2) fracture, (3) dislocation, (4) scoliosis, (5) kyphosis, (6) spinal tumor, and/or (7) failed previous fusion (pseudarthrosis).

    In addition, when used as a pedicle screw fixation system, the GII@ Spinal System is indicated for skeletally mature patients: (1) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar/first sacral (L5-S 1) joint; (2) who are receiving fusions using autogenous bone graft only; (3) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (4) who are having the device removed after the development of a solid fusion mass.

    The GII system is also a hook and sacral/iliac screw fixation system of the non-cervical spine indicated for (1) degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) degenerative spondylolisthesis with objective evidence of neurologic impairment, (3) fracture, (4) dislocation, (5) scoliosis, (7) spinal turnor, and/or (8) failed previous fusion (pseudarthrosis).

    Device Description

    The new plates are offered in a series of sizes designated by the number of slots or holes within the plates. The I ho w plates are single slot. The new plates are provided precurved to more closely match the natural lordotic curve of the lumbar spine. Two different curves are offered in each size.

    The devices are produced of a polyner matrix of Polyetherketonekctone (PEKEKK) Resins and long fiber carbon filaments.

    The new plates are intended to be used with the existing pedicle screws, nuts, washers, cross-links and instrumentation as currently provided with the cleared GII Spinal Fixation System. Because the new material is not a metallic material, precautions against mixing dissimilar metals or of a galvanic/corrosive interaction with the existing implants is eliminated.

    The previously cleared indications for use are unchanged.

    AI/ML Overview

    This document is a 510(k) premarket notification for the GII OstaPek® Plate from COLIGNE AG. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain a detailed study report with acceptance criteria and reported device performance in the way typically expected for an AI/ML device.

    This submission is for a medical implant (spinal fixation system), not a software/AI device that would involve performance metrics like sensitivity, specificity, or accuracy based on human interpretation or algorithm output. The "acceptance criteria" here refer to compliance with material, mechanical, and biocompatibility standards, which are physical tests rather than clinical study results comparing algorithm performance.

    Therefore, many of the requested points are not applicable or cannot be extracted from this type of regulatory document. I will answer the applicable points based on the information provided.


    1. Table of acceptance criteria and the reported device performance

    For this medical implant, acceptance criteria and performance are related to physical and biological properties.

    Acceptance Criteria CategorySpecific Criteria (Implicit from text)Reported Device Performance (Summary from text)
    Material StandardsCompliance with material standardsDevices produced of Polyetherketonekctone (PEKEKK) Resins and long fiber carbon filaments. Eliminates concerns about mixing dissimilar metals or galvanic/corrosive interaction.
    Mechanical TestingCompliance with mechanical testing standardsFatigue and static testing complete. Test results demonstrate new plates perform equivalently or superiorly to previously cleared GII Spinal Fixation System plates.
    BiocompatibilityCompliance with biocompatibility standardsBiocompatibility testing complete.
    LabelingSpecific warnings and precautionsLabeling contains required warnings and precautions.

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: Not explicitly stated for mechanical and biocompatibility testing. These tests typically involve a specific number of samples per test type according to the relevant standards (e.g., n=5 or n=10 per test for fatigue).
    • Data Provenance (Country of Origin, Retrospective/Prospective): Not applicable in the context of device performance testing. The company is based in Zurich, Switzerland, and the submission is to the US FDA. The testing would have been conducted in a laboratory setting, not on patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts & Qualifications: Not applicable. Ground truth for mechanical and biocompatibility testing is established through standardized laboratory methods and measurements, not by human expert consensus or clinical interpretation.

    4. Adjudication method for the test set

    • Adjudication Method: Not applicable. Standardized lab tests have defined pass/fail criteria.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    • MRMC Study: No. This is not a diagnostic device or an AI/ML software; it is a physical implant. Therefore, an MRMC study is not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used

    • Type of Ground Truth: For this device, ground truth is established through:
      • Validated Physical Measurements: Mechanical properties (e.g., strength, fatigue life) are measured using calibrated testing equipment against established engineering standards (e.g., ASTM standards).
      • Biocompatibility Testing: Assessed through in-vitro and in-vivo biological response tests following ISO standards.
      • Material Characterization: Chemical composition and properties verified against material specifications.

    8. The sample size for the training set

    • Training Set Sample Size: Not applicable. This device is not an AI/ML model that requires a training set. The "development" would involve engineering design and prototyping, not data training.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set: Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1