510(k) Data Aggregation

    K Number
    K050828

    Validate with FDA (Live)

    Date Cleared
    2006-02-28

    (333 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TROJAN SUPRA® Lubricate Polyurethane Male Condom is used for contraception and for prophylactic purposes (to help reduce the risk of pregnancy and the transmission of sexually transmitted diseases, STDs).

    Device Description

    The TROJAN SUPRA® Lubricated Polyurethane Condom is a male condom consisting of a sheath of polyurethane with a lubricant coating. The condom is a straight-walled, nipple-end condom with a nominal length of 190 mm and an approximate flatwidth of 58 mm.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the TROJAN SUPRA® Lubricated Polyurethane Male Condom. The acceptance criteria and the study that proves the device meets those criteria are detailed below.

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance CriteriaReported Device Performance (TROJAN SUPRA® Polyurethane Condom)Reference (Control Latex Condom)
    Clinical Breakage RateNot significantly different from a standard latex condom control.0.6%1.3%
    Clinically Significant Slippage RateNot significantly different from a standard latex condom control.1.1%0.5%

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Users/Cases): 206 couples.
    • Data Provenance: Prospective, conducted in France.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not explicitly provided in the document. The study involved "an In-Use Study" following an FDA guidance, and the outcomes were "clinical breakage rate" and "clinically significant slippage." It's customary for such clinical studies to involve medical professionals in assessing and recording these events, but the specific number and qualifications of experts are not detailed.

    4. Adjudication Method for the Test Set

    This information is not explicitly provided. Clinical studies often have adjudication committees for adverse events or endpoint assessment, but the document does not specify if one was used or the method (e.g., 2+1, 3+1).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Its Effect Size

    No, an MRMC comparative effectiveness study was not conducted as this is a medical device (condom) and not an imaging-based diagnostic or AI-assisted diagnostic tool. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, a standalone algorithm performance study was not done. The device is a physical medical device (condom) and does not involve AI algorithms.

    7. The Type of Ground Truth Used

    The ground truth used was clinical observation and user reporting of objective outcomes, specifically:

    • Clinical Breakage Rate: Direct observation/reporting of condom breakage during intercourse.
    • Clinically Significant Slippage Rate: Direct observation/reporting of condom slippage during intercourse.

    8. The Sample Size for the Training Set

    No training set information is applicable or provided as this is a physical medical device, not an AI-based system.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable as there is no training set for this type of medical device.