Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K050505

    Validate with FDA (Live)

    Device Name
    DEPOTONE
    Date Cleared
    2005-04-05

    (36 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    DepotOne is a sterile single use hypodermic single lumen needle intended for the injection and removal of fluids from below the surface of the skin in humans.
    DepotOne is a sterile hypodermic single lumen needle intended for the injection and removal of fluids from below the surface of the skin in humans. The needle is intended to interface with the male luer connector of a syringe or intravascular administration set. The needle is intended for single use only.

    Device Description

    The DepotOne needles comprise of a length of Hypodermic grade stainless steel tube connected to a female connector (hub) that is designed to mate with a male connector of a piston syringe or an intravascular administration set. The other end of the tube is sharpened with a lancet tip in accordance with ISO7864: 1993. The sharpened section of tube is of a reduced diameter. The fluid is injected or removed through a side aperture adjacent to the sharpened tip.

    AI/ML Overview

    The acceptance criteria and study proving the DepotOne™ device meets these criteria are detailed in the provided 510(k) submission.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device performance is demonstrated through adherence to various ISO and EN standards. While specific numerical targets for "acceptance criteria" are not always explicitly stated in the document for each test (e.g., "passed"), compliance with these standards is the acceptance criterion.

    Acceptance Criteria (Standard / Test)Reported Device Performance
    Material Composition
    Cannulae: Stainless SteelStainless Steel
    Hub (6% Locking Luer): Polypropylene (Predicate)Polycarbonate (DepotOne™) - Note: Difference in material stated not to impact sterilization and chosen for superior bond strength.
    Needle Lubricant: SiliconeSilicone
    Adhesive: EpoxyEpoxy
    Sterilisation Method: ETOETO
    Needle Shield: PolypropylenePolypropylene
    Physical and Functional Performance (per relevant ISO standards)
    ISO 7864:1993 Sterile Hypodermic Needles for Single Use complianceDepotOne™ conforms to the requirements.
    - Needle/Hub Bond TestPassed to EN ISO 7864:1993.
    - Patency of Lumen TestPassed to EN ISO 7864:1993.
    - Extractables TestPassed to EN ISO 7864:1993.
    - Acidity and Alkalinity TestPassed to EN ISO 7864:1993.
    EN ISO 9626:1995 Stainless Steel Tubing for Medical Devices
    - Bend Test (for tubing)Passed to EN ISO 9626:1995. Note: For 1.2mm (18G) XTW and 0.8mm (21G) XTW, passed requirements for nearest available data (1.1mm 19G XTW and 0.7mm 22G XTW) as no specific max deflection data in standard.
    - Breakage Test (for tubing)Passed to EN ISO 9626:1995.
    Additional Performance Tests
    Fluid Flow TestsPassed.
    Breakage Tests for completed needlesPassed.
    Bend Tests for complete needles (including swaged tip)Passed.
    Penetration force testsPassed.
    Sterilization & Biocompatibility
    EN550 (ETO sterilization validation)Needles EO sterilized by a validated method in accordance with EN550.
    EN556 (sterility certification)Sterilization certified to comply with EN556 and validated by an independent laboratory to a Sterility Assurance Level (SAL) of 10-6.
    Pyrogenicity (LAL test, European Pharmacopoeia)Tested for freedom of Pyrogens by LAL (limulus amebocyte lysate) test conducted by an independent laboratory in accordance with European Pharmacopoeia. Performed on samples from every production batch/lot.
    Residual Gas Contents (EN ISO10993-7)Tested in accordance with EN ISO10993-7. Limiting viewe 20mg ethylene oxide/24hrs.
    Luer Connector Compatibility
    ISO 594-2 (6% Luer connector)Proximal end securely attached to a 6% Luer connector, in accordance with ISO 594-2. (Note: Predicate device K-Pack II did not completely meet this requirement, but DepotOne™ does.)
    ISO 6009 (color coding of connector)Connector is color coded in accordance with ISO 6009.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the numerical sample sizes for each test mentioned (e.g., how many needles were subjected to bend, breakage, or bond tests). It only states that "DepotOne™M needles have passed the following tests" or "DepotOne™ needles have been subjected to and passed the following tests."

    The data provenance is from prospective testing conducted by Imprint Pharmaceuticals Ltd. and independent laboratories on the DepotOne™ device itself. The country of origin for the testing facilities is not explicitly stated, but Imprint Pharmaceuticals is based in Surrey, UK.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable to this type of device submission. The "ground truth" for hypodermic needles is established by compliance with international engineering and sterilization standards (e.g., ISO, EN, Pharmacopoeia) and validated laboratory testing, not by expert consensus on clinical interpretation or outcomes in the same way as, for example, an imaging diagnostic device. The expertise lies in engineering, materials science, microbiology, and quality assurance.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods are typically relevant for studies involving human interpretation (e.g., reading images), which is not the case for the physical and biological testing of a hypodermic needle. The "adjudication" is inherent in whether the device objectively passes or fails the specified criteria of the international standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This information is not applicable, as the device is a hypodermic needle and not an AI-enabled diagnostic or interpretive tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable, as the device is a hypodermic needle and does not involve AI algorithms.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device is based on International Standards and validated laboratory test results. This includes:

    • Engineering specifications for materials, dimensions, and physical properties (e.g., bend, breakage, bond strength, fluid flow, penetration force).
    • Biological safety and sterility standards (e.g., SAL of 10-6, pyrogenicity, residual gas content).
      These standards and test methods serve as the objective "truth" against which the device's performance is measured.

    8. The Sample Size for the Training Set

    This information is not applicable, as the device is a physical hypodermic needle and does not involve an AI algorithm or a "training set" in the context of machine learning.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no "training set" for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1