Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K043564

    Validate with FDA (Live)

    Date Cleared
    2005-01-27

    (31 days)

    Product Code
    Regulation Number
    870.1425
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Horizon Angio is a state-of-the-art computerized laboratory, capable of acquiring and displaying essential patient data such as ECG/Heart Rate, invasive blood pressure, pulse oximetry, respiration, cardiac output and body temperature.

    Heart rate, multi-lead ECG and BP waveforms from different heart sites are continuously presented on the Physiological Waveform Display. The hemodynamic data, waveform and numerical, can be stored, recorded, analyzed and presented in a variety of report formats.

    Device Description

    The Horizon Angio is a state-of-the-art computerized laboratory, capable of acquiring and displaying essential patient data such as ECG/Heart Rate, invasive blood pressure, pulse oximetry, respiration, cardiac output and body temperature. Heart rate, multi-lead ECG and BP waveforms from different heart sites are continuously presented on the Physiological Waveform Display. The hemodynamic data, waveform and numerical, can be stored, analyzed and presented in a variety of report formats.

    The Horizon Angio software includes graphic presentation of the abdominal, cranial and peripheral vascular system, to support reporting of vascular catheterization.

    The Sun Blade 150 workstation is a uni-processor system that runs the Cathlab program on a UNIX operating system. The workstation receives the digitized signals from the CFE via the Ethernet hub, displays real-time vital signs, analyzes, processes, and calculates the vital sign data and waveforms, cardiac status in real time during the catheterization process, creating a fully documented case history. The workstation continuously displays the vital signs waveforms and data on the local LCD displays.

    The hemodynamic data, waveform and numerical, can be stored, analyzed and presented in a variety of report formats.

    A Laser Printer is provided in the system. This provides printouts of textual and graphical summaries of all patient data and catheterization procedures.

    Base Configuration: Cathlab parameters

    • . 2 Invasive Blood Pressure channels
    • . Diagnostic 7 or 12 Lead ECG
    • . Non-invasive Blood Pressure
    • Pulse Oximetry (SpO2) .

    Horizon Angio Options:

    • Full Disclosure .
    • Off-line workstations ●
    • Remote Interactive terminal .
    • . Angiography Analysis Package
    • CDR, DVD or Optomagnetic drive .
    • Choice of Console Table regular, enhanced, compact or without consol .
    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study proving device efficacy:

    Device: Horizon Angio Cathlab (a sub-system of the Horizon 9000WS Cathlab)

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document provides extensive performance specifications for the Horizon Angio, which serve as the acceptance criteria. The accompanying information (e.g., "Meets or exceeds ANSI standard") indicates that the device's performance is intended to meet these criteria.

    ParameterAcceptance Criteria (from text)Reported Device Performance (Implied as meeting criteria)
    Chassis Leakage CurrentAll patient signal inputs fully isolated (<50 μA). Meets or exceeds ANSI standard: "Safe Current Limits for Electromedical Aparatus," (SCLE) Dec, 1978 item 2.1.1.The device is stated to meet or exceed these standards.
    ECG Frequency ResponseMonitor Mode: 0.5 to 40 Hz; Diagnostic: 0.05 to 150 Hz; Exercise: 1 to 25 Hz, -3 dBPerformance is stated to be exactly these specifications.
    ECG Input ImpedanceTypical 20 MΩ; Minimum greater than: 5 MΩ differential, DC to 10 Hz; 2.5 MΩ differential 10 to 100 Hz. 3 MΩ differential at 10 HzPerformance is stated to be exactly these specifications.
    ECG Common Mode RejectionAt least 100 dB at 50/60 Hz Without lead misbalance; 86 db with lead misbalance. In accordance with ANSI/AAMI EC11(9) Para. 3.2.11.Performance is stated to be exactly these specifications.
    ECG Input Dynamic Range±5mV p-p at a rate up to 320mV/sec, as per ANSI/AAMI EC13(8) Para. 3.2.9.1.Performance is stated to be exactly these specifications.
    ECG Input Offset± 300mV, as per ANSI/AAMI EC13 Para. 3.2.9.1.Performance is stated to be exactly these specifications.
    ECG NoiseLess than 30 μV p-p referenced to inputPerformance is stated to be exactly these specifications.
    Pacemaker Pulse RejectionReject pulses from: 2.0 mV to 700 mV pulses of 0.2 to 2.0 mSec pulse widths and ≥3.0mV for 0.1mSec pulse widthPerformance is stated to be exactly these specifications.
    Defibrillator ProtectionUp to 5 KV. Amplifier Recovery time: < 3 secondsPerformance is stated to be exactly these specifications.
    QRS Detection0.25 to 5.0 mV, 70-120 msec widthPerformance is stated to be exactly these specifications.
    Heart Rate Range20 to 350 bpmPerformance is stated to be exactly these specifications.
    Heart Rate AccuracyWithin 2 bpmPerformance is stated to be exactly these specifications.
    Heart Rate Response TimeLess than 7 sec for step change of 60 bpm from a base of 60 bpmPerformance is stated to be exactly these specifications.
    Blood Pressure Input Sensitivity5 μvolts/volt/mmHgPerformance is stated to be exactly these specifications.
    Blood Pressure Ranges-50 to +300 mmHgPerformance is stated to be exactly these specifications.
    Blood Pressure Zero Accuracy± 0.2 mmHgPerformance is stated to be exactly these specifications.
    Blood Pressure Zero DriftLess than +/- 0.2 mmHg in 24 hoursPerformance is stated to be exactly these specifications.
    Blood Pressure LinearityBetter than 1% of full scalePerformance is stated to be exactly these specifications.
    Blood Pressure Common Mode Rejection80 dB minimum (reference to chassis 50/60Hz)Performance is stated to be exactly these specifications.
    Blood Pressure Frequency ResponseDC to 12 Hz (DC to 40 Hz optional)Performance is stated to be exactly these specifications.
    Temperature Range27 °C to 45°C --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Performance is stated to be exactly these specifications. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Temperature Accuracy± 0.2°C --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Performance is stated to be exactly these specifications. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Respiration Frequency Response0.13 to 2.5 Hz., 3 dB bandwidth ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Performance is stated to be exactly these specifications. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Respiration Range8 to 150 bpmPerformance is stated to be exactly these specifications.
    SpO2 Range0% to 100% --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Performance is stated to be exactly these specifications. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Pulse Rate Range (SpO2)20-250 bpm, below 20 displays zero --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Performance is stated to be exactly these specifications. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Pulse Rate Accuracy (SpO2)± 3 bpm ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Performance is stated to be exactly these specifications. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    SpO2 AccuracyDetermined by specific sensor: Adult: ±2 digits between 70% and 100%, ±3 digits between 50% and 70%. Neonatal: ±3 digits between 70% and 95% --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Performance is stated to be exactly these specifications. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Non-Invasive Blood Pressure (NIBP) Pressure AccuracyOverall ± 3 mmHg, full scale ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Performance is stated to be exactly these specifications. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Defibrillation Pulse Protection5KV as per ANSI/AAMI EC13 (9), clause 3.2.2.2 and per IEC 60601-2-27 (12), clauses 17,101 and 102 --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------The device is stated to meet these standards. --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Degree of protection against electrical shockType CF and BF (ECG, IBP and CO = CF; NIBP and SpO2 = BF) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------The device is stated to meet these classifications. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
    Electrosurgical Interference SuppressionYes ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------The device is stated to have this feature. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

    2. Sample Size Used for the Test Set and Data Provenance:

    The document lacks explicit details regarding the sample size used for the test set in performance testing. It states: "Final testing for the whole Horizon Angio system included various performance tests designed to ensure that the device meets all functional requirements and performance specifications." This implies internal testing rather than a clinical study with a specific patient test set.

    The data provenance is not mentioned. Given the nature of a 510(k) submission for a device modification, the testing is likely retrospective on simulated or internal test data, not prospective clinical data from specific countries of origin.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    The document does not mention the use of experts to establish a "ground truth" for a test set in the context of clinical performance. The testing described focuses on engineering and performance specifications against industry standards.

    4. Adjudication Method for the Test Set:

    No adjudication method is described, as the testing focuses on meeting technical specifications rather than interpreting results that would require expert consensus.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC study is mentioned. This device is a physiological data acquisition and display system, not an AI diagnostic tool that assists human readers with interpretation. Therefore, a study comparing human reader improvement with or without AI assistance is not applicable to the information provided.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    The "Final testing for the whole Horizon Angio system included various performance tests designed to ensure that the device meets all functional requirements and performance specifications" implies standalone testing against its own specifications and relevant standards. The device is a "programmable diagnostic computer" that acquires and displays data, not an algorithm providing a diagnosis that would be independently evaluated for "algorithm-only" performance in a clinical context. Its performance is measured by the accuracy and reliability of its physiological measurements and display capabilities.

    7. Type of Ground Truth Used:

    The "ground truth" for the performance specifications are the defined electrical, physiological, and safety standards (e.g., ANSI/AAMI EC11, IEC 60601 series). For example, the accuracy of heart rate is measured against a known, controlled input signal, not against a clinical expert's interpretation or pathology.

    8. Sample Size for the Training Set:

    The concept of a "training set" is not applicable here as the Horizon Angio is not explicitly described as an AI/ML-driven device that learns from data. It's a system designed to acquire and display real-time physiological data based on pre-defined algorithms and hardware.

    9. How the Ground Truth for the Training Set Was Established:

    As there's no mention of a training set or AI/ML components, the establishment of ground truth for a training set is not pertinent to this document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1