510(k) Data Aggregation

    K Number
    K043242

    Validate with FDA (Live)

    Date Cleared
    2005-02-28

    (97 days)

    Product Code
    Regulation Number
    862.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Triage® TOX Drug Screen is a fluorescence immunoassay intended to be used with the Triage MeterPlus for the point-of-care qualitative determination of the presence of major metabolites above the threshold concentrations of up to 8 distinct drug classes, including assays for methamphetamines, acetaminophen/paracetamol, barbiturates, benzodiazepines, cocaine, opiates, phencyclidine, THC and tricyclic antidepressants in urine. The acetaminophen/paracetamol assay will yield positive results when acetaminophen/paracetamol is ingested at or above therapeutic doses.

    This test provides only a preliminary test result. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectroscopy (GC/MS) is the common confirmatory method.

    A quantitative serum acetaminophen/paracetamol measurement is the common confirmatory method for preliminary positive acetaminophen/paracetamol results.

    Device Description

    The Triage® TOX Drug Screen is a fluorescence immunoassay intended to be used with the Triage MeterPlus for the point-of-care qualitative determination of the presence of major metabolites above the threshold concentrations of up to 8 distinct drug classes, including assays for methamphetamines, acetaminophen/paracetamol, barbiturates, benzodiazepines, cocaine, opiates, phencyclidine, THC and tricyclic antidepressants in urine. The acetaminophen/paracetamol assay will yield positive results when acetaminophen/paracetamol is ingested at or above therapeutic doses.

    AI/ML Overview

    Acceptance Criteria and Study for Triage® TOX Drug Screen

    This response details the acceptance criteria and the study proving the Triage® TOX Drug Screen meets these criteria, based on the provided 510(k) summary.


    1. Acceptance Criteria and Reported Device Performance

    Drug ClassAnalyteThreshold Concentration (Acceptance Criteria)Reported Device Performance (Overall Agreement)
    Acetaminophen/ParacetamolAPAP5 µg/mL97.1% (with GC/MS)
    AmphetaminesAMP1000 ng/mLNot explicitly stated in the provided text
    MethamphetaminesmAMP1000 ng/mLNot explicitly stated in the provided text
    BarbituratesBAR300 ng/mLNot explicitly stated in the provided text
    BenzodiazepinesBZO300 ng/mLNot explicitly stated in the provided text
    CocaineCOC300 ng/mLNot explicitly stated in the provided text
    OpiatesOPI300 ng/mLNot explicitly stated in the provided text
    PhencyclidinePCP25 ng/mLNot explicitly stated in the provided text
    THCTHC50 ng/mLNot explicitly stated in the provided text
    Tricyclic AntidepressantsTCA1000 ng/mLNot explicitly stated in the provided text

    Note: The provided 510(k) summary only explicitly reports the agreement for Acetaminophen/Paracetamol. For other drug classes, the summary states that "The analytical performance characteristics of the assay were equivalent to predicate methods," but does not provide specific agreement percentages or studies for each individual drug class in the provided text. The table above reflects the information available in the given document. The acceptance criterion for each drug class is its specified threshold concentration.


    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:
      • For Acetaminophen/Paracetamol: 102 specimens were used for the method comparison study.
      • Specifically, 20 of these specimens were within ±25% of the cutoff concentration.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective). The summary only mentions "specimens" without further details on their source.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • The provided 510(k) summary does not mention the use of experts to establish ground truth for the test set.

    4. Adjudication Method for the Test Set

    • The provided 510(k) summary does not mention an adjudication method. For the Acetaminophen/Paracetamol study, the comparison was made directly against a reference method (GC/MS).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This device is an automated fluorescence immunoassay (point-of-care qualitative determination) and does not involve human readers interpreting images or data where AI assistance would typically be evaluated for reader improvement.

    6. Standalone Performance Study

    • Yes, a standalone performance study was done for the Acetaminophen/Paracetamol assay. The summary describes a "method comparison of acetaminophen results with GC/MS" and reports an "overall agreement of 97.1%." This is the algorithm's (device's) performance without human-in-the-loop. The summary also states, "The analytical performance characteristics of the assay were equivalent to predicate methods," implying similar standalone performance evaluations for other drug classes, although specific data is not provided in this extract.

    7. Type of Ground Truth Used

    • For the Acetaminophen/Paracetamol assay, the ground truth was established using an alternate chemical method: Gas Chromatography/Mass Spectroscopy (GC/MS). For acetaminophen/paracetamol specifically, "A quantitative serum acetaminophen/paracetamol measurement is the common confirmatory method." The summary implies GC/MS as the common confirmatory method for other drug classes generally.

    8. Sample Size for the Training Set

    • The provided 510(k) summary does not mention a training set or its sample size. This type of device (fluorescence immunoassay) typically relies on chemical and biological parameters rather than machine learning models that require training sets in the same way.

    9. How Ground Truth for the Training Set Was Established

    • As no training set is mentioned for this device, information on how its ground truth was established is not applicable based on the provided document.