    Intended Use

    The DakoCytomation ER/PR pharmDx™ assay is an immunohistochemical (IHC) kit system to identify human estrogen receptor (ER) and human progesterone receptor (PR) expression in breast cancer tissues routinely processed and paraffin-embedded for histological evaluation. ER/PR pharmDx specifically detects the ER alpha protein as well as the PR located in the cell nucleus of ER and/or PR-expressing cells.

    ER/PR pharmDx is indicated as an aid in identifying patients eligible for treatment with antihormonal or aromatase inhibitor therapies, as well as an aid in the prognosis and management of breast cancer.

    The DakoCytomation Monoclonal Mouse Anti-Human Progesterone Receptor, clone PgR 1294, is intended for laboratory use as a semi-quantitative detection of progesterone receptor by light microscopy in routinely processed normal and pathological human paraffinembedded tissue. This antibody is indicated for use as an aid in the management, prognosis and prediction of outcome of breast cancer. The clinical interpretation of any positive staining or its absence should be complemented by morphological and histological studies with proper controls. Evaluations should be made within the context of the patient's clinical history and other diagnostic tests by a qualified individual

    Device Description

    The DakoCytomation ER/PR pharmDx™ assay is a semi-quantitative immunohistochemical (IHC) kit system to identify estrogen receptor (ER) and progesterone receptor (PR) expression in normal and neoplastic tissues routinely processed and paraffin-embedded for histological expressing cells respectively.

    ER/PR pharmDx assay is available in two configurations, both manual and automated and is optimized for use with DakoCytomation detection systems.

    The DakoCytomation Monoclonal Mouse Anti-Human Progesterone Receptor, clone PgR 1294, is a component of the ER/PR pharmDx kit, which will also be commercially available as a concentrate.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary, specifically focusing on the acceptance criteria and study proving device performance for the DakoCytomation ER/PR pharmDx™ Kit and the Monoclonal Mouse Anti-Human Progesterone Receptor, clone PgR 1294:

    Acceptance Criteria and Device Performance

    The document does not explicitly present a table of numerical acceptance criteria with corresponding reported device performance values. Instead, it describes performance characteristics that were evaluated and states that the results demonstrated "a substantial degree of equivalency to the predicate devices" and "substantially equivalent staining results" to a reference method.

    The primary acceptance criteria appear to be substantial equivalence to predicate devices and high concordance with a recognized reference method.

    Table of Acceptance Criteria and Reported Device Performance (Inferred)

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance
    ConcordanceSubstantial equivalency (e.g., >95% or similar) with predicate devices and reference methods.99% concordance with the reference Allred method.
    SpecificityDemonstrates appropriate specificity compared to predicate devices.Demonstrated specificity; results deemed substantially equivalent to predicate devices.
    SensitivityDemonstrates appropriate sensitivity compared to predicate devices.Demonstrated sensitivity; results deemed substantially equivalent to predicate devices.
    ReproducibilityDemonstrates appropriate reproducibility compared to predicate devices.Demonstrated reproducibility; results deemed substantially equivalent to predicate devices.

    Study Description and Details:

    The summary refers to a single study that encompassed evaluations of specificity, sensitivity, reproducibility, and concordance testing.

    1. Sample Size and Data Provenance:

      • Test Set Sample Size: Not explicitly stated for any of the performance characteristics.
      • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective).
    2. Number of Experts and Qualifications for Ground Truth:

      • The document does not provide information on the number of experts used to establish ground truth or their specific qualifications.
    3. Adjudication Method:

      • The document does not specify any adjudication method used for the test set.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No MRMC comparative effectiveness study is mentioned. The comparison is between the new device and predicate devices/reference methods, not an assessment of human reader improvement with or without AI assistance.
    5. Standalone Performance (Algorithm Only):

      • Yes, the performance characteristics (specificity, sensitivity, reproducibility, and concordance) are described for the device itself (the IHC kit and its components) when used according to its intended purpose. This represents a standalone assessment of the assay's performance.
    6. Type of Ground Truth Used:

      • The ground truth for the concordance testing was established using the "Allred method." The Allred score is a semi-quantitative scoring system for ER/PR expression in breast cancer, combining the proportion of positive cells and the staining intensity. This can be considered a type of expert consensus/established semi-quantitative scoring method based on histopathology. For specificity, sensitivity, and reproducibility, the ground truth would also be based on established histopathological evaluation or comparison to the predicate devices, which are already considered validated.
    7. Training Set Sample Size:

      • Not specified. The document primarily focuses on the validation study demonstrating substantial equivalence. It does not provide details about any internal development or training data used in the creation of the assay itself, as this would typically be a chemical/biological assay rather than a machine learning algorithm requiring a separate "training set" in the computational sense.
    8. How Ground Truth for Training Set was Established:

      • Not applicable as this is not a machine learning device and no "training set" is described. The development of such IHC kits typically involves chemical formulation, antibody selection, and optimization against known positive and negative tissue samples, but these are not referred to as a "training set" in the context of this document.