510(k) Data Aggregation

    K Number
    K042772

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-12-13

    (69 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EBI® Array™ Spinal Fixation System is a non-cervical spinal fixation device intended for use as an anterolateral pedicle screw fixation system, a posterior hook and sacral/iliac screw fixations ystem, or as an anterplated for pedicle screw fixation system, a posterior november patients. The device is indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographie studiosis, spondylolisthesis, trauma, (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, spondylorisations) and lordosis), tumor, stenosis, psuedoarthrosis, and failed previous fusion.

    The EBI® VuePASS™ Portal Access System when used with the EBI® Array™ Spinal Firation System is I he EDI Vuel ABS

    The EBI® SpineLink II Spinal Fixation System is a non-cervical spinal fixation device intended for use as an anagerolateral The EBT Spinellilik II Spillar I Ration System.
    Pedicle screw fixation system, a posterior hook and sacrabiliates. The device is indicated for pedicle screw fixation system, a posterior nosition. In the patientive disactive is indicated for all of the following indications regardess of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studiosis discogenic back pain with defencialism or me
    spondylolisthesis, trauma, (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, spondylonstrations) in the many psuedoarthrosis, and failed previous fusion.

    The EBI® VuePASS™ Portal Access System when used with the EBI® Spinal Fixation The EBI * VuePASS * * Foral 766655 byers.
    System is indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery.

    Device Description

    The EBI® Array™ Spinal Fixation System is a single use spinal fixation device that utilizes rods, screws, and hooks. The EBT® VuePASS™ Portal Access Surgical System is a retractor system for a minimally invasive approach for posterior spinal surgery. The system includes a series of stainless steel sequential dilators and various sized reusable radiolucent oval cannulas and circular tubes. When used with the EBI® Array™ Spinal Fixation System, the use of the EBI® VuePASS™ Portal Access Surgical System is limited to the implant of a rod length of 50mm or less and excludes the use of system cross connectors or hooks.

    The EBI® SpineLink® II Spinal Fixation System is a fixation device that utilizes interconnecting links, screws (fixed and polydirectional designs), transverse connectors, hooks, endcaps, locknuts, and link ties. The EBI® VuePASS™ Portal Access Surgical System is a retractor system for a minimally invasive approach for posterior spinal surgery. The system includes a series of stainless steel sequential dilators and various sized reusable radiolucent oval cannulas and circular tubes. When used with the EBI® SpineLink® II Spinal Fixation System, the use of the EBI® VuePASS™ Portal Access Surgical System is limited to the implant of link lengths of 36mm or less and excludes the use of system cross links.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the EBI® Array™ Spinal Fixation System with the EBI® VuePASS™ Portal Access Surgical System, and the EBI® SpineLink® II Spinal Fixation System with the EBI® VuePASS™ Portal Access Surgical System:

    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the provided text, the submission focuses on demonstrating substantial equivalence to existing predicate devices rather than defining specific performance acceptance criteria with numerical targets. The "acceptance criteria" here is implicitly defined as successful implantation demonstrating substantial equivalence in design, intended use, and materials.

    Acceptance Criterion (Implicit)Reported Device Performance
    EBI® Array™ System: Substantial equivalence in design, intended use, and materials to the predicate EBI® Array™ Spinal Fixation System (K033312) when utilizing the EBI® VuePASS™ Portal Access Surgical System."Verification testing demonstrated successful implantation of the EBI® Array™ Spinal Fixation System utilizing the EBI® VuePASS™ Portal Access Surgical System. It is substantially equivalent to the predicate device in regards to design, intended use, and materials." (For EBI® Array™ System) "The EBI® Array™ Spinal Fixation System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and lordosis), tumor, stenosis, psuedoarthrosis, and failed previous fusion." (Intended Use from K042772)
    EBI® SpineLink® II System: Substantial equivalence in design, intended use, and materials to the predicate EBI® SpineLink® II Spinal Fixation System (K012516, K031355) when utilizing the EBI® VuePASS™ Portal Access Surgical System."Verification testing demonstrated successful implantation of the EBI® SpineLink® II Spinal Fixation System utilizing the EBI® VuePASS™ Portal Access Surgical System. It is substantially equivalent to the predicate device in regards to design, intended use, and materials." (For EBI® SpineLink® II System) "The EBI® SpineLink® II Spinal Fixation System is a non-cervical spinal fixation device intended for use as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients. The device is indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and lordosis), tumor, stenosis, psuedoarthrosis, and failed previous fusion." (Intended Use from K042772)

    2. Sample Size Used for the Test Set and Data Provenance

    The document only generically refers to "Verification testing". It does not specify a sample size for any test set or the data provenance (e.g., country of origin, retrospective or prospective). This type of information is typically detailed in a separate test report referenced in a 510(k) summary, but not explicitly provided in this summary itself.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention human experts being used to establish a "ground truth" for a test set. The term "verification testing" implies engineering or mechanical performance testing, not a clinical study involving expert assessment of outcomes or images.

    4. Adjudication Method for the Test Set

    As there's no mention of a test set requiring expert ground truth or assessment, there is no adjudication method described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done or reported in this 510(k) summary. This type of study would typically involve human readers (e.g., surgeons) evaluating cases with and without AI assistance, and measuring the effect size of the AI's improvement. The device in question (spinal fixation system and portal access system) is a physical surgical device, not a diagnostic AI system, making an MRMC study irrelevant in this context.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No standalone algorithm performance study was done or reported. The devices are physical surgical tools, not algorithms. The "verification testing" would assess the mechanical integrity and functional performance of the surgical tools themselves, likely through bench testing.

    7. Type of Ground Truth Used

    The concept of "ground truth" as typically applied to diagnostic AI (e.g., pathology, outcomes data) is not directly applicable here. The "ground truth" in this context would be the successful mechanical and functional performance of the medical devices as designed, verified through engineering tests matching established predicate device performance and safety standards. This is likely based on engineering specifications and test results.

    8. Sample Size for the Training Set

    The document does not specify a training set size. This is a physical device, not an AI algorithm that requires a training set in the conventional sense. If any form of "training" or development involved iterative testing, those details are not provided.

    9. How the Ground Truth for the Training Set Was Established

    As this is a physical device submission and not an AI submission, the concept of a "training set" and associated "ground truth" as used in AI development does not apply. The design and development of such devices are guided by engineering principles, biomechanical requirements, materials science, and conformity to established standards for medical implants. "Ground truth" for development would be the known and desired mechanical properties and biological compatibility standards.