Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K042713

    Validate with FDA (Live)

    Date Cleared
    2004-12-29

    (90 days)

    Product Code
    Regulation Number
    888.3060
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FIDJI type 5 is a vertebral body replacement device that is intended for use in the thoracic and/or thoracolumbar spine (T1-L5) to replace a collapsed, damaged or unstable vertebral body resected or excised (i.e., partial or total vertebrectomy procedures) due to turnor or trauma (i.e., fracture). The device is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period. This device is intended to be used with bone graft and supplemental fixation, such as the Spinal Concepts, Inc. InCompass System. Additionally, FIDJI type 5 is intended to be used in a paired configuration.

    Device Description

    The Spine Next FIDJI Vetebral Body Replacement Type 5 devices are hollow blocks that are machined from extruded Polyetheretherketone (PEEK); the blocks are tapered to aid in maintaining lordosis of the spine following implantation. The devices have angled teeth on the cephalad and caudal surfaces. The devices include tantalum inserts that serve as a location and orientation markers for radiographs.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the Spine Next FIDJI Vertebral Body Replacement Type 5. This type of submission relies on demonstrating substantial equivalence to a predicate device rather than conducting new clinical studies with specific acceptance criteria as you might see for a PMA (Premarket Approval) application for a novel device.

    Therefore, the information you're asking for regarding acceptance criteria, specific study details (sample sizes, experts, adjudication methods, MRMC studies, standalone performance), and ground truth establishment, is generally not applicable to a 510(k) submission like this one. 510(k)s focus on demonstrating that the new device is as safe and effective as a legally marketed predicate device, often through engineering and performance testing that shows the new device performs similarly within established standards, rather than generating new clinical data against specific performance targets.

    However, I can extract the relevant information from the provided text that aligns with what is typically done for a 510(k) in terms of "comparison to predicate" which serves as the "proof" of meeting the implicit acceptance criteria of substantial equivalence.


    Acceptance Criteria and Device Performance (Implicit through Predicate Comparison)

    For a 510(k) submission, the "acceptance criteria" are implicitly met by demonstrating "substantial equivalence" to a legally marketed predicate device. This is primarily done by showing similar technical characteristics, indications for use, and performance. When comparing to a predicate, the "performance" often refers to engineering and bench testing rather than clinical study outcomes.

    Acceptance Criteria (Demonstration of Substantial Equivalence to Predicate)Reported Device Performance (FIDJI vs. Predicate)
    Similar Indications for UseMet: Both Spine Next FIDJI and predicate devices are intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed or unstable vertebral body due to tumor or trauma. The FIDJI's indications are consistent with the predicate.
    Similar Materials of ConstructionMet: Both the Spine Next FIDJI and the predicate devices (Vertebral Spacer and Vertestack) are fabricated from PEEK Optima LT.
    Similar Principle of Operation and FunctionMet: Both are vertebral body replacements, hollow blocks with tapered and angled surfaces. They are "virtually identical in their function and principle of operation."
    Similar Load Bearing CapacityMet: Both devices are load-bearing, intended for use under load conditions. The submission states, "All devices meet mechanical performance requirements." Specific mechanical test data would typically be in the full submission, but this summary indicates compliance.
    Similar BiocompatibilityMet: Both are reported as biocompatible.
    Availability in Various SizesMet: Both are provided in various sizes.
    Sterile PresentationMet: Both are sold sterile.
    Similar Use/HandlingMet: While "subtle design differences will require minor adjustments to procedure by the implanting surgeon," the overall use and handling are considered similar.

    Study Details Not Applicable to This 510(k) Submission

    As this is a 510(k) submission, the device did not undergo a clinical study in the traditional sense that would generate new clinical performance data. Instead, it demonstrated equivalence to existing predicate devices based on design, materials, and mechanical performance testing (implied by "All devices meet mechanical performance requirements"). Therefore, the following information is not provided or required for this type of submission:

    1. Sample size used for the test set and the data provenance: Not applicable for a 510(k) unless specific clinical data was used to demonstrate equivalence (which is not indicated here). The "test set" would primarily refer to engineering bench tests.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as clinical ground truth establishment is not typically part of a 510(k) for this type of implant.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done: No, this is not indicated and generally not a requirement for an orthopedic implant 510(k).
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a physical implant, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable to a 510(k) for an orthopedic implant. Ground truth for performance would come from validated mechanical testing methodologies against industry standards or predicate device performance.
    7. The sample size for the training set: Not applicable, as this is not an AI/ML device requiring a training set.
    8. How the ground truth for the training set was established: Not applicable.

    In summary, the "study" demonstrating the device meets "acceptance criteria" for a 510(k) is the comparison to predicate devices, primarily focusing on technical characteristics, materials, and functional equivalence, often backed by bench testing to ensure mechanical performance is comparable. The FDA's decision to clear the device (K042713) signifies their acceptance that this comparison adequately demonstrates substantial equivalence to legally marketed predicate devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1