510(k) Data Aggregation

    K Number
    K041853

    Validate with FDA (Live)

    Device Name
    OTOGRAM
    Manufacturer
    Date Cleared
    2004-09-03

    (57 days)

    Product Code
    Regulation Number
    874.3310
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    "Otogram" can perform the functions of:
    Pure tone and speech audiometer, 1.
    Impedance audiometer (tympanometry, acoustic reflex), and 2.
    Distortion product otoacoustic emissions analyzer. 3.
    The unit is intended for use by a qualified/trained audiologist on both adult and pediatric subjects for measurement of acoustic impedance.

    Device Description

    The Otogram is a computer-controlled, audiometric instrument combining the functions of an Audiometer, a Impedance Instrument, and a DPOAE Instrument. The device is controlled through the use of ASCII commands transmitted over a standard RS-232 communication port. Control software is installed on the supplied PC. The device performs comprehensive audiometry; tympanometry, acoustic reflex; and otoacoustic emissions. Additional features include manual audiometer and NOAH v.3 compliant (hearing aid fitting software). Available in multiple languages, including English, Spanish, Russian, Mandarin, Vietnamese, and Korean.

    AI/ML Overview

    The provided text describes the regulatory submission for the Tympany Otogram device and its comparison to predicate devices, focusing on substantial equivalence rather than a study proving the device meets specific acceptance criteria in a clinical performance sense.

    Therefore, many of the requested elements (like sample size for test/training sets, number and qualifications of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and type of ground truth) either do not apply directly to this type of submission or are not detailed in the provided information. This document is a 510(k) summary, which focuses on demonstrating equivalence to existing legally marketed devices based on technical characteristics and intended use, often supported by bench and user testing rather than extensive clinical trials with human subjects.

    Here's an attempt to answer the questions based only on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state "acceptance criteria" in the format of specific quantitative thresholds to be met. Instead, it demonstrates "substantial equivalence" to predicate devices by comparing technical characteristics. The "reported device performance" is essentially the detailed technical specifications of the Otogram and its comparison to the predicate devices.

    The table below reconstructs the comparison from the provided Substantial Equivalence Chart as effectively serving the role of demonstrating performance relative to established benchmarks (the predicate devices). The implied "acceptance criterion" for each characteristic is that the Otogram's performance/specification is comparable to or within acceptable ranges of the predicate devices.

    CharacteristicAcceptance Criterion (Predicate Devices)Reported Device Performance (Tympany Otogram)
    Probe tone:
    Probe tone frequencyMadsen Zodiac 901: 226 Hz226 Hz, +/- 3%
    Probe tone levelMadsen Zodiac 901: 85 dB SPL70 dB SPL, +/- 3 dB
    Admittance measurement:
    Total rangeMadsen Zodiac 901: 0.1 ml to 8.0 ml0.2 ml to 6.0 ml
    Reflex rangeMadsen Zodiac 901: 0 ul to 375 ul0 ul to 375 ul
    CalibrationMadsen Zodiac 901: 2 cc cavity0.5, 2, 5 cc cavity
    Pressure system:
    Pump systemMadsen Zodiac 901: Plunger type; stepper motor, digitally controlledPeristaltic pump; stepper motor, digitally controlled
    Pressure range - normalMadsen Zodiac 901: +200 to -400 daPa+200 to -400 daPa
    Pressure range - extendedMadsen Zodiac 901: +400 to -600 daPaNA
    AccuracyMadsen Zodiac 901: +/- 10% or +/- 10 daPa (whichever is greater)+/- 15% or +/- 10 daPa (whichever is greater)
    Pump speedMadsen Zodiac 901: As fast as possible: 400 daPa/sec, 200 daPa/sec, 100 daPa/sec, 50 daPa/sec50 daPa/sec to 150 daPa/sec
    Air releaseMadsen Zodiac 901: Mechanical safety release valve set to +600 and -800 daPa. Manual & automatic air releaseMechanical safety release valve set to +600 and -800 daPa. Manual & automatic air release
    Ipsi/contralateral stimuli:
    Pure toneMadsen Zodiac 901: 0.5, 1, 2, 4 kHz0.5, 1, 2, 4 kHz
    Step sizeMadsen Zodiac 901: 1, 2, 5, 10 dB1 to 10 dB
    Attenuator range, ipsi lateral stimulusMadsen Zodiac 901: 0.5, 1.0, 2.0, kHz 50-115 dB HL, 4.0 kHz 50-90 dB HL0.5, 1.0, 2.0, 4.0 kHz 80-105 dB HL
    Attenuator range, contra lateral stimulusMadsen Zodiac 901: 0.5, 1.0, 2.0, 4.0 kHz 50-100 dB HL, white noise0.5, 1.0, 2.0, 4.0 kHz 80-100 dB HL
    Frequency accuracyMadsen Zodiac 901: +/- 0.5% (also Bio-Logic AuDX: +/- 0.5%)+/-0.5%
    Distortion Product Otoacoustic Emissions:
    F2 test frequencyBio-Logic AuDX: 2.0, 3.0, 4.0, 5.0 kHz1.0, 1.5, 2.0, 3.0, 4.0 kHz
    Stimulus level rangeBio-Logic AuDX: 55 dB SPL and 65 dB SPL45-70 dB SPL
    Standard and safety characteristics:
    Performance and calibrationIEC 1027, ANSI S3.39 (for Madsen Zodiac 901)IEC 1027, ANSI S3.39
    Electrical safetyEN 60601-1, class I, type B (for Madsen Zodiac 901); EN 55011: 1991 Group 1 Class B (for Bio-Logic AuDX)EN 60601-1, class I, type B

    The study proving the device meets these criteria is described as "bench, user, and laboratory testing" (Section 5. and 7.). These tests indicate that the new device is "as safe and effective as the predicate device," thus demonstrating substantial equivalence.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified. The document mentions "bench, user, and laboratory testing" but does not quantify the sample size (e.g., number of devices tested, number of users, number of measurements).
    • Data Provenance: Not specified. It's internal company testing ("Tympany Inc."). There is no indication of country of origin or whether the testing was retrospective or prospective in detail, although "user testing" implies prospective interaction with the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable/Not specified. This is a technical medical device, and the "ground truth" for its performance is typically established by comparing its physical outputs (e.g., sound frequencies, pressure measurements) against calibrated standards and predicate devices, rather than expert consensus on a diagnosis for a test set. The submission focuses on technical performance and safety.

    4. Adjudication method for the test set

    Not applicable/Not specified. Adjudication methods like 2+1 or 3+1 are typically used in studies where human readers evaluate cases and their agreement needs to be resolved. This document describes bench and user testing of a medical device's technical specifications and safety, not an evaluation of diagnostic accuracy by human readers.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This type of study (MRMC, AI assistance) is not mentioned and is typically associated with diagnostic algorithms or imaging interpretation, which is not the primary focus of this device's regulatory submission. The Otogram is a measurement device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable for this device in the context of "algorithm only." The performance of the device itself (its hardware and embedded software) was tested, implying a "standalone" evaluation of its technical capabilities against its specifications and predicate devices. However, this is not in the context of an "AI algorithm." The device's operation is "computer-controlled" (Section 4), but the submission focuses on its measurement functions, not on a standalone diagnostic algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the device's technical performance would be established by calibrated standards and the specifications of predicate devices. For example, to verify "Probe tone frequency: 226 Hz, +/- 3%", the device's output would be measured against a known, highly accurate frequency standard. For safety, compliance with standards like EN 60601-1 is the ground truth. There's no mention of patient-specific "outcomes data" or "pathology" as ground truth for this type of device submission.

    8. The sample size for the training set

    Not applicable. The device described (Otogram) is an audiometer, impedance instrument, and DPOAE instrument. It is a measurement device, not an AI/machine learning model that requires a "training set" in the conventional sense.

    9. How the ground truth for the training set was established

    Not applicable (as established in point 8).