510(k) Data Aggregation

    K Number
    K041337

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-05-28

    (9 days)

    Product Code
    Regulation Number
    890.3850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for medical purposes to provide mobility to persons restricted to a seated position.

    Device Description

    The PRO WALKER ML-300 Foldable Wheelchair is an indoor / outdoor wheelchair that has a base with four-wheeled with a seat. The device can be disassembled for transport and it is foldable easily. The device uses a standard sling type back and seat, the upholstery fabric meets the California Technical Bulletin CAL 117 standard for flame retardant.

    AI/ML Overview

    This 510(k) premarket notification for the PRO WALKER ML-300 Foldable Wheelchair does not describe a clinical study of the device's performance against acceptance criteria in the way a diagnostic or AI-driven medical device submission would. Instead, it demonstrates substantial equivalence to a predicate device through adherence to recognized performance standards for mechanical wheelchairs.

    Here's an analysis based on the provided text, focusing on how a device like this meets its "acceptance criteria":

    1. Table of Acceptance Criteria and Reported Device Performance

      Acceptance Criteria (Performance Requirements)Reported Device Performance (Compliance)
      ANSI/RESNA WC Vol. 1 Wheelchair StandardsMeets applicable performance requirements
      ISO 7176 Wheelchair StandardsMeets applicable performance requirements
      CA Technical Bulletin CAL 117 (Flame Retardant)Upholstery fabric meets standard
      Intended Use: Provide mobility to persons restricted to a seated positionDevice intended for this purpose
      Weight Limit: 100 kgsSame as predicate device (100 kgs)
      FoldabilityFoldable easily, essential feature same as predicate
    2. Sample Size used for the test set and the data provenance:
      The document does not specify a "test set" in terms of patient data or
      clinical images. The "testing" refers to mechanical and material
      performance evaluations of the wheelchair itself. Therefore, sample sizes
      would refer to the number of wheelchairs or components tested to verify
      compliance with the standards. This information is not explicitly provided,
      but typically involves a limited number of units or material samples from
      manufacturing for destructive and non-destructive testing per the
      standards.

      • Data Provenance: Not applicable in the context of clinical data. The
        testing would have been conducted by the manufacturer or a certified
        testing lab. The submitter's address is in Taiwan.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      Not applicable. "Ground truth" in this context refers to the defined
      parameters and limits within the ANSI/RESNA and ISO standards for
      wheelchair performance, as well as the California Technical Bulletin for
      flame retardancy. These standards themselves are established by expert
      committees in their respective fields (e.g., biomechanical engineers,
      rehabilitation specialists for wheelchair standards; fire safety experts for
      flammability standards). The submission implies that the device met these
      pre-established criteria, not that new ground truth was created by experts
      for this specific device's test set.

    4. Adjudication method:
      Not applicable. There's no clinical data or expert review process
      described that would require an adjudication method. Compliance with
      standards is typically determined by testing results falling within
      defined ranges or passing specific procedures.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      Not applicable. This device is a mechanical wheelchair and does not involve AI or human "readers" interpreting data.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Not applicable. This device is a mechanical wheelchair and does not
      involve an algorithm.

    7. The type of ground truth used:
      The "ground truth" for this device's performance is compliance with established international and national performance standards for wheelchairs (ANSI/RESNA WC Vol. 1 and ISO 7176) and material safety standards (California Technical Bulletin CAL 117 for flame retardance). These standards define objective, measurable criteria for safety and performance (e.g., stability, fatigue, static strength, braking effectiveness, flammability).

    8. The sample size for the training set:
      Not applicable. This device is a mechanical wheelchair; there is no "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established:
      Not applicable.

    Summary of the Study that Proves the Device Meets Acceptance Criteria:

    The PRO WALKER ML-300 Foldable Wheelchair demonstrated that it meets its acceptance criteria through performance testing against recognized national and international standards for mechanical wheelchairs and fire retardancy. The submission states, "ML-300 Foldable Wheelchair meet the applicable performance requirements as specified in ANSI/RESNA WC vol. 1 and ISO 7176 Wheelchair Standards... and the upholstery fabric meets the California Technical Bulletin CAL 117 standard for flame retardant."

    This type of submission for a Class I device typically relies on engineering and laboratory testing to assure compliance with these established standards, rather than clinical trials with human subjects. The substantial equivalence argument further reinforces that the device's design, materials, and performance characteristics are comparable to a device already deemed safe and effective (the BIOTECH B900 SUPER LIGHT Wheelchair, K020472), with any differences (like overall dimension and weight) not impacting safety aspects.