Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K040992

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-07-29

    (104 days)

    Product Code
    Regulation Number
    876.5540
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    THE MEDCOMP EXCELL™ SPLIT-TIP CATHETER IS INDICATED FOR USE IN ATTAINING LONG TERM VASCULAR ACCESS FOR HEMODIALYSIS AND APHERESIS. IT MAY BE INSERTED PERCUTANEOUSLY AND IS PRIMARILY PLACED IN THE INTERNAL JUGULAR VEIN. ALTERNATE INSERTION SITES INCLUDE THE SUBCLAVIAN VEIN AS REQUIRED. CATHETERS GREATER THAN 40CM ARE INTENDED FOR FEMORAL VEIN INSERTION.

    Device Description

    The Medcomp Excell™ Split-Tip Catheter is a 15F polyurethane, double lumen catheter used to remove and return blood through two, segregated furnen passages. Both lumens are "D" shaped, open at the distal tip, with two side holes. The distal venous lumen is tapered and extends beyond the arterial lumen to reduce recirculation. The fixed polyester cuff allows for tissue ingrowth for long-term placement. The arterial and venous lumens now are designed to be split, or peeled apart, prior to insertion to provide two free-floating lumens within the vessel. The lumens are connected to the extensions via a soft pliable hub with suture wing. The arterial and venous extensions are identified by red and blue luer connectors and clamps. Priming volume information is printed on the I.D. rings for ease in identification.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Medcomp Excell™ Split-Tip Catheter:

    Acceptance Criteria and Device Performance

    The provided document describes the modification of a legally marketed device and asserts its substantial equivalence to predicate devices. The "acceptance criteria" are implied by the performance data demonstrating "substantial equivalence." However, explicit, quantifiable acceptance criteria with pass/fail thresholds are not stated in the document.

    The performance data focuses on demonstrating that the modified device performs similarly to existing, legally marketed devices for hemodialysis and apheresis treatments.

    Here's a table based on the provided information:

    Acceptance Criteria (Implied)Reported Device Performance
    Substantial equivalence to legally marketed predicate devicesDemonstrated through in vitro testing.
    Performance in recirculation (similar to predicate devices)Data indicates performance is "substantially equivalent."
    Performance in gravity flow (similar to predicate devices)Data indicates performance is "substantially equivalent."
    Performance in flow vs. pressure (similar to predicate devices)Data indicates performance is "substantially equivalent."
    Lumen peel strength/integrity (similar to predicate devices)Data indicates performance is "substantially equivalent."
    Safety (similar to predicate devices)Clinical studies were not deemed necessary as in vitro testing was sufficient to demonstrate safety by way of comparison to legally marketed predicate devices.
    Effectiveness (similar to predicate devices)Clinical studies were not deemed necessary as in vitro testing was sufficient to demonstrate effectiveness by way of comparison to legally marketed predicate devices.
    Specific quantified criteria (e.g., recirculation <X%, flow >Y ml/min)Not explicitly stated in the provided document. The criteria are implicitly tied to the performance characteristics of the predicate devices.

    Note regarding "Acceptance Criteria": The document does not explicitly list quantified acceptance criteria (e.g., "recirculation must be less than 5%"). Instead, it states that the device was tested to demonstrate "substantial equivalence" to predicate devices for specific performance characteristics. This implies that the acceptance criteria were met if the device's performance in these tests was comparable to that of the predicate devices, falling within an acceptable range for the intended use.


    Study Details (Based on the provided text)

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not specified. The study involved "in vitro performance data," which typically refers to laboratory testing and might involve multiple samples of the device, but the exact number is not provided.
      • Data Provenance: The study was an in vitro performance test, meaning it was conducted in a laboratory setting, not on human subjects. The country of origin for the data is not explicitly mentioned, but the submitter is a U.S. company (Harleysville, PA). It is by definition a prospective study in that the testing was performed on the device designed for submission.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Number of Experts: Not applicable. This was an in vitro performance study comparing the device to predicate devices. There was no "ground truth" established by human experts in the context of clinical observations or diagnoses. Instead, the performance parameters of the device were measured and compared.
      • Qualifications of Experts: Not applicable for establishing ground truth.
    3. Adjudication method for the test set:

      • Adjudication Method: Not applicable. As an in vitro performance study, there was no need for expert adjudication of results in the way it would be required for clinical image analysis or diagnostic studies. The results were likely derived from standardized test protocols and measurements.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • MRMC Study: No, an MRMC comparative effectiveness study was not done. This device is a medical catheter, not an AI software intended for interpretation by human readers.
      • Effect Size of AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Standalone Performance: Not applicable. This is a physical medical device, not an algorithm. Therefore, "standalone" performance for an algorithm doesn't apply. The in vitro testing represents the "standalone" performance of the physical device in a controlled environment.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Type of Ground Truth: Not applicable in the traditional sense of clinical "ground truth." For in vitro performance studies, the "ground truth" is typically defined by standardized physical measurements, engineering specifications, and the established performance characteristics of the predicate devices. The "truth" is that the device should perform comparably to its predicates under controlled test conditions.
    7. The sample size for the training set:

      • Sample Size for Training Set: Not applicable. This is a physical medical device, not a machine learning model. There is no concept of a "training set" for the device itself.
    8. How the ground truth for the training set was established:

      • Ground Truth for Training Set: Not applicable for the reasons stated above (physical device, no machine learning model).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1