Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K032497

    Validate with FDA (Live)

    Date Cleared
    2003-09-25

    (43 days)

    Product Code
    Regulation Number
    882.5950
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sapphire Detachable Coils are intended for the endovascular embolization of intracranial aneurysms that - because of their morphology, their location, or the patient's general medical condition - are considered by the treating neurosurgical team to be a) very high risk for management by traditional operative techniques, or b) be inoperable.

    Device Description

    The Sapphire Detachable VDS™ Tension Safe Coil is manufactured from a platinum alloy wire which is first wound into primary coil and then formed into a secondary helical shape. The VDS Tension Safe has multiple coil segments with polymer "cap" at each segment ends, integrated into one coil system. The coil is welded to a positioning wire, which consist of ground stainless steel core wire with a stainless steel coil laser welded at the distal end and a Teflon outer jacket. The coil is detached by the battery operated power supply (SDS), which dissolves a small detachment element between the emboliozation coil and the positioning wire.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Sapphire Detachable Coil System, VDS™ Tension Safe, focusing on your requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Bench Testing (Acceptance Criteria)Reported Device Performance
    Coil DeformationMeet established acceptance criteria
    Ease of Delivery/Coil Frictional CharacteristicsMeet established acceptance criteria
    Reliability After Fatigue & Premature DetachmentMeet established acceptance criteria
    Tensile Strength of Coil & Detachment ZoneMeet established acceptance criteria
    Detachment TimeMeet established acceptance criteria
    Particulate GenerationMeet established acceptance criteria
    Physical DimensionsMeet established acceptance criteria
    RadiopacityMeet established acceptance criteria
    Aging and Shelf Life (6 months)Meet established acceptance criteria (for 6 months)

    Note: The document explicitly states "Meet established acceptance criteria" for each test, implying that specific quantitative or qualitative thresholds were defined and met. However, the exact numerical values or detailed qualitative descriptions of these criteria are not provided in the given text.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not explicitly state the sample size used for the test set for any of the "Bench Testing" categories. It also does not mention the data provenance (e.g., country of origin, retrospective or prospective) for these tests. The tests appear to be conducted in a laboratory or bench setting rather than using patient data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable and therefore, not provided. The acceptance criteria and performance are based on bench testing, not on clinical evaluation or expert interpretation of medical images.

    4. Adjudication Method for the Test Set

    This information is not applicable and therefore, not provided. The "tests" mentioned are bench tests of mechanical and physical properties, not clinical assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating substantial equivalence to a predicate device through bench testing of physical and mechanical properties. There's no mention of human reader performance or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable and therefore, not provided. The "device" is a physical medical implant (an embolization coil system), not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for the bench tests would have been the defined engineering specifications, performance standards, and material properties that the device was designed to meet. For instance:

    • Coil Deformation: A specific tolerance for deformation under a defined load.
    • Tensile Strength: A minimum tensile strength requirement.
    • Detachment Time: A specified range for detachment time.
    • Physical Dimensions: Exact dimensional specifications.

    However, the document does not explicitly detail these ground truths beyond stating that the device "Meet established acceptance criteria."

    8. The Sample Size for the Training Set

    This information is not applicable and therefore, not provided. As mentioned, the device is a physical medical implant, not an AI or algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and therefore, not provided, for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1