    Intended Use

    E-scan XQ is a magnetic resonance (MR) system that produces transversal, sagittal and coronal and oblique cross-section images of the limbs and joints. It is intended for imaging the arm, including the hand, wrist, forearm, elbow, upper arm and shoulder, and imaging the leg, including the foot, ankle, calf, knee, thigh and hip.

    E-scan XQ MR images correspond to the spatial distribution of protons (hydrogen nucle) that determine magnetic resonance properties and are dependent on the MR parameters, including spin-latice relaxation time (TI), spin-spin relaxation time (T2), nuclei density, flow velocity and "chemical shift". When interpreted by a medical expert trained in the use of MR equipment, the images can provide diagnostically useful information.

    Device Description

    The E-scan XQ is a magnetic resonance (MR) system. The modified system is composed of a patient positioning table, magnetic unit, operating console (PC unit, keyboard, mouse, monitor, operating table), electronics box with filter panel, and modular shielding box. The modifications from the cleared version (K020164) include a modified system configuration, new operating tables, upgrading of the electronics, and a new software release. These modifications are stated to not affect the intended use or alter the fundamental scientific technology of the device.

    AI/ML Overview

    This document is a 510(k) premarket notification for modifications made to the E-Scan XQ MRI system. It's not a study report proving a device meets acceptance criteria through clinical or technical performance testing with a specific outcome measure like diagnostic accuracy. Instead, it demonstrates substantial equivalence to a previously cleared device (E-Scan XQ K020164) based on similar technological characteristics and unchanged intended use.

    Therefore, many of the requested categories (acceptance criteria, reported performance, sample size, ground truth, expert consensus, MRMC study, standalone performance) are not applicable (N/A) in the context of this type of submission. The comparison is primarily against the specifications of the predicate device, not clinical performance metrics against a defined standard.

    However, I can extract the information provided regarding the modification and the comparison to its predicate.

    Here's a summary of the relevant information:

    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a 510(k) amendment for modifications, the "acceptance criteria" are implicitly the maintenance of, or improvement upon, the technological characteristics of the predicate device, while ensuring the intended use is not altered. The "reported device performance" is a comparison of the modified device's technical specifications against the predicate.

    CharacteristicPredicate Device (E-scan XQ K020164)Modified Device (E-scan XQ modified)Acceptance / Performance Comment
    Pulse SequencesOrthogonal Multi-planar Scout, Spin Echo T1 (set1), Spin Echo T2 (set2), Multi-Echo (se_pd_t2), Inversion Recovery (ir), Short TI Inversion Recovery (stir), Spin Echo Half Echo (set1he), Spin Echo Half Scan (set1hf), Turbo SE T2 weighted and Turbo ME (tse, tme), Gradient Echo (ge), Short Time Inversion Recovery Gradient Echo (ge_stir), Gradient Echo 3D (t3d_t1), Gradient Echo 3D contrast enhancement (3d_ce), Real TimeOrthogonal Multi-planar Scout, Spin Echo T1 (set1), Spin Echo T2 (set2), Multi-Echo (se_pd_t2), Inversion Recovery (ir), Short TI Inversion Recovery (stir), Spin Echo Half Echo (set1he), Spin Echo Half Scan (set1hf), Turbo SE T2 weighted and Turbo ME (tse, tme), Gradient Echo (ge), Short Time Inversion Recovery Gradient Echo (ge_stir), Gradient Echo 3D (t3d_t1), Gradient Echo 3D contrast enhancement (3d_ce), Real TimeUnchanged (Meets acceptance criterion of maintaining existing imaging capabilities)
    Sequence Parameters (High Res)Not explicitly detailed as "High Resolution" sequences with fixed TEs.Added: High Resolution se_26 (TE=26ms), se_26_hf (TE=26ms), se_18_he (TE=18ms), tse_80 (TE=80ms), tse_50 (TE=50ms), tme (first echo 28ms, second echo 90ms), ge_16 (TE=16ms), ge_stir_25 (TE=25ms)New feature/improved performance: These are particular versions of standard sequences with maximum acquisition matrix 512x512 (instead of 256x256 of non-High Resolution), aiming for best compromise between S/N and high resolution. This is an improvement.
    Acquisition Matrix2D FT: 192x128 to 256x256; 3D FT: 192x128 to 256x256, slice encoding 24 to 128.2D FT for non High Resolution: 192x128 to 256x256; 2D FT for High Resolution: 192x128 to 512x512; 3D FT: 192x128 to 256x256, slice encoding 24 to 128.Improved performance: Increased raw data matrix dimensions (up to 512x512 for High Resolution sequences) compared to the prior max of 256x256, allowing for higher resolution images.
    Fringe Field (0.5 mT line)X: 1.5 m front; 1.2 m rear; Y: 1.3 m; Z: 1.5 mX: 1.34 m front; 1.06 m rear; Y: 1.25 m; Z: 1.29 mMore precise characterization of data (Specification update, not necessarily a performance change in itself, but implies refinement)
    Gradients System - Control SystemDigital, based on DSP SHARC 66 Mips, 132 MFlops, 0.5 Mbit Memory; 1.1515 nsec Instruction rate; 4 independent channels (X - Y - Z - Bo); DAC 18 bit - updating every 7.2 µs.Digital electronic, based on DSP SHARC 21161@100 MHz, 400 MFLOPs, 128 KB On-Chip SRAM; 4 independent channels (X,Y,Z, Bo); DAC 18 bit - update every 7,2 µs.Technological updating: Improved DSP (higher MFLOPs, increased memory) suggests potential for improved control and efficiency.
    Magnetic Compensation SystemDigital electronic based on DSP SHARC 66 Mips, 132 MFlops, 0.5 Mbit Memory.Digital electronic, based on DSP SHARC 21161@100 MHz, 400 MFLOPs, 128 KB On-Chip SRAM.Technological updating: Improved DSP suggests potential for more robust magnetic field control.
    A/D ConversionBaseband conversion; 2 A/D converter 16 bit 78 kHz sampling; 2 digital low-pass filter (bandwidth up to 0.464xfc, fc = 78.125 KHz/n).For each channel: 3 MHz conversion of RF signal; A/D converter 14 bit 20 MHz sampling; digital demodulation; digital low-pass filter (bandwidth up to 0.43 x fc, fc from 156.25 to 4.882 kHz).Technological updating / Improved performance: Higher sampling rate (20 MHz vs 78 kHz) and different conversion method (3 MHz RF conversion vs baseband) indicates a significant upgrade in signal acquisition.
    SynthesizerDigital, through DSP SHARC 66 Mips, 132 MFlops, 0.5 Mbit Memory; resolutions: 1.2 Hz frequency, 256 levels amplitude, 1°4° phase; stability < 1 ppm.Digital, through DSP SHARC 21161@100 MHz, 400 MFLOPs, 128 KB On-Chip SRAM; resolutions: 28.4 µHz frequency, 4096 levels amplitude, 5'16" phase; stability < 1 ppm.Technological updating / Improved performance: Significantly finer frequency, amplitude, and phase resolutions suggest a more precise and stable RF excitation.
    Transmission CoilLinear saddle shape coil; IN-impedance = 50 ohm; passive detuning during receiving through BYW29 diodes.Linear saddle shape coil; IN-impedance = 50 ohm; active detuning during receiving through PIN diodes -5 V 100 mA.Improved performance: Active detuning with PIN diodes avoids distortion of lateral lobes of the transmitting pulse, which was an issue with BYW29 diodes, leading to image quality enhancement.
    Transmission ChainRF power amplifier until 600 W pep; bandwidth 7-9 MHz; gain stability 0.1 dB.RF power amplifier until 900 W pep; bandwidth 7-9 MHz; gain stability 0.1 dB.Improved performance: Increased RF power amplifier output (900 W vs 600 W) allows for potentially faster or more robust excitation.
    Central Processing Unit (CPU)CPU Pentium III 850 MHz or faster; main memory: 256 MB; secondary cache: 256 KB or upper.CPU Pentium IV 2.4 GHz or more; Main memory: 1 GB; Secondary cache: 512 KB or more.Technological updating: Significant upgrade in CPU speed, main memory, and cache, indicating substantial improvement in processing power.
    Control Processor36 MIPS, 72 MFlops, 0.25 MB memory.DSP SHARC 21161@100 MHz, 400 MFLOPs, 128 KB On-Chip SRAM + 384 KB SSRAM.Technological updating: Improved DSP (higher MFLOPs, more memory) for control.
    Acquisition & Reconstruction Processors36 MIPS, 72 MFlops, 0.5 MB memory + 128 MB DRAM.DSP SHARC 21161@100 MHz, 400 MFLOPs, 128 KB On-Chip SRAM + from 256 to 512 MB SDRAM + 1.28 MB SSRAM.Technological updating: Improved DSP and increased memory for acquisition and reconstruction.
    Hard Disk Unit3"1/2 hard disk; at least 20 GB, 7200 rpm.3"1/2 hard disk; at least 40 GB, 7200 rpm.Technological updating: Increased hard disk capacity.
    Image Reconstruction Matrix2D: 256x256; 3D: 256x256x8 to 256x256x128.2D: 128x128, 256x256, 512x512; 3D: 256x256x24 to 256x256x128.Improved performance: Added 512x512 visualization matrix for high-resolution sequences.
    Display Monitor19" CRT Colour OR 17" TFT Colour; 1280 x 1024 pixel; at least 70 Hz.18.1" TFT Color; 1280x 1024 pixel; not interleaved 60 Hz.Technological updating: Updated monitor specification (specific TFT size, 60 Hz not-interlaced).
    Patient Table Cushion MaterialWild Heather Ambla (DBL) foamed PVC on a fire retardant cotton support.Coverlim 27 New (LIMONTA) 56% Cotton 44% Polyurethane.Biocompatibility: New material evaluated as part of "Biocompatibility Summary" (not detailed in this section but mentioned). Maintaining safety requirements.
    Vibrations< -70 dBg da 0 a 100 Hz.< -65 dBg from 0 to 100 Hz; warning range with standard feets: 9-13 Hz; warning range with vibration-damping springs: 4-7 Hz.Improved adaptability/performance: Optional vibration-damping springs can lower magnet resonance frequency (from 11 Hz to 5.9 Hz) to enhance image quality in sites with significant vibrations (9-13 Hz range), by reducing ghost artifacts caused by magnetic field variations.
    Power Consumption1100 VA (quick heating); 800 VA (normal working); 200 VA (powered off).1000 VA (quick heating); 800 VA (normal working); 200 VA (powered off).More precise characterization of the data (Slight reduction in quick heating power, likely an optimization).
    Shielding BoxMade of Fe or Al.Only Al version available.Product streamlining/update: Fe shielding box no longer available. (Assumed that Al version meets shielding requirements).
    DICOM server/visualizerESA-VIEW Rel. 1.ESA-VIEW Rel.2.Technological updating: Updated software version. The ESA-VIEW Rel.2 is a cleared DICOM server/visualizer (K012093), ensuring continued compliance and functionality.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    N/A. This is a 510(k) for device modifications, not a clinical study. The changes relate to hardware and software upgrades, and their impact is assessed through engineering analysis and comparison to the predicate device's established performance specifications, not a test set of images or patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    N/A. No test set requiring expert ground truth was mentioned for this type of submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    N/A. No test set was mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A. This submission predates widespread AI integration in medical devices and is focused on hardware/software upgrades of an MRI scanner, not diagnostic AI performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    N/A. This is not an algorithm-only device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    N/A. The "ground truth" for this engineering submission is compliance with established technical specifications and demonstration of substantial equivalence to the predicate device without altering the fundamental scientific technology or intended use.

    8. The sample size for the training set

    N/A. This is a medical device modification notification, not a machine learning model.

    9. How the ground truth for the training set was established

    N/A. Not applicable as above.