510(k) Data Aggregation

    K Number
    K031852

    Validate with FDA (Live)

    Date Cleared
    2003-08-20

    (65 days)

    Product Code
    Regulation Number
    882.5950
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sapphire Detachable Coils are intended for the endovascular embolization of intracranial aneurysms that - because of their morphology, their location, or the patient's general medical condition - are considered by the treating neurosurgical team to be a) very high risk for management by traditional operative techniques, or b) be inoperable.

    The intended uses of the Sapphire fiber coils are narrowed relative to the predicate device GDC, but still fall within the 510(k) cleared intended use.

    Sapphire Detachable Fibered Coil - The Sapphire fibered coil is intended for embolization of neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Fibered Sapphire is also intended for arterial and venous embolizations in the peripheral vasculature.

    Device Description

    The Sapphire Detachable Coil (SDC) is manufactured from a platinum alloy wire which is first wound into primary coil and then formed into a secondary helical shape. The coil is welded to a positioning wire, which consist of ground stainless steel core wire with a stainless steel coil laser welded at the distal end and a Teflon outer jacket. The coil is detached by the battery operated power supply (Sapphire Detachment System, SDS), which dissolves a small detachment element between the emboliozation coil and the positioning wire. The fibered coil is manufactured with nylon fibers secured into the primary coil. The fibered coils are available in two shapes (Helix and Cyclone) and different sizes.

    AI/ML Overview

    The Sapphire Detachable Fibered Coil System demonstrates its acceptance criteria and performance through a series of bench, animal, and physician evaluation tests. The device's substantial equivalence to the predicate device, Target GDC fibered coils, is asserted based on similar intended use, operating principles, basic design, and sterilization processes.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategorySpecific TestAcceptance Criteria & Reported Performance
    Bench TestingCoil Strength (Coil Deformation and Softness)Meet established acceptance criteria
    Ease of DeliveryMeet established acceptance criteria
    Reliability After Fatigue & Premature DetachmentMeet established acceptance criteria
    Tensile Strength of Coil & Detachment ZoneMeet established acceptance criteria
    Tensile Strength of Delivery WireMeet established acceptance criteria
    Delivery Wire FlexibilityMeet established acceptance criteria
    Detachment TimeMeet established acceptance criteria
    RadiopacityMeet established acceptance criteria
    Particulate Generation of Detachment ZoneMeet established acceptance criteria
    Aging and Shelf Life (3 year)Meet established acceptance criteria
    Fiber Pull OutMeet established acceptance criteria
    Fiber EnduranceMeet established acceptance criteria
    Packaging validation after 3-year accelerated agingMeet established acceptance criteria
    Animal TestingCoil Per Aneurysm (Avg./S.D.)Comparable to the Predicate Device
    Detachment Time (seconds)Comparable to the Predicate Device
    Coil Detachment Reliability (detachment attempts/coil)The Sapphire demonstrated higher reliability compared to the Predicate Device
    Post Embolization Angiographic AssessmentComparable to the Predicate Device
    Angiographic Assessment One-Month Follow-UpComparable to the Predicate Device
    Angiographic Assessment Three-Month Follow-UpComparable to the Predicate Device
    Physician Device EvaluationEase of Coil PreparationComparable to acceptance criteria
    Ability to move coil within CatheterComparable to acceptance criteria
    AccessAcceptance Criteria
    Ability to Place catheter tip in desired locationComparable to acceptance criteria
    Trackability, Friction of coil through catheterComparable to acceptance criteria
    Coil VisualizationAcceptance Criteria
    Fluoroscopic visibility of CoilComparable to acceptance criteria
    Fluoroscopic visibility of Detachment MarkersComparable to acceptance criteria
    Coil Delivery ProcedureAcceptance Criteria
    Ability to Position Coil in SacComparable to acceptance criteria
    Detachment TimeComparable to acceptance criteria
    Ability to Pack Aneurysm SacComparable to acceptance criteria
    Ability to RepositionComparable to acceptance criteria
    Ease of use of the SDS Detachment SystemComparable to acceptance criteria
    Coil Positional Stability/Aneurysm OcclusionComparable to acceptance criteria
    Overall (Ease of Use)Comparable to acceptance criteria
    Overall (Overall performance)Comparable to acceptance criteria
    MRI CompatibilityMRI CompatibilityCompatible with 1.5 Tesla

    2. Sample size used for the Test Set and Data Provenance

    The provided document does not explicitly state the sample size for the test set used in the animal testing or physician device evaluation. Similarly, the data provenance (e.g., country of origin, retrospective or prospective nature) for these tests is not specified. The bench testing would likely be performed in a laboratory setting, not involving human or animal subjects in the same way.

    3. Number of Experts and their Qualifications for Ground Truth

    The document does not specify the number of experts used or their qualifications for establishing ground truth, particularly for the animal testing or physician device evaluation. However, the "Physician Device Evaluation" implies the involvement of medical professionals in assessing the device's practical performance.

    4. Adjudication Method

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1, none) for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study is mentioned in the provided text. The study focuses on comparing the Sapphire Detachable Fibered Coil System against a predicate device through bench, animal, and physician evaluations, rather than assessing human reader improvement with or without AI assistance.

    6. Standalone (Algorithm Only) Performance

    This device is a physical medical device (embolization coil system), not an AI-powered diagnostic or therapeutic algorithm. Therefore, "standalone (algorithm only without human-in-the-loop performance)" is not applicable to this submission.

    7. Type of Ground Truth Used

    Based on the descriptions:

    • Bench Testing: The ground truth for bench testing is inherent in established engineering and physicochemical standards and specifications for material properties, mechanical performance, and device functionality.
    • Animal Testing: The ground truth for animal testing would be based on physiological and anatomical observations, angiographic assessments, and potentially histopathological evaluations in the animal models. The comparison is made against the performance of the predicate device in a living system.
    • Physician Device Evaluation: The ground truth for physician evaluations is based on the subjective and objective assessments of experienced medical professionals regarding the device's ease of use, handling, visualization, and overall performance in a simulated or real clinical environment (though the nature of the evaluation is not fully detailed to determine if it's purely simulated or involves actual clinical use).

    8. Sample Size for the Training Set

    Since this is a physical medical device and not an AI/ML algorithm requiring a training set, the concept of a "training set" and its sample size is not applicable.

    9. How the Ground Truth for the Training Set was Established

    As mentioned above, this section is not applicable to the Sapphire Detachable Fibered Coil System as it is not an AI/ML device.