510(k) Data Aggregation

    K Number
    K031753

    Validate with FDA (Live)

    Date Cleared
    2004-10-28

    (511 days)

    Product Code
    Regulation Number
    886.5928
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AQuify Multipurpose Solution is indicated for cleaning, rinsing, chemical (not heat) disinfecting and storing soft (hydrophilic)lenses (including silicone hydrogel lenses) as recommended by your eye care practitioner. AQuify Multipurpose Solution is also indicated for daily protein removal when used with the no rub directions for use.

    Device Description

    AQuify Multi-Purpose Solution is a sterile aqueous solution containing dexpanthenol, preserved with polyhexanide 0.0001%. pluronic F127, edetate disodium dehydrate, sorbitol, tromethamine, sodium phosphate dihydrogren, and purified water.

    AI/ML Overview

    The provided text describes the safety and effectiveness studies for AQuify Multipurpose Solution, comparing it to predicate devices, particularly ReNu Multipurpose Solution. Here's an analysis of the acceptance criteria and study details based on the input:

    Acceptance Criteria and Device Performance

    The document does not explicitly state quantitative "acceptance criteria" in a table format with pass/fail thresholds. Instead, it describes studies and their findings, concluding that AQuify Multipurpose Solution is "substantially equivalent" to the predicate device. The performance is reported in terms of statistical significance and preference compared to the control.

    Here's a table summarizing the findings as they relate to implied acceptance:

    Acceptance Criteria (Implied)Reported Device Performance (AQuify MPS vs. ReNu Multiplus)
    Cleaning Capacity (Daily protein removal)AQuify MPS is substantially equivalent in terms of daily protein removal to previously submitted and reviewed predicate devices (K021635).
    Dehydration Effect / Water Holding CapabilitiesStatistically better than control lens multipurpose solution at increasing the steady-state water content within a lens group after 60 minutes at 50% relative humidity. AQuify had a higher water content.
    Microbiological Efficacy (Disinfection)Meets the stand-alone criteria of the FDA May 1, 1997 Guidance for Industry; Premarket Notification 510(k) Guidance Document for Contact Lens Care Products. Meets USP Modified criteria for Preservative Effective Testing and USP Sterility Test requirements (studies reviewed under K021635).
    ToxicologyToxicology studies completed, submitted, and reviewed under K021635. (Implied acceptable if approved).
    Clinical Safety (Symptoms)No significant differences in symptoms between AQuify MPS and the control solution.
    Clinical Efficacy (Overall Comfort - 2 weeks)Statistically significant differences in favor of AQuify MPS for overall comfort (1-10 scores), overall dryness, and overall lens awareness 1-10 scores. Statistically significant for lens awareness preference. Strong trends for overall comfort and lens awareness.
    Clinical Efficacy (Overall Comfort - 1 month)Statistically significant differences in favor of AQuify MPS for overall comfort and dryness (1-10 scores) and for overall comfort, dryness, and lens awareness preferences.
    Clinical Efficacy (Overall Comfort - 3 months)Statistically significant differences in favor of AQuify MPS for insertion and overall comfort (1-10 scores) and for insertion comfort, overall comfort, dryness, and lens awareness preferences. Overall comfort 1-10 scores and preferences were consistently statistically significantly better with AQuify MPS. AQuify MPS was also statistically significantly preferred for less dryness. AQuify MPS increasingly preferred, reaching almost a 4:1 ratio (42% vs 11%) by the final visit.
    Clinical Efficacy (Lens Cleanliness - Deposits/Wetting)No statistically significant differences in investigator appraisal of lens deposits, front surface wetting, and objective dark-field image analysis of lenses worn for 90 days. Both test and control provided clinically acceptable lens cleanliness.
    Overall Substantial Equivalence to Predicate DeviceAQuify Multipurpose Solution is substantially equivalent to ReNu Multiplus Multipurpose Solution for cleaning, rinsing, chemical disinfecting, and storing soft contact lenses.

    Study Details

    The primary study mentioned for clinical evaluation is a "three-month prospective, randomized, single masked trial."

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not explicitly stated as a number. The text mentions "over 50% of the subjects expressing a preference" by the final visit, suggesting a sizable cohort was involved, but the exact N is not provided.
      • Data Provenance: Prospective. The study was a "three-month prospective, randomized, single masked trial." Country of origin is not specified but implicitly in the USA as it's a 510(k) submission to the FDA.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document describes a clinical trial where subjects reported symptoms and efficacy scores, and an "investigator appraisal" for lens deposits and wetting.
      • Number of Experts: Not specified.
      • Qualifications of Experts: The "investigator" who performed appraisals would likely be an eye care professional (e.g., optometrist or ophthalmologist), but specific qualifications (e.g., years of experience) are not provided. The subjects themselves provided "ground truth" for subjective measures like comfort and dryness.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Adjudication methods like 2+1 or 3+1 are typically for reconciling discrepancies in expert interpretations of images or findings. This study primarily relies on direct subject reporting of subjective measures and investigator appraisal for objective measures.
      • Adjudication Method: Not applicable/Not mentioned. The study design ("single masked trial," "contra-lateral basis") means subjects compared the products on their own eyes, minimizing the need for adjudication of differing subjective experiences. Objective appraisals (e.g., dark field image analysis) would have predefined criteria.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • MRMC Study: No, this was not an MRMC study. This study evaluated the performance of a contact lens solution, not an AI-assisted diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Standalone Performance: Not applicable. This device is a contact lens solution, not an algorithm. Its performance is evaluated through direct use (cleaning, disinfection, clinical comfort/dryness) or laboratory testing (microbiology, water content).
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For Clinical Efficacy/Safety: Subjective self-reported data from study participants (e.g., comfort scores, dryness scores, preferences) and objective investigator appraisal (lens deposits, front surface wetting, dark-field image analysis).
      • For Microbiology/Cleaning/Dehydration: Laboratory standard tests, likely with pre-defined pass/fail criteria from regulatory guidance (e.g., FDA guidance, USP standards).
    7. The sample size for the training set:

      • Not applicable. This is a medical device (contact lens solution), not a machine learning model, so there is no "training set."
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for this type of device.