510(k) Data Aggregation

    K Number
    K031365

    Validate with FDA (Live)

    Date Cleared
    2003-07-18

    (79 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Prima III TotalSENSE Torso Coil is a receive-only phased array RF coil, used for obtaining diagnostic images of the abdominal and pelvic anatomy, including the liver, spleen, and renal arteries in Magnetic Resonance Imaging Systems. The Prima III TotalSENSE Torso Coil is designed for use with the Intera 3.0T MRI system manufactured by Philips Medical Systems, Inc. The indications for use are the same as for standard MR Imaging.

    The Prima III TotalSENSE Torso Coil is designed to provide Magnetic Resonance Images of the abdominal and pelvic anatomies. The Prima III TotalSENSE Torso Coil is designed for use with the Philips Intera 3.0T MRI scanner manufactured by Philips Medical Systems, Inc.

    The 3.0T MRI system is indicated for use as an NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR signal, (2) depend upon NMR parameters (proton density, spin lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.

    Device Description

    The Prima III TotalSENSE Torso Coil is a six channel receive only phased array coil. The coil is composed of a flexible top and a rigid bottom piece which allow for imaging of patients of different sizes. The open, patient friendly design minimizes claustrophobic effects and maximizes patient comfort. The coil elements and accessory electronics are enclosed in flexible and rigid plastic housings, which are fire rated.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for a medical device, the Prima III TotalSENSE Torso Coil for Magnetic Resonance Imaging (MRI). It is a declaration of "substantial equivalence" to predicate devices, rather than a study demonstrating meeting specific acceptance criteria in the way a clinical trial for a novel drug or AI diagnostic would.

    Therefore, many of the typical acceptance criteria and study design elements for an AI or diagnostic device are not applicable here. This document pertains to a hardware accessory for an existing medical imaging modality.

    However, I will extract what can be inferred about the "acceptance criteria" for a 510(k) submission, which in this context means demonstrating that the new device is as safe and effective as a legally marketed predicate device.


    Description of Acceptance Criteria and Device Performance

    The "acceptance criteria" for a 510(k) submission like this primarily revolve around demonstrating substantial equivalence to a predicate device in terms of intended use, indications for use, technological characteristics, and safety/effectiveness. There are no explicit quantitative performance metrics like sensitivity/specificity/accuracy with specific thresholds in this type of submission for an MRI coil.

    The "study" that proves the device meets these "acceptance criteria" is the 510(k) submission itself, where the manufacturer provides a comparison of their new device to existing predicate devices.

    Here's the information structured as requested, with "N/A" where the requested information is not relevant to this type of regulatory submission (an MRI coil, not an AI diagnostic):

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Demonstration of Substantial Equivalence)Reported Device Performance / Comparison
    Intended Use: Imaging of abdominal and pelvic anatomy, including liver, spleen, and renal arteries.Similar to: Vision 5000 Torso Coil (K013594) and Insight Plus 9000 Phased Array Torso and Pelvis Coil (K001209).
    Indications for Use: Identical to routine MRI imaging (diagnose based on distribution of protons, NMR parameters T1/T2, soft tissue structure).Similar to: Vision 5000 Torso Coil (K013594) and Insight Plus 9000 Phased Array Torso and Pelvis Coil (K001209). It is for use with the Philips Intera 3.0T MRI system, and the indications are the same as for that standard MRI system.
    Technological Characteristics:
    * Coil Enclosure Material: Flame Retardant Polyurethane, Vinyl Coated EVA foam, Flame Retardant Polycarbonate.Similar to: Mark 5000 Quadrature Shoulder Coil (K013854).
    * Coil Design: Receive-only phased array coil.Similar to: Insight Plus 9000 Phased Array Torso and Pelvis Coil (K001209).
    * Decoupling: Switching diode decoupling.Similar to: Insight Plus 9000 Phased Array Torso and Pelvis Coil (K001209).
    * Prevention of RF Burns: Does not transmit RF power, decoupling isolates coil elements, elements/circuitry enclosed in non-conductive housing.Similar to: Insight Plus 9000 Phased Array Torso and Pelvis Coil (K001209).
    * Radio Frequency Absorption: Receive-only coil, does not transmit RF power.Similar to: Insight Plus 9000 Phased Array Torso and Pelvis Coil (K001209).
    * Formation of Resonant Loop: Decoupling isolates coil elements, cable length and stiffness prevents looping.Similar to: Insight Plus 9000 Phased Array Torso and Pelvis Coil (K001209).
    Safety and Effectiveness: No new questions of safety or effectiveness regarding the device or its intended use are raised compared to the predicate devices. (Implicit acceptance criterion for 510k).The FDA's substantial equivalence determination implies that the device is considered as safe and effective as the predicate devices and does not raise new issues of safety or effectiveness. The comparison table directly addresses the technological characteristics related to safety (RF burns, RF absorption, resonant loops).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • N/A. This is a hardware device (MRI coil) submission, not an AI or diagnostic algorithm requiring a test set of patient data for performance evaluation. The "test" is the comparison to predicate devices and the technical specifications provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • N/A. Ground truth for patient data is not applicable to an MRI coil submission. The "ground truth" for the submission itself is the existing regulatory framework and the characteristics of predicate devices, assessed by the manufacturer and then reviewed by the FDA.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • N/A. Not applicable, as there is no test set of patient data requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A. Not applicable. No MRMC or AI component described.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • N/A. Not applicable. This is a hardware component.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • N/A. For this type of submission, the "ground truth" for demonstrating substantial equivalence relies on:
      • Technical specifications and engineering data proving the coil's physical properties, safety features, and performance within the MRI system.
      • Regulatory precedent (predicate devices cleared by the FDA).
      • Established principles of MRI physics and safety.

    8. The sample size for the training set

    • N/A. Not applicable. There is no AI component or training set.

    9. How the ground truth for the training set was established

    • N/A. Not applicable. There is no AI component or training set.