510(k) Data Aggregation

    K Number
    K030463

    Validate with FDA (Live)

    Date Cleared
    2003-02-26

    (14 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Maestro System™ endosseous implants may be used in the mandible and maxilla for use as an artificial root structure for single tooth replacement or as abutments for fixed bridgework and dental retention.

    Device Description

    The 7mm and 10.5mm length Maestro System dental implants are a machined titanium. screw-form implant supplied in diameters of 3.5mm, 4mm and 5mm. Implant raw material is titanium alloy as specified in ASTM F136 - Specification for Wrought Titanium 6AL-4V ELI Alloy for Surqical Implant Applications.

    The device is further processed by treating the surface with resorbable blast media (RBM) or coating with synthetic hydroxylapatite (HA). The product is packaged using materials known in the industry to be appropriate for medical device packaging and is provided with a minimum sterility assurance level of 108, validated in compliance to ANSI/AAMI/ISO 11137. Sterilization of healthcare products - Requirements for validation and routine control - Radiation sterilization.

    The Maestro System ™ is a comprehensive system containing implants, surgical components, and prosthetic components. The implants are specifically designed to optimize strain distribution to contiguous bone under functional loading in order to promote strain-induced bone growth and interface maintenance over the long term. This is achieved by optimizing implant designs based on bone quality.

    AI/ML Overview

    This document describes a Special 510(k) Notification for a device modification, specifically the addition of new sizes (7mm and 10.5mm lengths) to the existing BioHorizons Maestro System™ screw-type dental implants. This type of submission focuses on demonstrating substantial equivalence to a predicate device and typically does not involve new clinical studies or detailed performance acceptance criteria, as the fundamental technology and indications for use remain unchanged. Therefore, a direct response to many of the requested categories (acceptance criteria, specific study details, sample sizes, expert involvement, ground truth, etc.) is not applicable or cannot be extracted from the provided text.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    This information is not provided in the document. Special 510(k) applications for minor modifications like adding new sizes generally rely on demonstrating that the new sizes have the same technological characteristics and intended use as the predicate device, implying that the established performance of the predicate device applies. There is no new performance data or acceptance criteria presented for the modified device sizes.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable/provided. The submission is a Special 510(k) for a device modification, not a submission based on new clinical testing that would involve a test set. The focus is on demonstrating equivalence of the new sizes to the existing, predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/provided. As explained above, no new clinical test set requiring expert ground truth establishment is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided. No test set or associated adjudication methods are described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable/provided. This is a submission for a dental implant, not an AI-powered diagnostic device. Therefore, MRMC studies or AI assistance are not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable/provided. This is a submission for a dental implant, not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not applicable/provided. No new ground truth establishment is described as the submission focuses on demonstrating equivalence of new sizes to an already approved predicate device. The "ground truth" for this modification essentially relies on the established safety and effectiveness of the existing Maestro System implants.

    8. The sample size for the training set

    This information is not applicable/provided. No training set is described as this is not an AI/algorithm submission, nor does it include a de novo clinical study requiring a training set.

    9. How the ground truth for the training set was established

    This information is not applicable/provided. As above, no training set is described.


    Summary of the K030463 Application:

    The BioHorizons Implant Systems, Inc. K030463 submission is a Special 510(k) Notification for a Device Modification. The modification involves adding new lengths (7mm and 10.5mm) to the existing Maestro System™ screw-type dental implants.

    The core argument for acceptance is Substantial Equivalence to the predicate device (The Maestro System™ already marketed by BioHorizons, with various K-numbers listed).

    The justification for substantial equivalence is based on the following:

    • No change in fundamental scientific technology: The implants are still machined titanium, screw-form.
    • Same materials: Titanium alloy (ASTM F136 - 6AL-4V ELI) and surface treatments (RBM or HA) remain unchanged.
    • Same suppliers, processing, packaging, and sterilization methods.
    • No change in intended use or indications for use: The implants are for single tooth replacement or as abutments for fixed bridgework and dental retention in the mandible and maxilla.

    Because this is a submission for a modification to an existing, approved device, and not a de novo device or a significant change requiring new clinical data, the acceptance criteria are implicitly that the new sizes do not introduce new questions of safety or effectiveness and maintain the same technological characteristics as the predicate. The "study" proving this is the comparative analysis of technological characteristics and intended use against the predicate device, which is documented in the 510(k) submission itself. No new clinical trials or performance studies with specific acceptance criteria are mentioned or required for this type of submission.