Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K030109

    Validate with FDA (Live)

    Date Cleared
    2003-08-08

    (207 days)

    Product Code
    Regulation Number
    870.1220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Reprocessed Electrophysiology (EP) Catheters are indicated for temporary intracardiac sensing, recording, stimulation and electrophysiological mapping of cardiac structures.

    Reprocessed Electrophysiology (EP) Catheter Cables are indicated for use with the appropriate electrode catheter during electrophysiology studies.

    Device Description

    Diagnostic electrophysiology (EP) catheters are specially designed electrode catheters that transmit electrical impulses and can be positioned for endocardial recording or stimulation. Diagnostic EP catheters incorporate a handpiece, a flexible shaft and a distal tip section containing diagnostic electrodes. The distal tips of deflectable catheters can be deflected into a curve by manipulating the handpiece; fixed curve catheters have an established distal tip shape.

    AI/ML Overview

    The provided text is a 510(k) summary for the Alliance Medical Corporation's Reprocessed Electrophysiology (EP) Catheters. It describes the device, its intended use, and the testing conducted to demonstrate its safety and effectiveness.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state quantitative acceptance criteria or detailed performance metrics. Instead, it broadly claims that the reprocessed devices "perform as originally intended" and are "safe, effective and substantially equivalent to the predicate devices."

    Acceptance Criteria CategoryReported Device Performance (Summary)
    BiocompatibilityDemonstrated through testing
    Validation of ReprocessingDemonstrated through testing
    Sterilization ValidationDemonstrated through testing
    Function Test(s)Tested prior to packaging and labeling; deemed to perform "as originally intended"
    Overall Safety & EffectivenessSafe, effective, and substantially equivalent to predicate devices

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size for the test set used in the performance studies (biocompatibility, reprocessing validation, sterilization validation, function tests).

    The data provenance is implied to be prospective bench and laboratory testing conducted by Alliance Medical Corporation. There is no mention of country of origin for the data in this summary.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    The document does not provide information on the number or qualifications of experts used to establish ground truth for the test set. The nature of the tests (biocompatibility, sterilization, function) suggests objective laboratory measurements rather than expert human interpretation in a clinical setting.

    4. Adjudication Method for the Test Set

    Not applicable. The tests mentioned (biocompatibility, reprocessing, sterilization, function) are objective laboratory or bench tests, not requiring an adjudication method among experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a 510(k) submission for a reprocessed medical device (catheter), not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant or mentioned.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical device, not an algorithm. The testing focuses on the physical and functional properties of the reprocessed catheter.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance tests would be based on:

    • Established standards and protocols: for biocompatibility and sterilization.
    • Predetermined functional specifications: for the function tests, likely defined by the original manufacturer's specifications for new devices.
    • Validated reprocessing procedures: for reprocessing validation.

    8. The Sample Size for the Training Set

    Not applicable. This is a medical device approval, not a machine learning model. There is no concept of a "training set" in this context.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a machine learning model.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1