Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K023905

    Validate with FDA (Live)

    Date Cleared
    2003-04-30

    (159 days)

    Product Code
    Regulation Number
    876.5320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Athena Pelvic Muscle Trainer is intended to provide electrical stimulation and neuromuscular reeducation for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women.

    Device Description

    The Athena Pelvic Muscle Trainer is a vaginal two electrode stimulation probe operated through a FM remote control device. The probe is stored in an activator box that turns the probe on automatically when the probe is removed from the box.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for the Athena Pelvic Muscle Trainer. It focuses on demonstrating substantial equivalence to existing predicate devices, rather than presenting a standalone study with acceptance criteria and detailed performance metrics of the device itself. Therefore, many of the requested details about a study and its results cannot be fully extracted.

    Here's an analysis based on the provided text, addressing the points where information is available and indicating where it is not:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly define "acceptance criteria" in the context of a performance study for the Athena Pelvic Muscle Trainer. Instead, it compares the device's technical specifications to those of legally marketed predicate devices to establish substantial equivalence.

    FeatureAthena Pelvic Muscle Trainer (Reported Performance - as described in 510(k))Predicate Devices (Reference for Substantial Equivalence)
    Intended UseTreatment of urinary incontinenceTreatment of urinary incontinence (for all listed predicates)
    Programmable FeaturesIntensityIntensity, Session length (Pathway CTS 2000); Channel, Intensity, Continuous-Intermittent (Liberty PFS-200); Intensity (InCare Microgyn II)
    Channel SwitchYesYes (Pathway CTS 2000, Liberty PFS-200); No (InCare Microgyn II)
    Intensity Adj.YesYes (for all listed predicates)
    Duty Cycle2.5 sec stim / 2.5 sec rest2 sec stim/1 sec rest (Pathway CTS 2000); 5 sec stim/ 10 rest (Liberty PFS-200); 2 sec stim /4sec rest (InCare Microgyn II)
    Duration15 minutes0-30 minutes (Pathway CTS 2000); 30 minutes (Liberty PFS-200); 15 minutes (InCare Microgyn II)
    Power Source(Probe) 1 3.6 volt lithium battery (Remote) 3 AAA alkaline batteriesUnknown (Pathway CTS 2000); (2) 3 volt lithium batteries (Liberty PFS-200); 9V alkaline (InCare Microgyn II)
    Treatment Protocol15 min/twice daySet by physician (Pathway CTS 2000); 15 min/twice day (Liberty PFS-200, InCare Microgyn II)
    Mode of OperationIntermittentIntermittent (Pathway CTS 2000, Liberty PFS-200, InCare Microgyn II); Continuous (Liberty PFS-200)
    Ingress ProtectionIP67 – Protected from Dust and liquid ingressUnknown (Pathway CTS 2000); IPXO-no protection (Liberty PFS-200); Unknown (InCare Microgyn II)
    Output Current0 – 70 mA0-100 mA (Pathway CTS 2000); 0-65 mA (Liberty PFS-200); Unknown (InCare Microgyn II)
    Pulse Width$200 \mu S$$300 \mu s$ (Pathway CTS 2000); $300 \mu s$ (Liberty PFS-200); $50-300 \mu s$ (InCare Microgyn II)
    Output TypeConstant voltage over range of 100 to 1000 OhmsConstant current over unknown range (Pathway CTS 2000); Constant current over range of 100 to 1000 Ohms (Liberty PFS-200); Unknown (InCare Microgyn II)
    Output Isolation$>10^8$ OhmsUnknown (Pathway CTS 2000); $>10^8$ Ohms (Liberty PFS-200); $>10^{12}$ (InCare Microgyn II)
    Number of Electrodes2 stimulatorsUnknown (Pathway CTS 2000); 2 stimulators (Liberty PFS-200); 4 stimulators (InCare Microgyn II)
    Usage ConditionsReusable-single patient useReusable-single patient use (for all listed predicates)
    Electrode OrientationCircularCircular (for all listed predicates)
    Probe Length3.3 in. nominal2.9 in nominal (Pathway CTS 2000); 2.8 in. nominal (Liberty PFS-200); 4.5 in. nominal (InCare Microgyn II)
    Probe Diameter.99 in. nominal.70 in. nominal (Pathway CTS 2000); .95 in. nominal (Liberty PFS-200); 1 in. nominal (InCare Microgyn II)
    Electrode materialStainless steelUnknown (Pathway CTS 2000); Stainless steel (Liberty PFS-200); Metal (InCare Microgyn II)
    Device ConnectionRF RemoteAttached cord (for all listed predicates)
    Electrode PlacementVaginalVaginal (for all listed predicates)
    Waveform TypeSquareSquare (for all listed predicates)
    Waveform ShapeAsymmetricalAsymmetrical (Pathway CTS 2000); Symmetrical (Liberty PFS-200, InCare Microgyn II)
    Frequency12.5Hz/50 Hz12.5Hz/50/100/200Hz (Pathway CTS 2000); 12.5Hz/50 Hz (Liberty PFS-200); 10-400 Hz (InCare Microgyn II)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not describe a clinical performance study with a test set. This 510(k) is based on demonstrating substantial equivalence through comparison of technical specifications and intended use to predicate devices. No patient data or clinical study results for the Athena Pelvic Muscle Trainer are provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no clinical performance study involving a test set and ground truth establishment is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical performance study involving a test set is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an electrical stimulator, not an AI-powered diagnostic tool, and no MRMC study is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is an electrical stimulator, not an algorithm, and its performance is assessed against predicate devices' technical specifications, not by an algorithm's standalone performance.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    Not applicable, as no clinical performance study with defined ground truth is described. The "ground truth" in this context is the established safety and effectiveness of the predicate devices based on their prior clearance, to which the new device claims substantial equivalence.

    8. The sample size for the training set

    Not applicable, as no AI model or training set is mentioned for this device.

    9. How the ground truth for the training set was established

    Not applicable, as no AI model or training set is mentioned for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1