Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K023424

    Validate with FDA (Live)

    Device Name
    IGI-SYSTEM
    Date Cleared
    2003-05-09

    (210 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IGI Image-Guided Implantation system, is a computerized navigational system intended to provide assistance in both the planning (preoperative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides accurate navigational guidance of surgical instruments, with regard to the pre-operative planning in dental implantation procedure.

    Device Description

    The IGI-System is an electro-optical device, which was specifically designed to greatly improve the surgical implantation procedure by providing the surgeon an accurate guidance to the location of the surgical tools prior to, and more importantly, during the surgical operation. The novel device both increases surgical success rate, and reduces potential damage to adjacent anatomical structures and tissues. The IGI-System introduces, for the first time, the element of objective, state-of-the-art measurements to the field of dental implantation. Equipped with the IGI-System, the dental surgeon is able to plan the implantation procedure and align the actual placement of the implant in the pre-planned position with great accuracy. The IGI-System allows the user to view the reconstructed 2D images of the patient anatomy together with overlaid graphic information depicting the position of dental surgical instrument and pre-planned position of dental implant. The IGI-System utilises an infra-red (IR) optical tracking system, comprised of emitters, camera and a tracking data processor (computer), to determine the location and orientation of the dental hand-piece and the patient. The position and orientation of both the dental surgical instrument and the patient are transmitted to an image data processor (computer) which makes necessary calculations to provide the guidance graphic overlay depicting the dental surgical instrument and planned dental implant location on the medical image. Alignment of the patient and medical images is accomplished through a process termed fiducial registration. The goal is to indicate to the dentist, based on the preimplantation medical images, the exact position of the tracked dental surgical instrument together with the location of planned dental implant location, with regard to the patient's anatomy. The dentist will place a virtual implant image in the optimal location by use of the IGI software and available CT data. Once an implant has been optimally positioned, the system will correlate between the dentist's plan and the actual performance. In case of significant variance between the plan and the performance the system will alert the user. The system is used in two stages of the implantation procedure. The first stage is preplanning of the surgical implantation procedure. The second stage is the clinical one in which the system is used for accurate guidance of dental surgical instruments according to the pre-planned implant position. In addition, the system assists the dentist in avoiding critical anatomical structures. The system is used to pre-plan the surgical implantation procedure. Consequently, the system provides accurate guidance during the dental implantation, of surgical instruments with regard to pre-operative planning, and assists the dentist in avoiding the risk of causing damage to critical anatomical structures. The IGI-System is defined as a supporting device, providing a significant contribution to the decision-making process, which continuously takes place during the implantation process. It is by no means intended to replace human judgement. The final decisions as to the exact location, timing, intensity and depth of the drilling are the sole responsibility of the surgeon. The surgeon is at liberty, at any time throughout the surgical procedure, to determine the final drilling position, or to modify implants' planned position. Thus, under no circumstances does the device relieve the surgeon of his or her ultimate clinical responsibility.

    AI/ML Overview

    The provided text describes the IGI-System™, an image-guided implantation system for dental surgery, and its substantial equivalence to predicate devices. It discusses performance standards and the acceptance criteria for accuracy.

    Here's an analysis of the acceptance criteria and study information, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Navigational Accuracy: +/- 1mmNavigational Accuracy: +/- 0.5mm (Better than criterion)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a sample size for a test set or data provenance (e.g., country of origin, retrospective/prospective). The claim of +/- 0.5mm accuracy is made in the context of comparing the IGI-System to predicate devices, stating that "All three devices provide navigational accuracy of +/-0.5mm." This suggests the accuracy was likely established through internal testing or by referencing the accuracy of the predicate devices, rather than a specific clinical study described here.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document states, "The level of accuracy, defined to be acceptable in dental implantation navigation, was +/-1mm. This level of accuracy was defined based on review of state of the art technology, opinions by experts in the field..."
    However, the number and specific qualifications of these experts are not provided. It's unclear if these experts were involved in establishing ground truth for a test set, or merely in defining the acceptable accuracy threshold.

    4. Adjudication Method for the Test Set

    The document does not mention an adjudication method for a test set. This implies that there was no formal human adjudication process for evaluating the device's performance against a ground truth in a clinical or simulated setting described here.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or any effect size related to human reader improvement with AI assistance. The device is described as a "supporting device" that provides "accurate guidance" and "significant contribution to the decision-making process," but "is by no means intended to replace human judgement." This suggests its role is assistive rather than a direct replacement or an AI that significantly alters human reader performance in a diagnostic context.

    6. Standalone Performance Study

    The document implies a standalone performance due to the statement: "All three devices provide navigational accuracy of +/-0.5mm. This level of accuracy cannot be achieved by human visual inspection." This highlights that the device's accuracy is a standalone measurement of its navigational capability, independent of human input, though it is used with a human surgeon. However, a dedicated standalone study is not detailed. The accuracy is presented as an inherent characteristic of the device and its predicate technologies.

    7. Type of Ground Truth Used

    The type of ground truth used to establish the +/-0.5mm accuracy is not explicitly stated. Given it's a navigational system based on CT scans, the ground truth for accuracy would likely involve precise physical measurements against a known spatial reference, potentially using phantoms or highly accurate instruments to verify the device's ability to track and guide to precise locations defined by the CT data. The "optimal location" of an implant is planned using CT data and IGI software, and the system correlates this plan with "actual performance."

    8. Sample Size for the Training Set

    The document does not provide any information regarding a training set sample size. This is common for devices of this nature, which are typically based on established principles of optical tracking and image processing rather than machine learning models that require large training datasets in the conventional sense.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no mention of a training set, there is no information on how its ground truth was established. The device's functionality relies on algorithms for image reconstruction, instrument tracking, and registration, which are likely developed and validated through engineering principles and calibration rather than a machine learning training paradigm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1