510(k) Data Aggregation

    K Number
    K023030

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-06-18

    (280 days)

    Product Code
    Regulation Number
    878.5000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DemeTECH Polyester Nonabsorbable Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular and ophthalmic and neurological procedures.

    Device Description

    The DemeTECH Polyester Nonabsorbable Suture is a braided green, individually packaged nonabsorbable Polyester surgical suture available in a range of gauge sizes and lengths, and attached to needles of varying types and sizes.

    AI/ML Overview

    The DemeTECH Polyester Nonabsorbable Suture is a medical device and thus does not involve AI/ML performance.

    1. Table of Acceptance Criteria and Reported Device Performance:
    Acceptance Criteria (Set by USP)Reported Device Performance
    Meets United States Pharmacopeia (USP) 23 requirementsDevice demonstrated to meet USP 23 requirements
    Meets United States Pharmacopeia (USP) 24 requirementsDevice demonstrated to meet USP 24 requirements
    Equivalent to legally marketed predicate deviceDevice demonstrated to be equivalent to predicate device
    1. Sample size used for the test set and the data provenance: Not applicable. The study involved testing the physical properties of sutures against USP standards and comparing them to a predicate device, not analyzing patient data.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth was established by adherence to published United States Pharmacopeia (USP) standards, which are defined by scientific testing methodologies rather than expert consensus on individual cases.

    3. Adjudication method: Not applicable. The assessment was based on physical property testing against USP standards.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a medical device for surgical procedures, not an AI/ML diagnostic tool.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a surgical suture, not an algorithm.

    6. The type of ground truth used: The ground truth used was based on established United States Pharmacopeia (USP) 23 and 24 requirements for Nonabsorbable Surgical Sutures. This involves standardized physical and chemical testing methods to ensure quality and performance. Additionally, equivalence to a legally marketed predicate device (Grams Polyester Nonabsorbable Suture) served as a benchmark.

    7. The sample size for the training set: Not applicable. This is a medical device, not an AI/ML model that requires a training set. The study involves quality control testing for manufacturing standards.

    8. How the ground truth for the training set was established: Not applicable.