    Intended Use

    The over the counter Injex 30 Needle-Free Insulin Injector System is intended for the subcutaneous iniection of U-100 insulin. The Injex 30 Insulin Ampule is labeled as suitable for administration of 5 to 30 insulin units with a scale in 1 unit increments.

    Device Description

    Equidyne Systems Injex 30 Needle Free Insulin Injector System is a means of administering insulin in the subcutaneous tissue without the use of needles. The needle free injector utilizes a high velocity focused jet of liquid to penetrate the skin and deposit the insulin in the subcutaneous tissue. The process takes place in a fraction of a second.
    The Injex 30 Needle-Free Insulin Injector System consists of four main components:

    1. Iniex 30 Needle-Free Insulin Injector A small re-usable hand held Iniector.
    2. Reset Box - A compact re-usable Reset Box used to reset the Injex 30 Needle Free Insulin Injector,
    3. Injex 30 Insulin Ampule - A disposable, sterile 30 unit Injex 30 Insulin Ampule that functions like a needle syringe to draw the medication from the medicine bottle or vial using a hand draw with a plunger. Once the Injex 30 Insulin Ampule is loaded into the Injex 30 Needle Free Insulin Injector, the dose is injected into the individual by releasing the Safety and pressing the trigger.
    4. Vial Adapter - A disposable, sterile Vial Adapter facilitates the transfer of the insulin from the medicine bottle or vial to the Injex 30 Insulin Ampule.
      The Injex 30 Needle Free Insulin Injector can deliver variable doses of insulin from 5 to 30 units. The Injex 30 Insulin Ampule scale is in 1 unit increments.
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Injex 30 Needle-Free Insulin Injector System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
    Ease of Use/Instruction Adequacy1. Subjects find it easy/somewhat easy to self-administer an injection after viewing training materials. 2. Subjects are able to self-administer an injection after viewing training materials (i.e., not difficult or unable).1. 100% of subjects said it was either easy or somewhat easy to give themselves an injection after viewing the training video and reading the Instructions for Use. 2. None of the subjects said that it was difficult or they were not able to give themselves an injection after viewing the training video and reading the Instructions for Use.
    User PreferenceA significant portion of subjects prefer the Injex device or are ambivalent (would use either Injex or needle/syringe) compared to a traditional needle and syringe.84% of the subjects said that they either prefer the Injex 30 Needle-Free Insulin Injector or would use either the Injex 30 Needle-Free Insulin Injector or a needle and syringe in the future.
    Pain PerceptionPain from the Injex device is perceived as similar to, or less than, a traditional insulin needle and syringe by a significant portion of subjects.91% of the subjects said the pain from the Injex 30 Needle-Free Insulin Injector was either the same as, a little less or a lot less than the insulin needle and syringe.
    Comparative Ease of UseSubjects find the Injex device just as easy to use as an insulin needle and syringe.91% of the subjects felt that the Injex 30 Needle-Free Insulin Injector was just as easy to use as an insulin needle and syringe.
    Comparative Ease of Self-InjectionSubjects find it just as easy to give themselves an injection using the Injex device as with an insulin needle and syringe.100% of the subjects felt it was just as easy to give themselves an injection using the Injex 30 Needle-Free Insulin Injector as it was with an insulin needle and syringe.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not explicitly state the exact number of subjects in the clinical study. It mentions a "good distribution of age among the subjects with the average age being 48.5 years with a range of 25 to 77 years old" and provides percentages of education levels. This suggests a reasonably sized cohort, but the specific n is missing.
    • Data Provenance: The document does not specify the country of origin. It is a prospective study as subjects were observed administering injections and then surveyed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Not applicable. This study is based on subjective patient experience and self-reported perceptions (ease of use, pain, preference) when comparing the device to a traditional needle and syringe. There was no "ground truth" to be established by experts in the context of diagnostic accuracy or clinical correctness of the injection, as the purpose was to assess user experience and instruction adequacy.

    4. Adjudication Method for the Test Set

    • Not applicable. Since the study focuses on subjective user experience and self-reported outcomes, there was no need for an adjudication method by experts. The results are based on direct subject feedback via questionnaires.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. This was not an MRMC study. It was a clinical study evaluating user experience, ease of use, and pain perception for a medical device (insulin injector) compared to a predicate device (needle and syringe), not a diagnostic imaging study involving multiple readers and cases. Therefore, no effect size of human readers improving with or without AI assistance is reported.

    6. Standalone Performance Study

    • Yes, in spirit, but within a user context. The study assessed the Injex 30 Needle-Free Insulin Injector System alone (after subjects viewed training materials) for ease of use and ability to self-administer, comparing these aspects to a traditional needle and syringe. While not a "standalone algorithm" performance, it evaluated the device's functional performance in the hands of users without external assistance during the injection process itself, beyond the provided instructions.

    7. Type of Ground Truth Used

    • The "ground truth" for this study was subjective patient feedback and self-reported experiences regarding ease of use, pain, and preference, gathered through questionnaires after direct interaction with both the investigational device and the predicate.

    8. Sample Size for the Training Set

    • Not applicable. This device is a physical medical device (insulin injector), not a software algorithm requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As above, there is no training set for a physical medical device.