Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K021623

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-09-10

    (117 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. Spondylolisthesis Spinal Fixation Device System, 87MNH the intended use and indications for use are:
      The HAIDER-UCR Spinal System is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
    2. Pedicle Screw Spinal System, (Class II uses) 87MNI the intended use and indications for use are:
      The HAIDER-UCR Spinal System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine:
    • degenerative spondylolisthesis with objective evidence of neurological impairment,
    • fracture,
    • dislocation,
    • scoliosis,
    • kyphosis,
    • spinal tumor, and
    • failed previous fusion pseudarthrosis)
    Device Description

    The new 5.5mm screw will be used in all UCR Spinal System applications where a smaller screw would be of benefit. Therefore, the mechanical properties and fatigue performance are important. The remainder of the UCR Spinal System components will remain unchanged. Other than the major and minor diameter of the cancellous portion of the shaft and threads, the new 5.5mm screws are similar to the existing screws. Spinal rod attachment method are the same as cleared previously.

    The new 8.0mm Spinal Screws expand the options for screw insertion by providing an alternative to the smaller 6, 6.5 or 7.0mm pedicle screws of the available UCR Spinal System. This new screw diameter allows the surgeon to insert screws in cases where the spine deformity or the large size of the pedicle makes use of smaller diameter screws less desirable. The larger diameter of the 8.0mm Spinal Screws permits the surgeon to address the needs of an expanded patient population.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for the UCR Spinal System, specifically focusing on new 5.5mm and 8.0mm spinal screws. This document addresses the substantial equivalence of these new components to previously cleared devices.

    Based on the provided information, the device is a pedicle screw Spinal System, which falls under Class II uses. The "studies" conducted are non-clinical bench tests focused on mechanical and fatigue performance to ensure the new screws meet established standards and are equivalent to the predicate device.

    Here's an analysis of the requested information:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Standard/Guidance)Reported Device Performance
    ASTM F-1717-96 guidance documentBoth the 5.5mm and 8.0mm pedicular screws carry clinically useful loads out to the five-million cycle asymptotic endurance limit.
    Compliance with material standardsNot explicitly detailed, but implied by "Titanium alloy" and "Identical" material comparison to predicate.
    Compliance with mechanical testing standard"Mechanical and fatigue testing were carried out" in accordance with ASTM F-1717-96.
    Compliance with biocompatibility standardNot explicitly detailed but is a general control for such devices.
    Compliance with specified labeling requirementsNot explicitly detailed but is a general control for such devices.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: The document does not specify the exact number of samples tested for the mechanical and fatigue testing. It only states "Samples were tested".
    • Data Provenance: The data provenance is from non-clinical bench testing. The testing was conducted by SEASPINE, INC. The country of origin of the data would be the USA, where SEASPINE, INC. is located. It is prospective in the sense that the testing was performed specifically for this 510(k) submission, not gathered retrospectively from patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is a non-clinical mechanical testing study, not a clinical study involving diagnosis or interpretation by medical experts. Therefore:

    • Number of experts: Not applicable.
    • Qualifications of experts: Not applicable.

    The "ground truth" here is the performance under defined mechanical engineering standards, which is established by the test methods themselves and interpreted by engineers.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Since this is a non-clinical mechanical test, an adjudication method for human interpretation is not applicable. The results are quantifiable mechanical measurements.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No, an MRMC study was not done. This submission is for a medical device (spinal screws), not an AI/software device that assists human readers.
    • Effect Size: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: No, a standalone performance study in the context of an algorithm or AI was not done. This is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used for this premarket notification is based on engineering standards and mechanical performance metrics. Specifically, the ability of the screws to "carry clinically useful loads out to the five-million cycle asymptotic endurance limit" as defined by the ASTM F-1717-96 guidance document.

    8. The sample size for the training set

    This is a non-clinical mechanical testing study for a physical device. There is no "training set" in the context of machine learning or AI. The tests are designed to validate the physical properties against established standards.

    9. How the ground truth for the training set was established

    As there is no training set, this question is not applicable. The "ground truth" for the device's performance is established by the specified mechanical testing methods and standards (e.g., ASTM F-1717-96).