Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K021502

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-06-06

    (28 days)

    Product Code
    Regulation Number
    888.4580
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Orthosonics OSCAR OE3000 is intended to assist in the removal of polymethylmethacrylate (PMMA) bone cement during arthroplasty revision.

    Device Description

    The currently marketed Orthosonics OSCAR Ultrasonic Cement Removal System consists of a power module which generates the ultrasonic energy and provides overall control of the device, a handpiece, and a set of cement removal probes. The primary change to the system is the addition of a laparoscopic video system, which can aid the clinician in examining the inside of the bone that is having its glue removed. This part of the system would only be used when the ultrasound-powered probes are not being used. The laparoscope and optical coupler are currently marketed medical devices. In order to make room for the video module in the main system unit of the OE3000, the ultrasonic power modules were reduced in physical size. In addition, the currently marketed device now has an optional accessory that was added to the system via a letter-to-file, without filing a new 510(k). The change involves adding a "slap hammer" that can be attached to an "extraction" probe, to aid in removal of a PMMA cement plug that is NOT firmly bonded to the endosteal interface. The Extraction Probe to be used with the slap hammer is a modified version of the piercer probe that was cleared with the original 510(k).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Orthosonics OSCAR OE3000, a device intended to assist in the removal of polymethylmethacrylate (PMMA) bone cement during arthroplasty revision.

    It explicitly states that the device was tested against safety and EMC standards. However, the document does not contain information regarding acceptance criteria for device performance (e.g., efficacy in removing cement, accuracy of the laparoscopic system, or strength of the slap hammer) nor does it describe a study to prove such performance criteria were met.

    Instead, the submission focuses on demonstrating "substantial equivalence" to predicate devices, primarily through technological characteristics and safety testing. The statement "The technological characteristics of the OSCAR OE3000 are exactly the same as those of the predicate devices" is central to its claim of substantial equivalence.

    Therefore, many of the requested sections regarding performance studies, sample sizes, ground truth, and expert adjudication cannot be extracted from this document.

    Here's a summary of what can be extracted:


    Acceptance Criteria and Device Performance Study for Orthosonics OSCAR OE3000

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Safety: Device meets general requirements for safety (EN60601-1, EN60601-1-1)."The OSCAR OE3000 was tested to the requirements of... EN60601-1, EN60601-1-1."
    Electrical Safety: Device meets electrical safety requirements (EN60601-1-2)."The OSCAR OE3000 was tested to the requirements of... EN60601-1-2."
    Electromagnetic Compatibility (EMC): Device meets EMC requirements (EN60601-1-2)."The OSCAR OE3000 was tested to the requirements of... EN60601-1-2."
    Substantial Equivalence: Device has the same indications for use and target population as legally marketed predicate devices, and the same technological characteristics."The modified OE3000 is a medical device, and it has the same indications for use and the same target population as the legally marketed predicate devices. The modified OE3000 has the same technological characteristics as the predicate devices."

    2. Sample Size Used for the Test Set and Data Provenance:

    • Not Applicable. The document describes testing to safety and EMC standards, and a demonstration of substantial equivalence based on technological characteristics and intended use. It does not describe a clinical performance test set with a specific sample size of subjects or data, nor does it specify data provenance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • Not Applicable. As no clinical performance test set is described, there's no mention of experts establishing ground truth for such a set.

    4. Adjudication Method for the Test Set:

    • Not Applicable. No clinical performance test set is described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No. The document does not describe any MRMC comparative effectiveness study where human readers' performance with and without AI assistance was evaluated. The device is a surgical tool, not an AI diagnostic aid.

    6. Standalone (Algorithm Only) Performance Study:

    • No. The device is a physical surgical instrument system (ultrasonic cement removal, a laparoscopic video system, and a slap hammer accessory). It is not an algorithm that performs in a standalone capacity.

    7. Type of Ground Truth Used:

    • For Safety and EMC: The ground truth for safety and EMC testing would be the "pass/fail" criteria defined by the respective EN standards (EN60601-1, EN60601-1-1, EN60601-1-2).
    • For Substantial Equivalence: The ground truth for demonstrating substantial equivalence was established by comparing the device's technological characteristics, indications for use, and target population against legally marketed predicate devices (K961725, K914084, K903458, K961101) as reviewed by the FDA.

    8. Sample Size for the Training Set:

    • Not Applicable. This is a hardware device submission, not an AI/machine learning algorithm that requires a training set. The term "training set" is not relevant in this context.

    9. How the Ground Truth for the Training Set Was Established:

    • Not Applicable. As no training set is described (see point 8), there is no ground truth establishment method for it mentioned.