510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ENDOPATH and PROXIMATE Linear Cutters and Staplers are intended for use in multiple open or minimally invasive surgical procedures for the transection and resection of tissues. Linear cutters are also intended for creation of anastomoses in these procedures. Specific indications are provided in specific instrument product insert labeling.

    The ENDOPATH ETS Endoscopic Linear Cutter and ETS FLEX Endoscopic Articulating Linear Cutter have application in general, gynecologic, and thoracic surgery for transection, resection, and/or creation of anastomoses.

    The ENDOPATH ETS45 Endoscopic Linear Cutters, the ETS Flex45 Endoscopic Articulating Linear Cutters , the ETS Compact Flex45 Articulating Linear Cutters are intended for transection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic (including radical prostatectomy), thoracic, and pediatric surgical procedures. They can be used with staple line or tissue buttressing materials such as bovine pericardium.

    The ENDOPATH ETS Flex45 No-Knife Articulating Linear Staplers and the ETS Compact Flex45 No-Knife Articulating Linear Staplers are intended for transection, and /or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic (including radical prostatectomy), thoracic, and pediatric surgical procedures. They can be used with staple line or tissue buttressing materials such as bovine pericardium.

    The ENDOPATH EZ45 Endoscopic Linear Cutter has application in multiple open and other minimally invasive surgical procedures for transection, resection of anastomoses, and can be used with staple line or tissue buttressing materials, such as bovine pericardium.

    The ENDOPATH EZ45 No-Knife Endoscopic Linear Stapler has application in multiple open or other minimally invasive surgical procedures for transection, and can be used with staple line or tissue buttressing materials, such as bovine pericardium.

    The ENDOPATH EZ45 Thoracic Endoscopic Linear Cutter has application in thoracic and other minimally invasive surgical procedures for transection, resection of anastomoses, and can be used with staple line or tissue buttressing materials, such as bovine pericardium.

    The PROXIMATE ACCESS 55 Articulating Linear Staplers have application in gastrointestinal, gynecologic, thoracic, and pediatric surgery for transection and resection of tissues.

    The PROXIMATE Linear and Vascular Linear Cutters with Safety Lockout have application in gastrointestinal, gynecologic, thoracic, and pediatric surgery for transection, and/or creation of anastomoses and can be used with staple line or tissue buttressing materials, such as bovine pericardium.

    The PROXIMATE Linear Four Row Stapler with Safety Lockout have application in gastrointestinal and thoracic surgery for transection and can be used with staple buttressing materials, such as bovine pericardium.

    The PROXIMATE Reloadable Linear Stapler has application throughout the alimentary tract and in thoracic surgery for transection and resection of internal tissues.

    The PROXIMATE Reloadable Vascular Linear Stapler has application for use on internal tissue that can easily be compressed to 1.0 mm in thickness. The staplers may also be used to ligate pulmonary vessels.

    Device Description

    The ENDOPATH and PROXIMATE Linear Cutter models are sterile single use instruments that deliver staples while simultaneously dividing tissue between rows. The ENDOPATH No-Knife Stapler and PROXIMATE Stapler models are sterile single use instruments that deliver staples, but do not cut. Depending upon the particular model, they deliver 2 to 3 staggered rows of staples, 2 double-staggered rows, or 2 triple-staggered rows of staples. Staple line lengths vary between 35 mm and 100 mm. These instruments may be used in either open or endoscopic procedures, depending upon the design. Some instruments are reloadable and, if so, they may be reloaded with various reloads (i.e. vascular/thin, standard, thick) depending on the thickness of tissue that is to be transected or resected. Vascular models use specific vascular reload cartridges. The instruments may be reloaded a number of times according to product insert instructions for a maximum number of firings per instrument. Some instruments have articulating heads and flex features. The length of the shaft may vary as well. A safety lock-out feature on linear cutter instruments prevents a used reload from being fired again. The common element among all these instruments is that they are all mechanical surgical stapling devices.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for "ENDOPATH™ and PROXIMATE™ Linear Cutters and Staplers". Here's an analysis based on your requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Implied Criteria: Demonstrated safety and effectiveness for intended use, substantial equivalence to predicate devices.Demonstrated Performance: "Animal testing demonstrated satisfactory performance of the ENDOPATH and PROXIMATE Linear Cutters and Staplers to support the revised product insert labeling. The performance data and information in this submission demonstrate that these devices, when used in accordance with directions for use with respect to tissue thickness, can be used in procedures where pulmonary vessels are encountered."
    New Indication: Use in procedures where pulmonary vessels are encountered (implies a specific performance standard for vessel transection/ligation).Adequately demonstrated for this use.

    Important Note: The document is a 510(k) summary, which typically focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and specific numerical performance data against those criteria. The acceptance criteria for a 510(k) are generally implicit: the device must be as safe and effective as a legally marketed predicate device. The performance data described is a summary to support this claim, particularly for a revised indication (pulmonary vessels).

    Beyond the provided text, specific numerical acceptance criteria (e.g., staple line integrity, burst pressure, or tissue thickness range) would typically be found in the full 510(k) submission, internal design verification and validation documents, or industry standards applied to surgical staplers.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated. The document only mentions "Animal testing."
    • Data Provenance: Animal testing. The country of origin is not specified, but the submission is to the U.S. FDA. The testing would be prospective in nature as it's designed to evaluate the device's performance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This information is not provided in the document. For animal testing, the "ground truth" would likely be established by the observed physiological outcomes, successful stapling, and lack of adverse events, assessed by veterinary or surgical experts involved in the study.

    4. Adjudication Method for the Test Set

    • This information is not provided in the document. In animal studies, adjudication would typically involve veterinary pathologists or surgeons evaluating the results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. This type of study is more common for diagnostic imaging AI devices, rather than a mechanical surgical stapler. The evaluation here is based on physical performance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • No, this is not applicable. The device is a mechanical surgical stapler, not an algorithm. Its performance is inherent to its design and how it's used by a human surgeon.

    7. Type of Ground Truth Used

    • The ground truth used was based on physiological outcomes and direct observation of device performance during animal testing (e.g., successful transection, anastomoses creation, hemostasis, and absence of leaks or defects in the stapled tissue). The specific details of what was considered "satisfactory performance" are not elaborated upon.

    8. Sample Size for the Training Set

    • This concept is not applicable to a traditional mechanical surgical device like a stapler. There is no "training set" in the context of machine learning. The device's design and manufacturing processes are refined through engineering, material science, and prior testing, not through a data-driven training set in the AI sense.

    9. How the Ground Truth for the Training Set Was Established

    • This concept is not applicable for the reasons stated above.