510(k) Data Aggregation

    K Number
    K020769

    Validate with FDA (Live)

    Date Cleared
    2002-04-03

    (27 days)

    Product Code
    Regulation Number
    862.3200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Single Analyte Urine Drugs of Abuse (DAU) Calibrators are intended for in vitro diagnostic use for the calibration of their respective DAU enzyme immunoassays to detect phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.

    The Single Analyte Urine Drugs of Abuse (DAU) Controls are intended for in vitro diagnostic use for the validation of their respective DAU enzyme immunoassays to detect phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.

    Device Description

    All of the Single Analyte Urine DAU Calibrators and Controls are human urine-based liquid, and ready to use. These Calibrators and Controls do not have any especially unique technical characteristics. When applicable, they contain known concentration of a specific drug analyte.

    The Negative DAU Calibrator is a processed, drug-free human urine matrix. The Low, Cutoff, Intermediate, and High Calibrators, as well as the 2 levels of Controls are prepared by spiking known concentrations of drug analyte into the Negative DAU Calibrator matrix. The various concentrations of each drug analyte in their corresponding calibrators and controls are summarized as follows:

    Phencyclidine EIAOpiate EIACocaine Metabolite EIAAmphetamines EIA
    Reference MaterialPhencyclidineMorphineBenzoylecgonined-Methamphetamine
    Low Calibrator12.5 ng/mL150 ng/mL150 ng/mL500 ng/mL
    Cutoff Calibrator25 ng/mL300 ng/mL300 ng/mL1000 ng/mL
    Intermediate Calibrator50 ng/mL600 ng/mL1000 ng/mL1500 ng/mL
    High Calibrator100 ng/mL1000 ng/mL3000 ng/mL2000 ng/mL
    Control Level 118 ng/mL225 ng/mL225 ng/mL750 ng/mL
    Control Level 232 ng/mL375 ng/mL375 ng/mL1250 ng/mL

    The nominal concentrations of the analyte in the calibrators and controls are determined and confirmed by GC/MS.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for "Single Analyte Urine Drugs of Abuse Calibrators and Controls". This device is a set of calibrators and controls used for in vitro diagnostic tests, rather than an AI-powered diagnostic device. As such, many of the requested categories in the prompt (e.g., AI performance, expert adjudication, MRMC studies, training set details) are not applicable.

    Here's an analysis based on the information available:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state "acceptance criteria" for the LZI device. Instead, it demonstrates substantial equivalence to a predicate device (DRI's Drugs of Abuse Urine Calibrators and Controls) by highlighting similarities in intended use, matrix, and performance characteristics. The performance characteristics mentioned as similar are precision, accuracy, and stability. However, specific numerical targets or benchmarks for these characteristics for the LZI device are not provided in this summary.

    The table below summarizes the device's characteristics and how they compare to the predicate, which implicitly defines the performance expectations for equivalence.

    CharacteristicPredicate Device (DRI's DAU Urine Cals/Controls)LZI's Single Analyte Urine DAU Cals/ControlsAcceptance Criteria / Performance
    Intended UseCalibration and validation of DAU immunoassays for phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.Calibration and validation of DAU immunoassays for phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.Similar (Intended use is the same)
    MatrixUrine-based liquidsHuman urine-based liquidsSimilar (Urine-based)
    Cutoff ConcentrationSame as SAMHSA recommendationsSame as SAMHSA recommendationsSimilar (Matches SAMHSA guidelines)
    Control Levels2 levels, ± 25% of cutoff per SAMHSA guideline2 levels, concentrations (18/32 ng/mL for PCP, 225/375 ng/mL for Opiate, 225/375 ng/mL for Cocaine, 750/1250 ng/mL for Amphetamines) set around cutoffSimilar (2 levels, concentrations set around cutoff per SAMHSA guideline implicitly met)
    Analyte Concentration ConfirmationConfirmed by GC/MSConfirmed by GC/MSSimilar (Uses GC/MS for nominal concentration confirmation)
    Storage Condition2℃ to 8℃2℃ to 8℃Similar (Same storage conditions)
    Performance Characteristics (General)Precision, accuracy, stability are similar to LZI's productPrecision, accuracy, stability are similar to DRI's productSimilar (General statement of similar precision, accuracy, and stability. Specific numerical data not provided in the summary.)

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not provide details on sample size or data provenance for any "test set" in the context of device performance testing. The evaluation focused on demonstrating substantial equivalence to a predicate device based on characteristics and general performance attributes, rather than presenting a detailed clinical study with a test set. The nominal concentrations of analytes in the calibrators and controls are "determined and confirmed by GC/MS," which implies internal analytical testing, but specifics are not given.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    Not applicable. This device is a set of calibrators and controls for laboratory assays, not an AI or diagnostic device that requires expert ground truthing of clinical data for performance evaluation. The "ground truth" for the calibrators and controls is the precisely known concentration of drug analytes, which is confirmed by analytical methods like GC/MS.

    4. Adjudication Method:

    Not applicable for the reasons stated above.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is not an AI-powered device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an AI-powered device with an algorithm for standalone performance. The device itself (calibrators and controls) is a component used in an assay.

    7. The Type of Ground Truth Used:

    The ground truth for the device's characteristics is the precisely known (nominal) concentrations of drug analytes spiked into the urine matrix, which are confirmed by Gas Chromatography/Mass Spectrometry (GC/MS).

    8. The Sample Size for the Training Set:

    Not applicable. This is not an AI-powered device that requires a "training set." The development of calibrators and controls relies on precise chemical formulations and analytical validation.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable. As there is no training set for this type of device, this question is not relevant. The "ground truth" for the product itself (known analyte concentrations) is established through analytical chemistry methods like GC/MS.