510(k) Data Aggregation

    K Number
    K020703

    Validate with FDA (Live)

    Date Cleared
    2002-11-13

    (254 days)

    Product Code
    Regulation Number
    888.3560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Scorpio® CR and PS Femoral Components with Titanium Plasma Spray Coating and the Scorpio Tibial Tray with Titanium Plasma Spray Coating are intended for single-use, and are intended to be marketed for cemented fixation only. These femoral components and tibial tray are intended to be used with Osteonics tibial inserts and patellar components in total knee arthroplasty.

    Indications:

    • Painful, disabling joint disease of the knee resulting from: degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis.
    • Post-traumatic loss of knee joint configuration and function.
    • Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
    • Revision of previous unsuccessful knee replacement or other procedure.

    Additional Indications for Posterior Stabilized Components:

    • Ligamentous instability requiring implant bearing surface geometries with increased constraint.
    • Absent or non-functioning posterior cruciate ligament.
    Device Description

    The Scorpio CR and PS Femoral Components with Titanium Plasma Spray Coating, and the Scorpio® Tibial Tray with Titanium Plasma Spray Coating, are substantially equivalent to the predicate Cobalt Chromium alloy (CoCr) MicroStructured® Scorpio® CR (#K974556) and PS (#K962152) femoral components, and the MicroStructured® standard tibial trays of the Osteonics® Series 7000 Total Knee System (#K910990) in all material and design aspects. The following reflect the new characteristics of the subject components:

    • The interior surfaces of the subject Scorpio® CR and PS femoral components with Titanium Plasma Spray Coating receive a titanium plasma spray coating instead of the MicroStructured® porous coating. All other design features of the femoral components are identical to the predicate devices.
    • For the Scorpio Tibial Tray with Titanium Plasma Spray Coating, the inferior surface of the tibial tray receives a titanium plasma spray coating (as opposed to a MicroStructured® porous coating), and the keel of the tray receives a thin titanium plasma spray coating. The plate of the subject tibial tray is 0.75mm thicker than that of the predicate tibial tray, and has been filled in with cobalt chromium alloy. Also, the screw hole configuration on the titanium plasma sprayed trays has been modified to allow greater angulation of the screws.
    AI/ML Overview

    This document is a 510(k) summary for the Scorpio® CR and PS Femoral Components and Tibial Tray with Titanium Plasma Spray Coating. It does not describe a study involving an AI/ML device, but rather a medical device (knee implants). Therefore, many of the requested categories for AI/ML studies are not applicable.

    Here's a breakdown of the relevant information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) submission is for demonstrating substantial equivalence to predicate devices, not for meeting specific performance acceptance criteria for a novel device. The "acceptance criteria" here implicitly refer to the FDA's requirements for demonstrating substantial equivalence.

    Acceptance Criteria (Implicit for Substantial Equivalence)Reported Device Performance (as claimed by manufacturer)
    Material Equivalence: The new device's materials are comparable to predicate devices.The plasma-sprayed commercially pure ("CP") Titanium coating is identical to the coating used on predicate Howmedica Osteonics® femoral stems. The combination of a CoCr substrate and a plasma sprayed CP Titanium coating on a knee component is substantially equivalent to the CoCr substrate/Titanium alloy coating metal combination employed by the legally marketed predicate MCK® Total Knee System manufactured by Biomet, Inc.
    Design Equivalence: The new device's design is comparable to predicate devices.The interior surfaces of the subject Scorpio® CR and PS femoral components with Titanium Plasma Spray Coating receive a titanium plasma spray coating instead of the MicroStructured® porous coating. All other design features of the femoral components are identical to the predicate devices. For the Scorpio® Tibial Tray with Titanium Plasma Spray Coating, the inferior surface receives a titanium plasma spray coating, the keel receives a thin titanium plasma spray coating. The plate of the subject tibial tray is 0.75mm thicker than that of the predicate tibial tray, and has been filled in with cobalt chromium alloy. The screw hole configuration was modified. These changes maintain substantial equivalence in design given the overall context of knee implants.
    Intended Use Equivalence: The new device has the same intended use as predicate devices.The intended use for the new Scorpio® components is for cemented fixation only, with Osteonics tibial inserts and patellar components in total knee arthroplasty, which is consistent with the predicate devices. Specific indications and contraindications are provided, matching established uses for similar knee implants.
    Safety and Effectiveness Equivalence: The new device is as safe and effective as predicate devices."Testing was provided to support the claim of substantial equivalence." (Specific test results are not detailed in this summary, but would have been part of the full 510(k) submission). The FDA concurrence letter confirms their determination of substantial equivalence.

    2. Sample Size Used for the Test Set and the Data Provenance

    This document describes a 510(k) submission for a medical device (knee implants), not an AI/ML study. Therefore, the concepts of "test set" and "data provenance" in the context of AI/ML are not applicable here. The "testing" referred to would typically involve mechanical, material, and potentially biological compatibility testing, not data analysis on a clinical dataset in the AI/ML sense.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. This is not an AI/ML study involving "ground truth" establishment from expert consensus. The "ground truth" for a medical device like a knee implant is established through rigorous engineering standards, material science, and clinical performance over time, not expert labeling of data.

    4. Adjudication Method (2+1, 3+1, none) for the Test Set

    Not applicable. This is not an AI/ML study.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device submission, not an AI/ML comparative effectiveness study.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    Not applicable in the AI/ML sense. For this type of device, "ground truth" would be related to:

    • Material properties: Established through laboratory testing (e.g., tensile strength, fatigue life, biocompatibility standards).
    • Mechanical performance: Established through in-vitro biomechanical testing (e.g., wear testing, fixation strength, range of motion).
    • Clinical outcomes (for predicate devices): Long-term clinical data on the safety and effectiveness of the predicate devices.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML study.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. This is not an AI/ML study.


    Summary of the Document's Purpose:

    The provided document is a 510(k) summary for new knee implant components. The primary goal of a 510(k) submission is to demonstrate that a new medical device is substantially equivalent to a legally marketed predicate device. This means the device has the same intended use and either has the same technological characteristics as the predicate device or has different technological characteristics that do not raise different questions of safety and effectiveness and demonstrate that the device is as safe and effective as the legally marketed device. The document explicitly states that the new components are "substantially equivalent" to existing predicate devices in terms of materials, design (with minor, justified modifications), and intended use.