510(k) Data Aggregation

    K Number
    K012960

    Validate with FDA (Live)

    Date Cleared
    2002-03-08

    (185 days)

    Product Code
    Regulation Number
    864.1860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BioGenex Mouse Monoclonal Anti-Progesterone Receptor Antibody (Clone PR88) is an immunohistochemical (IHC) assay and is intended for laboratory use to qualitatively identify by light microscopy human progesterone receptor in normal and/or pathological paraffin-embedded, formalin-fixed tissues. The PR88 antibody specifically binds to antigens located in the nucleus of cell populations that express progesterone receptor in normal and abnormal tissues. This antibody is indicated as an aid in assessing patient response to hormonal therapy and as an aid in the prognosis and management of breast cancer patients. The clinical interpretation of any staining or its absence should be complemented by morphological studies using proper controls and should be evaluated within the context of the patient's clinical history and other diagnostic tests by a qualified pathologist.

    Device Description

    BioGenex PR88 is a monoclonal antibody, which specifically binds to progesterone Dreceptor antigen located in the nuclear region of a variety of normal and abnormal tissues. It is a mouse monoclonal anti-progesterone receptor antibody from mouse ascites fluid diluted in phosphate buffered saline pH 7.6 containing bovine serum albumin as carrier and on and 0.09% sodium azide as preservative. The antibody is available in concentrated proximand 0.5970 bounds ready to use form (AM328-5M and AM328-10M). Refer to package insert for details.

    AI/ML Overview

    This document appears to be a 510(k) summary for a medical device, the BioGenex Mouse Monoclonal Anti-Progesterone Receptor Antibody (Clone PR88). It does not contain explicit "acceptance criteria" in the format typically used for performance studies with predefined thresholds for success (e.g., a specific sensitivity or specificity target to be met). Instead, it describes a "performance data" section to demonstrate substantial equivalence to a predicate device and a reference method.

    Therefore, the response below will present the reported device performance and infer the "acceptance criteria" based on the described concordance and the overall conclusion of substantial equivalence.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Inferred)Reported Device Performance
    Substantial equivalence to predicate device (Ventana K990618) and reference DCC techniqueVerified by comparison of features and performance study results.
    Specificity with normal human tissuesPR88 antibody demonstrated negative immunoreactivity with most normal human tissues. Positive immunoreactivity observed in specific normal tissues (Langerhans islet cells of pancreas, endometrium of uterus, stromal cells of cervix, parenchymal cells of ovary, convoluted and collecting tubes of kidney, epithelial cells of colon, adenohypophysis of pituitary).
    Intra-run (within-run) reproducibilityNo significant variation among slides of the same tissue stained in a single run.
    Inter-run (run-to-run) reproducibilityNo significant variation among slides of the same tissue stained in different runs.
    Agreement between instrumental vs. manual runsNo significant variation between slides stained manually and by the automated process.
    Concordance with reference DCC assay for PROverall binary concordance of 74% (99/134) with a 2-sided 95% confidence interval of 66% - 81% (p<0.0001). This level was considered "similar."
    StabilityStable for at least 24 months at 2-8°C. Stable after continuous 48-hour exposure at 45°C.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Specificity Test Set: 88 formalin-fixed and paraffin-embedded tissues covering a wide range of normal human tissue types. The document does not specify the country of origin or whether it was retrospective or prospective.
    • Reproducibility Test Set:
      • Intra-run: 10 slides of the same tissues within a single run.
      • Inter-run: Slides containing the same tissues over 10 different runs.
      • Instrumental vs. manual: 10 slides for each tissue (selected for a wide range of reactivity) stained manually and using BioGenex Automated Stainer.
      • The document does not specify the country of origin or whether it was retrospective or prospective.
    • Sensitivity (DCC Comparison) Test Set: A total of 134 specimens.
      • First batch: 41 specimens assayed for DCC at the University of Texas Health Science Center at San Antonio, Texas.
      • Second batch: 93 specimens assayed for DCC at Genesee Hospital, Rochester, NY.
      • The document does not explicitly state if the data was retrospective or prospective, but the description of "specimens were selected according to ... 1) formalin-fixed and paraffin-embedded tissue sections were available; 2) each tissue was initially assayed for PR by DCC" suggests a retrospective collection of existing samples. The origin appears to be the USA.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • For the Sensitivity (DCC Comparison) Test Set: Each resulting slide was read independently by two pathologists.
    • Qualifications: The document states they were "pathologists," but does not provide additional details such as years of experience. They reportedly had "no knowledge of any other laboratory or clinical data of the specimens."

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • For the Sensitivity (DCC Comparison) Test Set: The document states "each resulting slide was read independently by two pathologists." It does not describe an adjudication method if there was disagreement between the two pathologists. The scoring was based on percentage of cells with positive nuclear staining, and a cut-off value of >10% of positive tumor cells was used to score a slide as positive or negative. The document reports a single concordance figure, implying agreement or a method not detailed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly conducted as described in the context of human readers improving with AI assistance. This device is an immunohistochemical antibody, not an AI software. The study compared the IHC staining (read by pathologists) against a biochemical assay (DCC).

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No, a standalone algorithm-only performance assessment was not performed, as this is not an AI device. The device itself is an antibody reagent, and its performance (IHC staining) is evaluated by human pathologists. The "specificity" and "reproducibility" sections effectively describe the standalone performance of the antibody and staining system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the Sensitivity (DCC Comparison) Test Set: The primary ground truth for comparison was the dextran-coated charcoal (DCC) technique, described as a biochemical assay and "considered the gold standard for PR assay" at the time. Pathologists' readings of IHC slides were then compared against this DCC ground truth.

    8. The sample size for the training set

    • This document does not describe a "training set" in the context of machine learning or AI. This is a traditional IVD device (antibody). The development and optimization of the antibody and staining protocols would have involved various internal tests, but these are not referred to as a "training set" in the provided text.

    9. How the ground truth for the training set was established

    • As there is no mention of a "training set" in the context of an AI/ML algorithm, this question is not applicable to the provided document. The ground truth for the performance evaluation (test set) was established by the DCC assay.