510(k) Data Aggregation

    K Number
    K012935

    Validate with FDA (Live)

    Date Cleared
    2001-10-11

    (41 days)

    Product Code
    Regulation Number
    884.5250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    As the Select IUI® (predicate device), TheCurve™ (new device) is a single use, sterile, disposable, flexible catheter designed to be used for intrauterine artificial insemination.

    Device Description

    TheCurve, such as the Select IUI, comprises a long clear polyethylene tube that is tightly surrounded on three quarters of its proximal length by an outer, clear polypropylene sheath. This leaves 4.5 cm of the inner polyethylene tube exposed at the distal end of the device. The proximal inner tube and outer sheath circumferences are molded directly onto the distal end of a luer lock hub so that only the lumen of the inner tube can provide a channel for flow. This design results in a single-channel catheter with two segments each having different degrees of flexibility. The extreme distal tip of TheCurve™ is open as compared to the extreme distal tip of the Select IUI® which is closed and has two side openings. Both TheCurve and Select IUI have six (6) coloured graduation markings placed at 5, 6, 7, 8, 9, and 10 cm distance from the extreme distal tip of the catheter. The distal portion of TheCurve is curved while the distal portion of the Select IUI is straight.

    AI/ML Overview

    The provided text describes a 510(k) submission for a medical device called "TheCurve™," an intrauterine insemination catheter. The submission aims to demonstrate substantial equivalence to a predicate device, the "Select IUI®."

    Here's an analysis based on your requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The documents do not explicitly state "acceptance criteria" in a quantitative sense as might be found in a performance study with defined thresholds (e.g., minimum sensitivity, specificity). Instead, the submission focuses on demonstrating substantial equivalence to a predicate device based on design, materials, function, and intended use. The "performance" is implicitly deemed acceptable if the device is found substantially equivalent to the already approved predicate.

    FeatureTheCurve™ (New Device) PerformanceSelect IUI® (Predicate Device) PerformanceEquivalence Justification
    Intended UseSingle use, sterile, disposable, flexible catheter designed for intrauterine artificial insemination.Single use, sterile, disposable, flexible catheter designed for intrauterine artificial insemination.Identical
    Constituent MaterialsIdentical to Select IUI®.Identical to TheCurve™.Identical
    Length (Effective)17.2 cm17 cmMinor difference (1 mm shorter), not cited as a concern for effectiveness or safety.
    Length (Overall)19.9 cm20 cmMinor difference (1 mm shorter), not cited as a concern for effectiveness or safety.
    ID (Inner Diameter)1.1 mm for effective length1.6 mm for effective lengthMinor difference (0.5 mm smaller). The implication on performance (e.g., flow rate) is not discussed in detail but is implicitly deemed acceptable given the substantial equivalence claim.
    OD (Outer Diameter)2.6 mm stepped down to 1.6 mm for distal 4.5 cm of length2.6 mm stepped down to 2.1 mm for distal 4.5 cm of lengthDifferences in distal OD are minor and related to the curvature of TheCurve™. Implicitly deemed acceptable.
    Graduation MarksIdentical to Select IUI® (six colored markings at 5, 6, 7, 8, 9, and 10 cm from the distal tip).Identical to TheCurve™.Identical
    DesignInner polyethylene tube within an outer polypropylene sheath. Distal portion is curved. Extreme distal tip is open.Inner polyethylene tube within an outer polypropylene sheath. Distal portion is straight. Extreme distal tip is closed with two side openings.Differences (curved distal tip, open vs. closed distal tip) are described as improvements to ease insertion. Implicitly deemed acceptable for maintaining or improving safety and effectiveness.
    FunctionIdentical to Select IUI® (channel for flow).Identical to TheCurve™.Identical

    2. Sample Size Used for the Test Set and Data Provenance

    The documents do not describe a formal "test set" or a study with human subjects (clinical trial) to evaluate the device's performance against specific acceptance criteria. This submission is a 510(k) for substantial equivalence based on device comparison and technical specifications, not a clinical performance study. Therefore, there's no sample size or data provenance in the context of a clinical test set.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. As no clinical study or performance evaluation with a "ground truth" was conducted or reported, no experts were involved in establishing such ground truth for a test set. The "truth" in this submission relies on the comparison of engineering specifications and intended use to a predicate device.

    4. Adjudication Method for the Test Set

    Not applicable. No test set requiring adjudication was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a manual medical instrument (an intrauterine insemination catheter), not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study related to AI performance is irrelevant.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a manual medical instrument and does not involve an algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    Not applicable in the conventional sense of a clinical study. The "ground truth" for this 510(k) submission is the pre-existing regulatory approval and established safety and effectiveness of the predicate device (Select IUI®). The new device, TheCurve™, is compared against the predicate's specifications and intended use.

    8. The Sample Size for the Training Set

    Not applicable. Since there's no algorithm involved, there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable. Since there's no algorithm involved, there is no "training set" or ground truth for it.

    In summary:

    This 510(k) submission relies on a "benchmarking" or "comparison" approach rather than a clinical performance study with a defined test set and acceptance criteria. The "acceptance criteria" are implicitly met if the FDA determines the new device is substantially equivalent to the predicate device in terms of safety and effectiveness, despite some minor physical differences that are argued to either have no impact or slightly improve usability (e.g., curved tip for easier insertion). The evidence provided is primarily a detailed technical comparison of the new device with the legally marketed predicate device.