Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    DEN230019

    Validate with FDA (Live)

    Date Cleared
    2023-08-18

    (144 days)

    Product Code
    Regulation Number
    876.5016
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AXIOS Stent and Electrocautery Enhanced Delivery System is indicated for use to facilitate transgastric or transquodenal endoscopic drainage of the gallbladder in patients with acute cholecystitis who are at high risk for surgery.

    Device Description

    The AXIOS Stent and Electrocautery-Enhanced Delivery system is designed to help facilitate the transgastric or transduodenal endoscopic drainage of the gallbladder. The AXIOS Stent remains identical to the stents cleared previously per K220112. The AXIOS Stent is a flexible, magnetic resonance conditional, fully covered, self-expanding, braided nitinol stent, which comes preloaded into the delivery system. The AXIOS stent is designed with two flanges, one on each end, to prevent migration and to enable tissue plane apposition and a "saddle" in between the flanges to span the tissue implant distance. The stent is fully covered with silicone to prevent leakage, discourage tissue in-growth within the nitinol woven braid, and facilitate removal. The AXIOS Electrocautery-Enhanced Delivery System consists of a catheter and an integrated handle with manual controls for positioning and deploying the AXIOS™ Stent. The AXIOS™ Electrocautery-Enhanced Delivery System is designed to be used in the gastrointestinal tract with commercially available echoendoscopes with a 3.7 mm diameter or larger working chamel and is compatible with commercially available 0.035-inch insulated endoscopic guidewires. The Electrocautery-Enhanced Delivery System connects with an off-the-shelf electrosurgical unit or generator that is compliant to IEC 60601-1-2 and IEC 60601-2-2.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves the AXIOS Stent and Electrocautery-Enhanced Delivery System meets those criteria. The information is primarily from non-clinical/bench testing and a pivotal clinical study, supplemented by a meta-analysis of clinical literature.

    Here's the breakdown of the requested information:


    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents two main categories of performance data: non-clinical/bench testing and clinical study results.

    Table 1: Non-Clinical/Bench Testing Performance for AXIOS Stent and Delivery System

    TestPurposeAcceptance CriteriaReported Device Performance (Result)
    Deployed Stent Saddle LengthVerify dimensions of deployed stentDeployed Stent Saddle Length (inside flange to inside flange) @37°C $\geq$ 8.5 mm $\leq$ 13 mm for both AXIOS 10 mm x 10 mm & AXIOS 15 mm x 10 mmPass
    Deployed Stent Saddle Outer DiameterVerify dimensions of deployed stentAXIOS 10mm x 10 mm stent: $\geq$ 8 mm; AXIOS 15mm x 10mm stent: $\geq$ 13 mmPass
    Deployed Stent Flange WidthVerify dimensions of deployed stentAXIOS 10 mm x 10 mm: $\geq$ 3 mm; AXIOS 15 mm x 10 mm: $\geq$ 3 mmPass
    Stent Pull-Out ForceVerify stent will remain in place when pull-out force is appliedAXIOS 10 mm x 10 mm: $\geq$ 2.54 N; AXIOS 15 mm x 10 mm stent: Not explicitly stated, implied same as 10mmx10mm via "Pass" result column.Pass
    Stent (Saddle) Radial Strength – Compression & ExpansionVerify expansion strength of the compressed stent meets product specificationIn compression @ 37°C $\geq$ 0.06 N/mm of length @ 50% Ø; In expansion @ 37°C $\geq$ 0.02 N/mm of length @ 50% Ø (for both AXIOS 10mm x 10 mm and 15 mm x 10 mm)Pass
    Deployment (Unsheathing) ForceVerify force needed to deploy stentAXIOS 10 mm x 10 mm: $\leq$ 40 N; AXIOS 15 mm x 10 mm stent: $\leq$ 40 NPass
    Implant Anchor Function – Retention (Tensile)Verify force needed to achieve apposition of distal flange to target structureAXIOS 10 mm x 10 mm: $\geq$ 2 N; AXIOS 15 mm x 10 mm stent $\geq$ 2 NPass
    Magnetic Resonance TestingVerify stent will not prevent ability to use MRIThe stent should not prevent the ability of the physician to use Magnetic Resonance Imaging on the patient.Pass
    Fatigue TestingEnsure stent remains intact throughout the period of useThe stent must remain intact after 940,000 translation (distal flange with respect to proximal flange) cycles.Pass
    Corrosion TestingEvaluate corrosion resistance of stentThe metal stent and stent covering must show corrosion resistance when subjected to an in vitro corrosive environment equivalent to sixty days.Pass
    Delivery System Working LengthVerify working length of delivery deviceWorking Length (distal point of the handle to distal end of the catheter tip) $\leq$ 138 cmPass
    Slider lock hold force testingEvaluate performance of lock under compressional and torsional forcesLock hold force $\geq$ 22 NPass
    Nose lock hold forceEvaluate performance of lock under compressional and torsional forcesLock hold force $\geq$ 22 NPass
    Nose/slide lock cyclingInspect function of nose/slide lock featureNose lock must function after 6 cycles; slide lock must function after 4 cycles.Pass
    Joint Strength TestingVerify joint strength meets design requirementsPer EN ISO 10555-1, the device shall meet specific minimum force.Pass
    Luer to Nose Joint StrengthVerify peak tensile force of luer to nose jointLuer to Nose Joint Strength $\geq$ 22 NPass
    Outer Sheath to Handle Torque StrengthVerify ability of AXIOS catheter to withstand torque when the device is rotated 360°The catheter system must withstand a minimum of 1 full turn.Pass
    Tracking ForceVerify tracking force of outer sheath through a 3.7 mm working channelTrack force $\leq$ 8.83 NPass

    Table 2: Clinical Study Performance (Pivotal Study and Meta-Analysis)

    Endpoint/OutcomeAcceptance Criteria (Performance Goal)Reported Device Performance (Pivotal Study, N=30 ITT)Reported Device Performance (Meta-Analysis, N=713 combined)
    Primary Effectiveness: Days to resolution of Acute Cholecystitis (AC)$< 3.5$ days (upper bound of 97.8% one-sided CI)Primary Analysis (ITT, censored): Mean 5.30 $\pm$ 14.90 days. 97.8% one-sided CI [0.00, 10.78]. Failed acceptance criteria on this specific analysis using censoring. Additional Analysis (ITT, all patients observed): Mean 1.60 $\pm$ 1.50 days. 97.8% one-sided CI [0.0, 2.15]. Met acceptance criteria. Treated/Per-Protocol (N=28): Mean 1.39 $\pm$ 0.92 days. 97.8% one-sided CI [0.0, 1.74]. Met acceptance criteria.Not applicable (this specific endpoint was not directly reported in the meta-analysis in the same way).
    Secondary Effectiveness: Rate of re-intervention (migration/occlusion)$< 46.2$% (upper bound of 99.7% one-sided CI)ITT Patients (N=30): 16.7% (5/30). 99.7% one-sided CI [0.0%, 42.0%]. Met acceptance criteria. Treated/Per-Protocol Patients (N=28): 14.3% (4/28). 99.7% one-sided CI [0.0%, 40.2%]. Met acceptance criteria.10.3% (2.6%, 22.1%) - Broader definition of re-interventions likely.
    Additional Endpoints: Technical success (stent placement with patency)Not explicitly stated (implied high percentage)ITT Patients (N=30): 93.3% (28/30). 97.9% CI [77.9%, 99.2%].97.0% (95.1%, 98.5%)
    Technical stent removal successNot explicitly stated (implied high percentage)Attempted (N=19): 100% (19/19). 95% CI [82.4%, 100.0%].Not directly reported in meta-analysis.
    Stent patency at stent removal visitNot explicitly stated (implied high percentage)Assessed (N=22): 100% (22/22). 95% CI [84.6%, 100.0%].Not directly reported in meta-analysis.
    Recurrence of acute cholecystitis symptomsNot explicitly stated (implied low percentage)ITT Patients (N=30): 10.0% (3/30). 95% CI [2.1%, 26.5%].1.8% (0.4%, 4.2%)
    Clinical SuccessNot explicitly stated (implied high percentage)Not directly defined as a primary or secondary endpoint with a specific numerical goal in the pivotal study. The "Symptom resolution" primary endpoint implies clinical success.92.0% (89.2%, 94.3%)
    Procedure Related AEsNot explicitly stated (implied low percentage)3.6% (4/111 of AEs at procedure), 0.0% (0/11 of related AEs). The overall rate of related AEs was 30% (9/30 subjects) with a count of 11 related AEs.12.0% (8.5%, 16.0%)
    Overall AEsNot explicitly stated (implied low percentage)93.3% (28/30 subjects). Total AEs: 111. SAEs: 35. One SAE was related to the device/procedure.22.7% (17.5%, 28.4%)

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Pivotal Clinical Study (Test Set):

      • Sample Size: 30 subjects (Intent-to-Treat, ITT). 28 subjects formed the Treated/Per-protocol cohort.
      • Data Provenance: The study was a "Multicenter, Prospective Study" (from G170190), indicating it was a forward-looking study where data was collected specifically for this evaluation. The document does not explicitly state the country of origin, but Boston Scientific Corporation is based in Marlborough, Massachusetts, USA, suggesting a likely multi-site international or US-based study framework.
    • Meta-Analysis of Clinical Literature:

      • Sample Size: 17 articles encompassing 713 patients.
      • Data Provenance: Retrospective, derived from published literature from various sources. One notable included study was a "randomized multicentre controlled superiority trial (DRAC 1)" (Teoh et al., Gut, 2020) which compared the AXIOS stent to percutaneous cholecystostomy, providing prospective, comparative data within the meta-analysis.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not provide information regarding the number of experts or their specific qualifications (e.g., radiologists with X years of experience) used to establish ground truth for the clinical study data. It is a clinical trial setting, so subject outcomes, diagnoses, and adverse events would be determined by the treating physicians and study staff according to study protocols, often involving standard clinical diagnostic criteria and adjudication by a clinical events committee, though none of these details are given in the provided text.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe a formal adjudication method (e.g., 2+1, 3+1) for the primary and secondary endpoints in the pivotal study. Clinical trials typically have predefined methods for recording and verifying endpoints and adverse events, often involving a clinical events committee (CEC), but these specific details are not provided. The determination of "symptom resolution" was based on objective clinical measures (temperature, pain score, WBC).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study typically assesses the performance of human readers with and without AI assistance on medical images. The AXIOS Stent is a medical device for drainage, not an imaging diagnostic AI device.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This question is not applicable in the context of this device. The AXIOS stent is a physical medical device and delivery system, not an AI algorithm. Its performance is evaluated through bench testing and clinical outcomes, reflecting the device's function directly and its interaction with human operators (physicians) during placement and use.

    7. The Type of Ground Truth Used

    • For Non-Clinical/Bench Studies: The ground truth was established by physical measurements, standardized test methods (e.g., ASTM, EN ISO standards), and direct observation against predefined engineering specifications and performance criteria.
    • For the Pivotal Clinical Study: The ground truth for effectiveness endpoints (e.g., symptom resolution, re-intervention) was based on clinical outcomes data from the patients enrolled in the prospective study, including physiological parameters (temperature, WBC), patient-reported pain scores, physician assessments of stent patency, and documented adverse events.
    • For the Meta-Analysis: The ground truth was based on the reported outcomes from published clinical literature, which would similarly derive from clinical assessments, diagnostic findings, and patient outcomes within those individual studies.

    8. The Sample Size for the Training Set

    This question is not applicable. The AXIOS stent is a physical medical device, not an AI algorithm that requires a "training set" for machine learning. The term "training set" is relevant for AI/ML model development.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable, as there is no "training set" in the context of this device's development or evaluation as described.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1