510(k) Data Aggregation

    K Number
    K250045

    Validate with FDA (Live)

    Device Name
    uWS-Angio Pro
    Date Cleared
    2025-09-24

    (257 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The uWS-Angio Pro is a post-processing software based on the uWS-Angio Basic for viewing, processing, evaluating and analyzing CT, MR, 3D XA images, and overlaying 2D medical images to support image guidance during interventional procedures. It has the following application:

    The uWS-Angio Pro uTarget view application is intended to provide tools for adding the trajectory to the lesion. It also supports overlay of 3D datasets on 2D fluoroscopy images.

    The uWS-Angio Pro uEmbo View application is intended to provide users with tools for planning and guiding interventional procedures. It offers a suite of features to mark targets, detect feeder vessels, and overlay the 3D image with the feeders and the corresponding anatomical X-ray image. When the overlay function is activated, it supports automatic 2D-3D image registration specifically for head neuroimaging, while the manual registration function is not restricted to any specific anatomical region.

    The uWS-Angio Pro uCardiac View application is intended to provide users tools to plan and guide the interventional procedures. This application can use CT angiography images for vascular access analysis, aortic root analysis and angle management, and overlay registration display with 2D XA images.

    Device Description

    uWS-Angio Pro is a software intended for viewing, processing, evaluating and analyzing medical images that comply with the DICOM 3.0 protocol. It supports interpretation and evaluation of examinations within healthcare institutions. It can be deployed on independent hardware such as a stand-alone diagnostic review and post-processing workstation. It can also be configured within a network to send and receive DICOM data. Furthermore, it can be deployed on systems of the United Imaging Healthcare Angiography system family.

    uWS-Angio Pro contains the following applications:

    • uWS-Angio Pro uTarget View
    • uWS-Angio Pro uEmbo View
    • uWS-Angio Pro uCardiac View
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250040

    Validate with FDA (Live)

    Device Name
    uWS-Angio
    Date Cleared
    2025-09-12

    (247 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    uWS-Angio is a post-processing software based on the uWS-Angio Basic for viewing, processing, evaluating and analyzing XA images. It has the following application:

    • The uWS-Angio 2D Vessel Analysis is intended to provide users a tool for quantifying the dimensions of coronary and peripheral vessels from 2D angiographic images. It supports vessel segmentation and output associated diameter parameters and curves.

    • The uWS-Angio Left Ventricular Analysis application is intended to provide users a tool for analyzing left ventricular functions. It supports editing contours of ventricular and outputs a set of parameters to show the ventricular functions and wall motion.

    • The uWS-Angio 2D Blood Flow Analysis application is intended to provide users a tool for analyzing blood flow of region of interest in peripheral vessels. It supports adding interest region and output multiple blood flow parameters for further perfusion analysis.

    • The uWS-Angio Stitching is intended to provide users a tool for lower extremities stitching bolus-chase acquired images into a stitching image.

    Device Description

    uWS-Angio is a post-processing software intended for viewing, processing, evaluating, and analyzing medical images that comply with the DICOM 3.0 protocol. It supports interpretation and evaluation of examinations within healthcare institutions. It can be deployed on independent hardware such as a stand-alone diagnostic review and post-processing workstation. It can also be configured within a network to send and receive DICOM data. Furthermore, it can be deployed on systems of the United Imaging Healthcare Angiography system family.

    uWS-Angio contains the following applications:

    • uWS-Angio 2D Vessel Analysis
    • uWS-Angio Left Ventricular Analysis
    • uWS-Angio 2D Blood Flow Analysis
    • uWS-Angio Stitching
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243632

    Validate with FDA (Live)

    Device Name
    uWS-Angio Basic
    Date Cleared
    2025-05-12

    (168 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    uWS-Angio Basic is intended to display and process XA (X-Ray Angiographic), CT (Computed Tomography), and MR (Magnetic Resonance) images that comply with the DICOM3.0 protocol.

    Device Description

    uWS-Angio Basic is an image processing software that matches the use of medical vascular angiography X-ray machines. It contains the following functions:

    • Patient Administration
    • Review 2D: This application can be used to load 2D images and perform related processing. The function provides Basic processing tools for
      1. 2D image viewing,
      2. 2D image processing,
      3. 2D DSA image viewing and processing,
      4. Calibration and Measurement.
    • Review 3D: This application can be used for loading and processing 3D data. The function provides Basic processing tools for
      1. 3D image viewing;
      2. 3D image processing;
      3. CTA bone removal.
    • Saving
    • Filming

    uWS-Angio Basic can be deployed on independent hardware such as a stand-alone diagnostic review and post processing workstation. It can also be configured within a network to send and receive DICOM data. Furthermore, it can be deployed on systems of the United-imaging Angiography system family.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the uWS-Angio Basic device does not contain the detailed performance data, acceptance criteria, or study design information typically found in a clinical study report or a more comprehensive technical and clinical assessment. The letter is a regulatory document affirming substantial equivalence to predicate devices, and while it mentions "Performance Testing" for certain algorithms, it only provides very high-level summaries and no specific quantitative acceptance criteria or detailed study results.

    Therefore, I cannot extract the full information requested for acceptance criteria and the study proving the device meets them from this document. However, I can infer and state what is present and what is absent.

    Here's an analysis based on the provided text, highlighting the limitations:

    Information NOT available in the provided document:

    • Detailed acceptance criteria with specific thresholds (e.g., AUPRC > 0.95, sensitivity > X%, specificity > Y%).
    • Specific reported device performance metrics against those criteria.
    • Sample size used for the test set.
    • Country of origin of data, or whether it was retrospective/prospective.
    • Number of experts used for ground truth, or their specific qualifications.
    • Adjudication method for the test set.
    • Whether a multi-reader multi-case (MRMC) comparative effectiveness study was done, or any effect size for human readers.
    • Specific details on standalone algorithm performance (beyond a general statement that one performs "as well as" and another has "high accuracy").
    • The type of ground truth used (e.g., pathology, outcomes data).
    • The sample size for the training set.
    • How the ground truth for the training set was established.

    Information that can be inferred or is explicitly mentioned (though not in the requested format):

    The document states:
    "Testing was conducted between uWS-Angio Basic and the predicate device to evaluate the performance of the algorithms mentioned above. It is shown that catheter calibration of uWS-Angio Basic has high accuracy with average error rates consistently lower than those of the predicate device. CTA Removing Bone of the uWS-Angio Basic perform as well as the predicate devices."

    This indicates that comparative performance against predicate devices was the primary method of demonstrating "safety and effectiveness," rather than meeting pre-defined quantitative performance metrics against a clinical ground truth.


    Attempted Fulfilling of Request (with significant caveats due to missing information):

    Given the limitations of the provided text, most of the requested table and details cannot be populated. The performance data section is extremely brief and qualitative.

    1. Table of acceptance criteria and reported device performance:

    Feature/AlgorithmAcceptance Criteria (Inferred from text: Comparative Performance)Reported Device Performance (Qualitative from text)
    Catheter CalibrationPerformance comparable to or better than predicate device."High accuracy with average error rates consistently lower than those of the predicate device."
    CTA Removing BonePerformance comparable to predicate device."Perform as well as the predicate devices."
    Overall Software Safety & EfficacyDemonstrated through verification and validation (hazard analysis, SRS, architecture, V&V, cybersecurity)"Found to have a safety and effectiveness profile that is similar to the predicate device and Secondary predicate devices."

    2. Sample size used for the test set and data provenance:

    • Sample Size: Not specified.
    • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

    • Number of experts: Not specified.
    • Qualifications of experts: Not specified.
    • Note: It's unclear if expert consensus was even the ground truth method, or if it was comparative to predicate output. This document does not suggest an independent ground truth for "accuracy" other than comparison to the predicate.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not specified.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study is mentioned. The study described focuses on algorithm-to-algorithm comparison (uWS-Angio Basic vs. predicate device algorithms), not human-in-the-loop performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, implicitly. The "Performance Testing" section describes evaluation of the "Catheter Calibration Algorithm for Review 2D" and "CTA Removing Bone Algorithm for Review 3D." This refers to the performance of these algorithms themselves, likely in a "standalone" fashion, by comparing their output to that of predicate device algorithms.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not explicitly stated. The document implies a "ground truth" derived from the performance or output of the predicate devices. For example, "average error rates consistently lower than those of the predicate device" for catheter calibration suggests a definition of "accuracy" relative to the predicate's output. It's not stated that an independent, gold-standard clinical ground truth (e.g., pathology, clinical outcomes, or expertconsensus on the true measurement/absence/presence of bone) was established for the performance testing of these algorithms.

    8. The sample size for the training set:

    • Not specified. The document does not discuss the training of the algorithms.

    9. How the ground truth for the training set was established:

    • Not specified, as training set details are not provided.

    Summary of what can be gleaned vs. what is missing:

    This 510(k) clearance letter primarily focuses on demonstrating "substantial equivalence" based on similar intended use, technological characteristics, and a basic comparative performance assessment against already cleared predicate devices. It does not provide the detailed breakdown of quantitative acceptance criteria, robust clinical study design, or comprehensive ground truth establishment methods that would be seen for higher-risk devices or novel AI functions requiring extensive clinical validation. The "Performance Testing" described is rudimentary comparative testing without specified quantitative metrics or detailed study populations.