510(k) Data Aggregation

    K Number
    K250434

    Validate with FDA (Live)

    Device Name
    trophon2
    Manufacturer
    Date Cleared
    2025-08-05

    (172 days)

    Product Code
    Regulation Number
    892.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The trophon2 is designed to provide High-Level Disinfection (HLD) of validated ultrasound transducers. High-Level Disinfection is achieved by surface exposure to a controlled dose of hydrogen peroxide mist delivered to a disinfection chamber contain the ultrasound probe.

    The trophon2 system consists of a multiple use instrument combined with a single use disinfectant "trophon Sonex-HL", delivered from a multi-dose cartridge.

    The trophon2 is suitable for use in general hospital and health care facilities by trained personnel.

    The trophon Sonex-HL should be used with the following contact conditions:

    • Minimum Operational Cycle Time: 4 minutes
    • Minimum Concentration: 31.5%
    • Minimum Disinfectant Dose: 1.0 g
    • Minimum Chamber Temperature: 56°C
    Device Description

    The trophon2 is a software-controlled device which provides High- Level Disinfection of validated ultrasound probes. The device consists of a sealed disinfection chamber and operates in conjunction with a multi-dose cartridge of concentrated hydrogen peroxide disinfectant, supplied as an accessory to the device. Pre-cleaned and dried ultrasound transducers are placed within the trophon2 chamber and disinfected by means of an automated disinfection and aeration cycle.

    AI/ML Overview

    The provided text is a 510(k) clearance letter for a medical device called "trophon2". It focuses on proving the new version of trophon2 is substantially equivalent to a previously cleared version, primarily through non-clinical testing. This document does not describe a study involving an AI model for medical image analysis, nor does it detail a multi-reader multi-case (MRMC) study or expert adjudication for ground truth related to medical imaging.

    The device, trophon2, is a High-Level Disinfection (HLD) system for ultrasound transducers, a piece of equipment for cleaning medical devices, not an AI diagnostic tool. Therefore, many of the requested categories (e.g., number of experts for ground truth, MRMC study, training set ground truth) are not applicable to the information provided in this regulatory document.

    However, based on the non-clinical testing section, I can extract and present the acceptance criteria and reported device performance related to its disinfection capabilities.


    Analysis of Acceptance Criteria and Device Performance for trophon2 (K250434)

    The trophon2 device, as described in the 510(k) submission K250434, is a High-Level Disinfection (HLD) system for ultrasound transducers. The evaluation presented in the document is primarily a demonstration of its substantial equivalence to a previously cleared version of the trophon2 (K173865) by fulfilling specified non-clinical performance and safety standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the trophon2 are based on its ability to perform High-Level Disinfection effectively and safely, as well as adherence to relevant industry standards.

    Test CategoryTest / ParameterApplicable Standard / GuidanceAcceptance CriteriaReported Device Performance (Result)
    EfficacyMycobactericidal Efficacy TestAOAC 6.3.06:2012; FDA Guidance "Content and Format Premarket Notification [510(k)] Submission for Liquid Chemical Sterilants/ High Level Disinfectants" Jan 2000, Section III. H.5aMeets recommendations of Section III. H.5aPass
    EfficacyFungicidal Efficacy TestAOAC 6.3.02:2006Not explicitly stated, implied by standard compliancePass
    EfficacyBactericidal Efficacy TestAOAC 6.2.02:2006, AOAC 6.2.03:2006, AOAC 6.2.05:2006Not explicitly stated, implied by standard compliancePass
    EfficacyVirucidal Efficacy Test (Poliovirus type 1)N/A (Internal protocol implied)Not explicitly stated, implied by Pass resultPass
    EfficacyVirucidal Efficacy Test (Herpes simplex virus type 1)N/A (Internal protocol implied)Not explicitly stated, implied by Pass resultPass
    Simulated UseSimulated Use TestASTM E1837-96(2014); FDA Guidance "Content and Format Premarket Notification [510(k)] Submission for Liquid Chemical Sterilants/ High Level Disinfectants" Jan 2000, Section III. H.4Meets recommendations of Section III. H.4Pass
    Critical Process Parameters1. Temperature SensorN/A (Internal protocol, same as K173865)The verification sensor temperature is equal to or greater than the lower verification sensor temperature limit at any time during all disinfection cycles at temperature setpoint.Pass
    Critical Process Parameters2. Dose Sensor vs Consumption AssessmentN/A (Internal protocol, same as K173865)The dose limits for a disinfection cycle shall correspond to the defined operational range of the device.Pass
    Critical Process Parameters3. Dosage MeasurementN/A (Internal protocol, same as K173865)Within the upper and lower limits of the dosage sensors corresponding to the defined operational range of the device.Pass
    Critical Process Parameters4. Flow Rate MeasurementN/A (Internal protocol, same as K173865)A statistically linear relationship (R² ≥ 0.99) exists between the flow rate and the max nebuliser fan rpm.Pass
    Safety / ResiduesChamber Venting AssessmentN/A (Internal protocol, same as K173865); FDA Guidance "Content and Format Premarket Notification [510(k)] Submission for Liquid Chemical Sterilants/ High Level Disinfectants" Jan 2000, Section III.I.2 and III. J.2Meets recommendations of Section III.I.2 and III. J.2Pass
    Safety / ResiduesChemical Indicator (CI) AssessmentN/A (Internal protocol, same as K173865); FDA Guidance "Content and Format Premarket Notification [510(k)] Submission for Liquid Chemical Sterilants/ High Level Disinfectants" Jan 2000, Section III.J.3Meets recommendations of Section III.J.3Pass
    Safety / ResiduesUltrasound Probe Residual H2O2 AssessmentN/A (Internal protocol, same as K173865); FDA Guidance "Content and Format Premarket Notification [510(k)] Submission for Liquid Chemical Sterilants/ High Level Disinfectants" Jan 2000, Section III. I.2Meets recommendations of Section III. I.2Pass
    Safety / ResiduesChemical Indicator Residual H2O2 AssessmentN/A (Internal protocol, same as K173865); Implied by guidance.Not explicitly stated, implied by Pass resultPass
    Safety / LeakageLeak TestN/A (Internal protocol, same as K173865); FDA Guidance "Content and Format Premarket Notification [510(k)] Submission for Liquid Chemical Sterilants/ High Level Disinfectants" Jan 2000, Section III. J.2Meets recommendations of Section III. J.2Pass

    2. Sample size used for the test set and the data provenance

    The document states that "Potency tests were conducted under worst case conditions" and "Simulated use testing was conducted under worst case conditions." However, specific sample sizes (e.g., number of tested devices, number of disinfection cycles, number of contaminated probes) for these non-clinical tests are not provided in the provided 510(k) summary. The data provenance is implied to be from Nanosonics Limited's internal testing in Australia, as the company is based in Macquarie Park, NSW, Australia. The document does not specify if the data was retrospective or prospective, though it pertains to performance testing of the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to the provided document. The trophon2 is a disinfection device, not an AI model that interprets medical images. Therefore, the concept of "ground truth" established by medical experts for a test set, as is common in AI/ML medical devices, does not apply here. The "ground truth" for this device's performance is objective measurements of microbial kill, chemical residue levels, and functional parameters, validated against established standards (e.g., AOAC, ASTM, FDA guidance).

    4. Adjudication method for the test set

    This question is not applicable to the provided document, as it does not describe studies involving human readers or interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable to the provided document. The trophon2 is not an AI diagnostic tool, and no MRMC study or AI assistance for human readers is described.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable to the provided document. While the trophon2 is a software-controlled system, it is a physical device performing disinfection, not an algorithm providing diagnostic output. The "software verification and validation testing" mentioned refers to the functional correctness of the device's control software, not diagnostic performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the trophon2's performance is based on microbiological standards (e.g., complete kill of specific microorganisms), chemical concentration measurements, temperature parameters, and physical operational limits. This is established by well-defined laboratory testing protocols outlined in standards like AOAC and ASTM, and FDA guidance for liquid chemical sterilants/high-level disinfectants. It is a deterministic, objective ground truth based on the device's physical and chemical disinfection capabilities, not interpretive expert consensus or patient outcomes data for diagnosis.

    8. The sample size for the training set

    This question is not applicable to the provided document. The trophon2 is a hardware device with controlling software; it does not involve machine learning or a "training set" in the context of AI model development for diagnostic purposes. Its software is developed and validated through traditional software engineering principles.

    9. How the ground truth for the training set was established

    This question is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173865

    Validate with FDA (Live)

    Device Name
    trophon2
    Manufacturer
    Date Cleared
    2018-04-24

    (125 days)

    Product Code
    Regulation Number
    892.1570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The trophon2 is designed to provide High-Level Disinfection (HLD) of validated ultrasound probes. High-Level Disinfection is achieved by surface exposure to a controlled dose of hydrogen peroxide mist delivered to a disinfection chamber containing the ultrasound probe.

    The trophon2 system consists of a multiple use instrument combined with a single use disinfectant "trophon Sonex-HL", delivered from a multi-dose cartridge.

    The trophon2 is suitable for use in general hospital and health care facilities by trained personnel.

    The trophon Sonex-HL should be used with the following contact conditions: Minimum Operational Cycle Time: 7 minutes Minimum Concentration: 31.5% Minimum Disinfectant Dose: 1.0 g Minimum Chamber Temperature: 56℃

    Device Description

    The trophon2 is a software controlled device which provides High-Level Disinfection of ultrasound transducers. The device consists of a sealed disinfection chamber and operates in conjunction with a multi-dose cartridge of concentrated hydrogen peroxide disinfectant "trophon Sonex- HL".

    Pre-cleaned and dried ultrasound transducers are placed within the trophon2 chamber and disinfected by means of an automated disinfection and aeration cycle.

    The disinfected ultrasound transducer is removed from the chamber, wiped and is ready for immediate use.

    AI/ML Overview

    The provided document is a 510(k) Pre-Market Notification for the trophon2 device, establishing its substantial equivalence to a predicate device (Trophon EPR) for High-Level Disinfection (HLD) of ultrasound transducers. It focuses on the device's technical specifications, comparisons to the predicate, and non-clinical testing performed to demonstrate its safety and effectiveness.

    Based on the provided document, here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't provide specific quantitative acceptance criteria or detailed reported performance values for each test. Instead, it makes a general statement:

    Acceptance Criteria (Implied)Reported Device Performance
    Device meets requirements for Electromechanical and EMC testing.The results demonstrated that the subject device met the acceptance criteria for each non-clinical test.
    Device meets requirements for Biocompatibility.The results demonstrated that the subject device met the acceptance criteria for each non-clinical test.
    Device meets requirements for Microbiological Efficacy Testing.The results demonstrated that the subject device met the acceptance criteria for each non-clinical test.
    Device meets requirements for Validation testing of process parameters.The results demonstrated that the subject device met the acceptance criteria for each non-clinical test.
    Device meets requirements for Materials Compatibility.The results demonstrated that the subject device met the acceptance criteria for each non-clinical test.
    Device meets requirements for Stability testing of Sonex-HL.The results demonstrated that the subject device met the acceptance criteria for each non-clinical test.
    Device meets requirements for Verification Testing of the Chemical Indicator.The results demonstrated that the subject device met the acceptance criteria for each non-clinical test.
    Critical operational parameters for HLD are maintained.The trophon2 maintains the same minimum operational cycle time (7 minutes), minimum concentration of hydrogen peroxide (31.5%), minimum disinfectant dose (1.0g), and minimum chamber temperature (56°C) as the predicate device.
    Device effectively removes residues from disinfected transducers."Effectively removes residues from disinfected transducers" (stated as a "Same" characteristic as the predicate).
    Device meets relevant performance standards (IEC 61010-1, IEC61010-2-040, IEC 61326, IEC 62304, ISO 62366 -1 and -2, ISO10993, ISO14971).Listed as "Same" as predicate, implying adherence to these standards.

    2. Sample size used for the test set and the data provenance:

    The document refers to "Non-clinical Testing" but does not specify the sample sizes used for any of the tests, nor does it explicitly state the data provenance (e.g., country of origin, retrospective or prospective). The testing appears to be primarily laboratory-based to validate the device's functional and biological performance rather than clinical studies with human subjects or medical images.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The type of testing described (electromechanical, biocompatibility, microbiological, etc.) typically relies on established scientific protocols and instrumentation rather than expert human interpretation for ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not provided in the document. Adjudication methods are typically relevant for studies involving human interpretation (e.g., image reading), which is not the primary focus of the non-clinical testing described here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, an MRMC comparative effectiveness study was not conducted or described in this document. This device is an automated high-level disinfection system, not an AI-assisted diagnostic tool that aids human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    The core of the device is an automated, standalone system for high-level disinfection. The "Non-clinical Testing" section describes validation of the device's automated functions: "Microbiological Efficacy Testing," "Validation testing of process parameters," and "Verification Testing of the Chemical Indicator." This implies that the device's performance in achieving HLD is evaluated independently of human intervention during the disinfection cycle.

    7. The type of ground truth used:

    The "ground truth" for the non-clinical tests would be established by:

    • Microbiological Standards: For microbiological efficacy, ground truth would be determined by established standards for killing specified microorganisms (e.g., AOAC methods mentioned on page 6).
    • Physical/Chemical Measurements: For process parameters (temperature, concentration, dose, time), ground truth would be established by precise measurements using calibrated instruments.
    • Engineering Specifications: For electromechanical and material compatibility, ground truth would be established by compliance with engineering design specifications and relevant international standards (e.g., IEC, ISO standards listed).

    8. The sample size for the training set:

    This information is not applicable and not provided. This device is a hardware system for disinfection and the document does not describe the use of machine learning or AI models that would require a "training set."

    9. How the ground truth for the training set was established:

    This information is not applicable and not provided for the same reason as above. There is no mention of a training set for an AI/ML model.