Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K160144

    Validate with FDA (Live)

    Device Name
    rainbow LS Block
    Manufacturer
    Date Cleared
    2016-07-21

    (182 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    rainbow™ LS Block can be used for manufacturing Crown, Veneer, Inlay and Onlay.

    Device Description

    rainbow™ LS Block is a dental ceramic made out of Lithium disilicate. rainbow™ LS Block is milled into cores of teeth and then is fired in a furnace to harden the Li2Si2O3. Then, the core is layered with porcelain to make a finished tooth.

    AI/ML Overview

    The provided information describes the acceptance criteria and performance data for a dental ceramic device named "rainbow™ LS Block," submitted for 510(k) premarket notification. The study primarily relies on bench testing and biocompatibility testing to demonstrate substantial equivalence to a predicate device (IPS e.max CAD).

    Here's the breakdown of the information that can be extracted from the document, organized according to your requested points.

    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a table summarizing the performance data, including criteria and results, primarily from bench testing and biocompatibility.

    No.ItemsCriteriaReported Device Performance (Result)Standard
    Biocompatibility Testing
    1CytotoxicityNone cytotoxicityNone cytotoxicityISO 10993-5
    2Sensitization (Local Lymph Node Assay)None sensitizationNone sensitizationISO 10993-10
    3Oral mucosa irritationLess than stimulus thresholdMinimal irritationISO 10993-10
    4Acute systemic toxicityNone acute Systemic toxicityNone acute Systemic toxicityISO 10993-11
    5Genotoxicity (Ames, Micronucleus)None genotoxicityNone genotoxicityISO 10993-3
    Performance Bench Testing
    1VisualNo impurities and No specific changesNo impurities and No specific changesISO 6872
    2SizeSize error of; Standard Size < ±5%(C-type) W 0.03%, L 3.90%, H 0.21%ISO 6872
    (Disk type) W 3.42%, L2.36%, H 0.07%ISO 6872
    3PackageNo damageNo damageISO 6872
    4UniformityUniformUniformISO 6872
    5Freedom from extraneous materialsNot Freedom from extraneous materialsNot Freedom from extraneous materialsISO 6872
    6Radioactivity238U Less than 1.0 Bq/g238U: < 0.00628 Bq/gISO 6872
    7Chemical solubilityLess than 100 $ \mu $g/cm²Solubility: 12 $ \mu $g/cm²ISO 6872
    8Flexural strengthMore than 100MPaAverage: 370MPaISO 6872
    9Linear thermal expansion ($10^{-6}K^{-1}$)9.95($\pm0.5$) $X 10^{-6}K^{-1}$Average: 9.95 $X 10^{-6}K^{-1}$ISO 6872
    10Glass-transition temperatureLess than 579$\pm20$ °C579 °CISO 6872

    Study Details

    2. Sample size used for the test set and the data provenance

    The document indicates that the evaluation was based on bench testing and biocompatibility testing. It does not specify a "test set" in the context of clinical data or images. The sample sizes for the specific tests (e.g., number of samples for flexural strength, chemical solubility) are not explicitly stated in this summary but are implied to be sufficient for compliance with the referenced ISO standards.

    The data provenance is from GENOSS Co., Ltd., a company based in Korea (1F, Gyeonggi R&DB Center / 226, 2F, GSBC, 105 Gwanggyo-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea). The study appears to be retrospective in the sense that it's a summary of completed lab tests rather than a prospective clinical trial.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to this study. The ground truth for this device (a dental material) is established through physical and chemical property measurements and standardized biological tests, not through expert consensus on medical images or clinical outcomes.

    4. Adjudication method for the test set

    This question is not applicable. There is no "test set" requiring human adjudication as this is not a diagnostic AI device or a clinical outcome study. The tests are benchtop measurements guided by ISO standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. This is a submission for a dental material (porcelain block) and does not involve AI or human readers for diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. There is no algorithm or AI component mentioned for this device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for this device is based on standardized physical, chemical, and biological measurements specified by international standards like ISO 6872 (dental ceramics) and ISO 10993 (biological evaluation of medical devices). For example:

    • Flexural strength: Measured in MPa against a standard criterion.
    • Chemical solubility: Measured in $ \mu $g/cm² against a standard criterion.
    • Biocompatibility: Assessed against criteria like "None cytotoxicity," "None sensitization," etc., based on established biological assays.

    8. The sample size for the training set

    This question is not applicable. This submission is for a physical dental material, not a machine learning or AI device that would require a "training set."

    9. How the ground truth for the training set was established

    This question is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1