510(k) Data Aggregation

    K Number
    K251610

    Validate with FDA (Live)

    Device Name
    qER-CTA (v1.0)
    Date Cleared
    2025-09-08

    (104 days)

    Product Code
    Regulation Number
    892.2080
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    qER-CTA is a notification-only, parallel workflow tool for use by hospital networks and trained clinicians to identify and communicate images of specific patients to a specialist, independent of the standard of care workflow. qER-CTA uses a deep learning algorithm to analyze images for findings suggestive of a pre-specified clinical condition and to notify an appropriate medical specialist in parallel to standard of care image interpretation. Identification of suspected findings is not for diagnostic use beyond notification. Specifically, the device analyses CT angiogram images of the brain acquired in the acute setting and sends notifications to a neurovascular specialist that a suspected large vessel occlusion has been identified, recommending review of those images. Images can be previewed through a mobile application. qER-CTA is intended to analyze the internal carotid artery (ICA) and M1 segment of the middle cerebral artery (MCA) for LVOs on CTA scans of adults (≥ 22 years of age). Images previewed through the mobile application are compressed and for informational purposes only, not intended for diagnostic use beyond notification. Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer, conducting appropriate patient evaluation, and engaging in relevant discussions with the treating physician before making care-related decisions or requests. qER-CTA is limited to the analysis of imaging data and should not be used as a substitute for full patient evaluation or relied upon to make or confirm a diagnosis.

    Device Description

    qER-CTA is a radiological computer-aided triage and notification (CADt) software designed to assist trained clinicians and radiologists in analyzing and triaging head CTA scans for suspected LVO (Large Vessel Occlusion) in the anterior circulation.

    The software uses a deep learning algorithm to analyze CTA images and provide a case-level output available in the PACS or workstation for worklist prioritization or triage. It does not alter the original image, change the worklist order, or send proactive alerts directly to the end user. Instead, the end user can sort the worklist based on the passive notification flag. Images can be previewed through a mobile application also. There are two alternatives' users can choose from engaging with qER-CTA.

    1. For de-identified CTA scans, they are sent to qER-CTA via transmission functions built within the user's PACS or workstation. Results are pushed back to the user's PACS or other user-specified radiology software database once the processing is complete.

    2. For the client system that does not have de-identification and re-identification capabilities, qER-CTA interacts with on-premises gateway rather than directly with the PACS.

    qER-CTA is not intended to direct attention to specific portions of the image, rule out target conditions, or be used as a standalone tool for clinical decision-making. It operates as a parallel workflow tool, independent of the standard of care, to assist in identifying and communicating suspected LVO cases to appropriate medical specialists for further review. Images previewed through the mobile application are compressed and are for informational purposes only.

    AI/ML Overview

    1. Acceptance Criteria and Reported Device Performance

    AbnormalityAcceptance CriteriaReported Device Performance (95% CI)
    Large Vessel OcclusionNot explicitly stated in the provided text. Likely compared against predicate.AUC: 0.959 (0.943 – 0.975)
    Sensitivity: 91.35% (87.54%-94.07%)
    Specificity: 91.86% (88.18% -94.47%)
    Time to NotificationNot explicitly stated in the provided text. Likely compared against predicate.Mean: 6.36 minutes (6.06-6.66)

    2. Sample Size for Test Set and Data Provenance

    • Sample Size: 584 head CTA scans (289 LVO, 295 non-LVO).
    • Data Provenance: Not explicitly stated in the provided text (e.g., country of origin, retrospective or prospective).

    3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set)

    • Number of Experts: Three.
    • Qualifications: U.S. board-certified neuroradiologists with at least 10 years of experience.

    4. Adjudication Method for the Test Set

    • The adjudication method is not explicitly mentioned. It states that "Three U.S. board certified neuroradiologists with at least 10 years of experience did the ground truthing," implying a consensus or majority vote might have been used, but specific details (e.g., 2+1, 3+1) are not provided.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a MRMC comparative effectiveness study was not done. The document explicitly states that the performance was assessed using a "standalone study."

    6. Standalone Performance (Algorithm Only) Study

    • Yes, a standalone study was done. The performance of the qER-CTA device was assessed using a standalone study, evaluating its classification of large vessel occlusion.

    7. Type of Ground Truth Used

    • Expert Consensus: The ground truth was established by three U.S. board-certified neuroradiologists with at least 10 years of experience.

    8. Sample Size for the Training Set

    • The sample size for the training set is not provided in the given text.

    9. How Ground Truth for the Training Set Was Established

    • How the ground truth for the training set was established is not provided in the given text. The document only mentions the ground truthing for the clinical performance testing (test set).
    Ask a Question

    Ask a specific question about this device

    K Number
    K211222

    Validate with FDA (Live)

    Device Name
    qER-Quant
    Date Cleared
    2021-07-30

    (98 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The qER-Quant device is intended for automatic labeling, visualization of segmentable brain structures from a set of Non-Contrast head CT (NCCT) images. The software is intended to automate the current manual process of identifying, labeling and quantifying the volume of segmentable brain structures identified on NCCT images. qER-Quant provides volumes from NCCT images acquired at a single time point and provides a table with comparative analysis for two or more images that were acquired on the same image acquisition protocol for the same individual at multiple time points.

    The qER-Quant software is indicated for use in the following structures: Intracranial Hyperdensities, Lateral Ventricles and Midline Shift.

    Device Description

    qER-Quant is a standalone software device that processes non-contrast head CT scans to outline and quantify the structures described in the intended use. The qER-Quant software interacts with the user's picture archiving and communication system (PACS) to receive scans and returns the results to the same destination.

    The analysis module of the qER-Quant software contains of a set of pre-trained convolutional neural networks (CNNs), that form the core processing component shown in Figure 1. This core processing component is coupled with a pre-processing module to prepare input digital imaging and communications in medicine (DICOMs) for processing by the CNNs and a post-processing module to convert the output into visual and tabular output for users.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the qER-Quant device, based on the provided text:


    qER-Quant Device Performance Study Details

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria were defined based on the accuracy of the qER-Quant system when compared against manually labeled ground truth. The reported device performance met these pre-set criteria.

    MetricAcceptance Criteria (Implied / Context)Reported Device Performance (Mean ± SD / Mean (95% CI) / Median (10th-90th Percentile))
    Intracranial Hyperdensity
    Absolute Error (Volume)Exceeds preset acceptance criteria6.56 (7.33) ml (Mean ± SD) 3.98 (0.52 - 17.35) ml (Median (10th - 90th percentile))
    Dice Score (Segmentation Accuracy)Exceeds preset acceptance criteria0.75 (0.72 - 0.78) (Mean (95% CI))
    Midline Shift
    Absolute Error (Shift)Exceeds preset acceptance criteria1.37 (1.23) mm (Mean ± SD) 1.15 (0.23 - 2.59) mm (Median (10th - 90th percentile))
    Dice Score (Segmentation Accuracy)Not ApplicableNot applicable
    Left Lateral Ventricle
    Absolute Error (Volume)Exceeds preset acceptance criteria2.09 (1.88) ml (Mean ± SD) 1.60 (0.29 - 4.24) ml (Median (10th - 90th percentile))
    Dice Score (Segmentation Accuracy)Exceeds preset acceptance criteria0.79 (0.78 - 0.81) (Mean (95% CI))
    Right Lateral Ventricle
    Absolute Error (Volume)Exceeds preset acceptance criteria2.18 (1.72) ml (Mean ± SD) 1.88 (0.40 - 4.53) ml (Median (10th - 90th percentile))
    Dice Score (Segmentation Accuracy)Exceeds preset acceptance criteria0.75 (0.73 - 0.77) (Mean (95% CI))

    2. Sample Size and Data Provenance

    • Test Set Sample Sizes:
      • Intracranial Hyperdensity: 183 scans
      • Midline Shift: 188 scans
      • Left Lateral Ventricle: 210 scans
      • Right Lateral Ventricle: 210 scans
      • Reproducibility testing was done on 20% of these CT scans.
    • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. It uses "a set of head CT scans."

    3. Number of Experts and Qualifications for Ground Truth Establishment

    • Number of Experts: The document states "experts" (plural) were used but does not specify the exact number.
    • Qualifications of Experts: Not specified beyond being "experts" in the context of manually labeling CT scans.

    4. Adjudication Method for the Test Set

    The document does not explicitly state an adjudication method (e.g., 2+1, 3+1). It only mentions that the ground truth was established by "manually labeled by experts."

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not reported. The performance testing was a "standalone" evaluation of the device's accuracy against expert-generated ground truth.

    6. Standalone Performance (Algorithm Only)

    • Yes, a standalone performance study was conducted. The document states: "Qure.ai performed standalone consisted of a set of head CT scans with the outlines of the target structures manually labeled by experts." The results detailed in Table 2 are of this standalone performance.

    7. Type of Ground Truth Used

    • The ground truth used was expert consensus / manual labeling. The document clearly states: "manually labeled by experts."

    8. Sample Size for the Training Set

    • The document does not provide the sample size for the training set. It only describes the architecture of the analysis module as "a set of pre-trained convolutional neural networks (CNNs)."

    9. How Ground Truth for the Training Set Was Established

    • The document does not explicitly state how the ground truth for the training set was established. It describes the CNNs as "pre-trained," which implies a training phase using labeled data, but the method of ground truth establishment for that specific data is not detailed.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200921

    Validate with FDA (Live)

    Device Name
    qER
    Date Cleared
    2020-06-17

    (72 days)

    Product Code
    Regulation Number
    892.2080
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    qER is a radiological computer aided triage and notification software in the analysis of non-contrast head CT images.

    The device is intended to assist hospital networks and trained medical specialists in workflow triage by flagging the following suspected positive findings of pathologies in head CT images: intracranial hemorrhage, mass effect, midline shift and cranial fracture.

    qER uses an artificial intelligence algorithm to analyze images on a standalone cloud-based application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include non-diagnostic preview images that are meant for informational purposes only. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

    The results of the device are intended to be used in conjunction information and based on professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

    Device Description

    Qure.ai Head CT scan interpretation software, qER, is a deep-learning-based software device that analyses head CT scans for signs of intracranial hemorrhage, midline shift, mass effect or cranial fractures in order to prioritize them for clinical review. The standalone software device consists of an on-premise module and a cloud module. qER accepts non-contrast adult head CT scan DICOM files as input and provides a priority flag indicating critical scans. Additionally, the software has the preview of critical scans to the medical specialist.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the qER device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    The core purpose of the qER device is workflow triage by identifying suspected positive findings of pathologies in head CT images. The performance data presented focuses on the device's ability to accurately detect these pathologies in a standalone setting.

    1. Table of Acceptance Criteria and Reported Device Performance

    While the document doesn't explicitly state "acceptance criteria" as numerical thresholds beyond "exceeded the predefined success criteria, as well as the required performance criteria for triage and notification software as per the special controls for QAS," the reported sensitivities and specificities for each pathology effectively serve as the demonstrated "acceptance" level the device achieved.

    AbnormalityAcceptance Criteria (Implied Success)Reported Device Performance (Sensitivity [95% CI])Reported Device Performance (Specificity [95% CI])Reported Device Performance (AUC [95% CI])
    Intracranial HemorrhageHigh sensitivity & specificity for triage96.98 (95.32 - 98.17)93.92 (91.87 - 95.58)98.53 (98.00 - 99.15)
    Cranial FractureHigh sensitivity & specificity for triage96.77 (93.74 - 98.60)92.72 (91.00 - 94.21)97.66 (96.88 - 98.57)
    Mass EffectHigh sensitivity & specificity for triage96.39 (94.28 - 97.88)96.00 (94.45 - 97.21)99.09 (98.73 - 99.52)
    Midline ShiftHigh sensitivity & specificity for triage97.34 (95.30 - 98.67)95.36 (93.79 - 96.64)99.09 (98.74 - 99.51)
    Any of the 4 target abnormalitiesHigh sensitivity & specificity for triage98.53 (97.45 - 99.24)91.22 (88.39 - 93.55)NA

    Additionally, a key performance metric for a triage device is the time to notification:

    ParameterAcceptance Criteria (Implied improvement over std. care)Reported Device Performance (Mean [95% CI])Reported Device Performance (Median [95% CI])
    Time to open exam in the standard of careBenchmark for comparison65.54 (59.14 - 71.76) min60.01 (54.57 - 77.63) min
    Time-to-notification with qERSignificantly lower than standard of care2.11 (1.45 - 2.61) min1.21 (1.12 - 1.25) min

    Study Details

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 1320 head CT scans.
    • Data Provenance: Retrospective, multicenter study. Data originated from multiple locations within the United States.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: 3 board-certified radiologists.
    • Qualifications: The document explicitly states "board-certified radiologists." No further details on years of experience are provided.

    4. Adjudication Method for the Test Set

    • The text states that the ground truth was established by "3 board-certified radiologists reading the scans." It does not explicitly mention an adjudication method (e.g., 2+1, 3+1 consensus). It is implied that their readings defined the ground truth, but the process of resolving discrepancies among the three readers is not detailed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and what was the effect size of how much human readers improve with AI vs without AI assistance

    • The provided text does not describe an MRMC comparative effectiveness study where human readers' performance with and without AI assistance was directly compared. The study primarily focuses on the standalone performance of the qER algorithm and its ability to reduce the "time to notification" compared to standard of care "time to open." While the "time-to-notification" analysis suggests a significant workflow improvement when using qER for triage (2.11 mins vs. 65.54 mins), this is not a direct measure of human reader diagnostic accuracy improvement with AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, a standalone performance study was done. The "Performance Data" section explicitly states, "A retrospective, multicenter, blinded clinical study was conducted to test the accuracy of qER at triaging head CT scans... Sensitivity and specificity exceeded the predefined success criteria... demonstrating the ability of the qER device to effectively triage studies containing one of these conditions." The results in Table 2 are for the qER algorithm's accuracy independently.

    7. The Type of Ground Truth Used

    • Expert Consensus: The ground truth for the pathologies (Intracranial hemorrhage, cranial fractures, mass effect, midline shift, and absence of these abnormalities) was established by "3 board-certified radiologists reading the scans." This indicates an expert consensus approach to defining the ground truth.

    8. The Sample Size for the Training Set

    • The document does not specify the sample size used for the training set. It mentions that the qER software uses "a pre-trained artificial intelligence algorithm" and "a pre-trained classification convolutional neural network (CNN) that has been trained to detect a specific abnormality from head CT scan images." However, the size of the dataset used for this training is not disclosed in the provided text.

    9. How the Ground Truth for the Training Set was Established

    • The document does not explicitly describe how the ground truth for the training set was established. It only states that the CNN was "pre-trained" on medical images to detect specific abnormalities. It is common practice for such training to also rely on expert annotations, but this is not detailed for the training set in this document.