510(k) Data Aggregation

    K Number
    K161032

    Validate with FDA (Live)

    Date Cleared
    2016-12-22

    (253 days)

    Product Code
    Regulation Number
    888.3075
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    neon3™ is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical spine (C1 to C7) and the thoracic spine (T1 to T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease. including intractable radiculopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. neon3™ is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration of fusion.

    Device Description

    neon3™ is a modular, posterior system used for the surgical stabilization and fixation of the occipital, cervical and thoracic regions of the spine. The system components include longitudinal rods, screw and hook anchors, and interconnecting devices such as anchor- to-rod connectors, and rod-to-rod and screw-to-screw crosslinks.

    AI/ML Overview

    The document describes the neon3™ universal OCT spinal stabilization device, a modular posterior system for spinal stabilization and fixation.

    However, the provided text does not contain details about acceptance criteria, device performance metrics, or a study comparing its performance against specific quantitative criteria for diagnostic accuracy or clinical outcomes as would be found in a study proving a device meets acceptance criteria.

    Instead, the document details the device's 510(k) premarket notification for FDA clearance, focusing on substantial equivalence to predicate devices based on mechanical testing.

    Here's an analysis based on the information available regarding the provided request:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of acceptance criteria as quantitative metrics (e.g., sensitivity, specificity, accuracy) for device performance in a clinical or diagnostic context.

    The performance data section mentions:

    • "Mechanical testing of worst case neon3TM craniocervical constructs included static and dynamic compression bending and static torsion according to ASTM F1717 as well as static and dynamic compression bending and static and dynamic torsion per ASTM F2706."
    • "The mechanical test results demonstrate that neon3TM performance is substantially equivalent to the predicate devices."

    This indicates that the "acceptance criteria" were likely related to meeting the performance specifications outlined in the ASTM standards (F1717 and F2706) for spinal implant mechanical properties, and demonstrating substantial equivalence to predicate devices (Synapse Occipital-Cervical-Thoracic (OCT) System, Halifax Interlaminar Clamp, neon3TM (K150650), and Ascent™ POCT System) in these mechanical properties. The specific numerical values or success/failure thresholds for these mechanical tests are not presented in this document.

    2. Sample sized used for the test set and the data provenance

    Not applicable. The "test set" in this context refers to mechanical testing of device constructs, not a clinical data set. The sample size for the mechanical testing is not specified beyond "worst case neon3TM craniocervical constructs". Data provenance is the manufacturing process and materials used for the constructs.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a mechanical testing study, not a study requiring expert clinical assessment for ground truth.

    4. Adjudication method for the test set

    Not applicable. This is a mechanical testing study, not a study involving human reader adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a mechanical testing study for a spinal stabilization device, not an AI-assisted diagnostic or imaging device for which an MRMC study would be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a mechanical testing study for a spinal stabilization device.

    7. The type of ground truth used

    For the mechanical testing, the "ground truth" would be established by the physical and mechanical properties of the materials and constructs themselves, as measured by standardized testing procedures (ASTM F1717 and F2706). The comparison point is the performance of predicate devices, implying that their mechanical properties served as a benchmark for substantial equivalence.

    8. The sample size for the training set

    Not applicable. There is no training set mentioned, as this is not an AI or machine learning study.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150650

    Validate with FDA (Live)

    Device Name
    neon3
    Date Cleared
    2015-09-24

    (196 days)

    Product Code
    Regulation Number
    888.3075
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    neon3™ is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1 to C7) and the thoracic spine (T1 to T3): traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability.

    neon?™ is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

    Device Description

    neon 3™ is a modular, posterior system used for the surgical stabilization and fixation of the cervical and thoracic regions of the spine. The system components include longitudinal members, anchors and interconnection devices.

    AI/ML Overview

    The acceptance criteria and study proving the device meets the criteria are described below:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Mechanical Testing: Static and dynamic compression bending and torsion according to ASTM F1717.Mechanical testing of worst-case neon3™ constructs included static and dynamic compression bending and torsion according to ASTM F1717.
    Tulip/Shank Dissociation Testing: Performed on worst-case screws.Tulip/shank dissociation testing was performed on the worst-case neon3™ screws.
    Substantial Equivalence: Performance is substantially equivalent to predicate devices based on mechanical test results.Published literature and the mechanical test results demonstrate that neon3™ performance is substantially equivalent to the predicate devices.

    Study Details

    1. Sample size used for the test set and the data provenance: Not applicable. This study primarily involved mechanical testing of the device components, not human subject data.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. The ground truth for mechanical testing is established by industry standards (ASTM F1717) and engineering principles rather than expert consensus.

    3. Adjudication method for the test set: Not applicable. This was a mechanical testing study, not a clinical study requiring adjudication of expert opinions.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a spinal fixation system, not an AI software or imaging device that would involve human readers.

    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is a hardware medical device, not an algorithm.

    6. The type of ground truth used:

      • Mechanical performance: Established by adherence to ASTM F1717 (standard specification for spinal implant constructs) and engineering design principles.
      • Substantial Equivalence: Demonstrated by comparing the mechanical test results of neon3™ to those of the predicate devices.
    7. The sample size for the training set: Not applicable. This refers to a medical device's mechanical performance evaluation, not a machine learning model.

    8. How the ground truth for the training set was established: Not applicable.