    K Number
    K231683

    Validate with FDA (Live)

    Device Name
    inHEART Models
    Manufacturer
    Date Cleared
    2024-02-29

    (265 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    inHEART MODELS comprises a suite of medical imaging software modules that are intended to provide qualified medical professionals with tools to aid them in reading, interpreting, and treatment planning, inHEART MODELS accepts DICOM compliant medical images acquired from a variety of imaging devices, including CT and MR*. The software is designed to be used by qualified medical professionals (including physicians, cardiologists, radiologists, and technicians) and the users are solely responsible for making all final patient management decisions.

    • inHEART Models AI software module is indicated for adults only and is designed for CT images only.
    Device Description

    inHEART MODELS is a suite of medical image processing software tools that enables 3D visualization and analysis of anatomical structures.

    This software suite is composed of three software as a medical device components:

    • . inHEART MODELS AI: a medical image processing software for automatic 3D modelling, used to pre-process medical images (acquired only by CT devices).
      This software module uses a machine-learning based approach with the following characteristics:

    • Training dataset: 796 cases (3D CT original images and the segmentation । masks) from previously manually performed segmentations (time period 2018-2022); origins of the data are public and private clinical and hospital institutions located in US (40%) and Europe (60%).

    • -Training process: Machine learning algorithm (UNet) is trained applying data augmentation (including patch mirroring) and regularization (including Instance Normalization, Dropout, Data Augmentation, Weight Initialization, Leaky ReLU Activation) methods. Loss functions (soft dice loss and binary cross entropy) and optimization methods (stochastic gradient descent) are used during learning over a fixed number of 1000 epochs, each consisting of 250 iterations with a batch size of 2.

    • Data anonymized prior to its processing by inHEART team, therefore no । personal data (gender, age, ethnicity) is exploitable. CT scanner manufacturers include Siemens (40%). GE Medical Systems (30%). Toshiba (10%) and other manufacturers (20%) among which Philips and Canon.

    • inHEART MODELS Shaper: a standalone image processing software used to . generate digital 3D models of the patient heart from medical images acquired by CT and/or MR devices and;

    • inHEART MODELS Explorer: a web-based 3D visualization software that permits . display, review, analysis, annotation and export of the cardiac 3D models generated from inHEART MODELS Shaper.

    Specifically, these software modules read DICOM compatible pre-operative CT and MR images acquired by commercially available imaging devices. These images are then processed to generate 3D models of the anatomy to allow qualified medical professionals to display, review, analyze, annotate, interpret, export, and plan therapeutic interventions.

    inHEART MODELS software suite also includes two non-device Medical Device Data Systems (MDDS) modules that are only intended to transfer, store and convert formats.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device's performance, based on the provided text:

    Acceptance Criteria and Study Details for inHEART MODELS

    The inHEART MODELS AI software module's performance was evaluated to ensure its automated segmentation capabilities, when reviewed and edited by expert users, yield results comparable to those obtained using the previously cleared predicate device alone.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are defined based on the objective of providing an initial segmentation that accelerates the process for expert segmenters, rather than replacing manual expert segmentation entirely. The reported performance metrics are computed between the revised segmentations made by experts based on AI results and the Ground Truth (GT).

    MetricAcceptance Criteria (for acceptance without manual correction)Overall Reported Device Performance (Mean)Overall Reported Device Performance (Median)
    Dice Coefficient> 0.90.940.94
    Average Symmetric Surface Distance (ASSD)< 5 mm1.17 mm1.05 mm (calculated from median of individual structures)
    Volumetric Analysis (Absolute Difference)< 20 mL9.18 mL6.85 mL (calculated from median of individual structures)
    Volumetric Analysis (Relative Difference)Not explicitly stated for acceptance without manual correction, but for main chambers: <10% after correction.7%4.5% (calculated from median of individual structures)

    Note on Volumetric Analysis: The table in the document provides values for individual structures. The "Overall Reported Device Performance" rows above are calculated as the average/median of the means and medians respectively across all listed structures from the provided table.

    Individual structure performance:

    StructureDICE (Median)DICE (Mean)DICE (StdDev)ASSD (Median)ASSD (Mean)ASSD (StdDev)Volume diff. (mL) (Median)Volume diff. (mL) (Mean)Volume diff. (mL) (StdDev)Volume diff. (%) (Median)Volume diff. (%) (Mean)Volume diff. (%) (StdDev)
    Aorta0.950.930.100.640.770.543.954.744.075%8%11%
    Left Atrium0.950.950.021.161.240.615.977.848.983%5%6%
    Left Ventricle Endocardium0.970.970.020.970.970.024.926.365.322%3%3%
    LV Epicardium0.970.970.010.710.760.307.2210.109.962%3%2%
    Pulmonary Artery Trunk0.930.910.071.412.022.157.3810.8411.389%14%20%
    Right Atrium & Inf. Vena Cava0.930.920.041.171.300.577.289.8710.124%6%6%
    Right Ventricle Endocardium0.920.910.051.221.250.569.8914.9914.987%11%12%
    RV Epicardium0.940.930.030.951.060.456.428.678.074%4%4%

    2. Sample Size and Data Provenance

    • Test Set Sample Size: 100 CT cases.
    • Data Provenance (Test Set): Randomly selected from the time period year -2023 to ensure independence from training datasets. The sources are similar to the training dataset, including new clients. The overall geographic origin for training data (which implies similar origins for test data) is US (40%) and Europe (60%).
    • Retrospective/Prospective: Not explicitly stated, but the selection from a time period "year -2023" for testing suggests a retrospective approach.

    3. Number and Qualifications of Experts for Ground Truth

    • Number of Experts: Two external experts.
    • Qualifications of Experts: Radiologists. No specific years of experience are detailed, but they are referred to as "external experts" and were tasked with evaluating concordance of manual segmentations.

    4. Adjudication Method for the Test Set

    The ground truth was established by "Approval by external experts (radiologists) of segmentations previously obtained with the predicate device, that will constitute the Ground Truth (GT)". This implies a consensus or agreement method, but the specific adjudication method (e.g., 2+1, 3+1) is not explicitly stated. It seems the two experts had to agree for the segmentations to form the GT.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC study was not explicitly described in terms of human readers improving with AI vs. without AI assistance. The study compares the performance of the AI-assisted, expert-reviewed results against the originally expert-segmented predicate device results (Ground Truth). The stated "effect size" is the performance metrics (Dice, ASSD, Volume difference) demonstrating that the AI-assisted workflow, after expert review and editing, is substantially equivalent to the manual predicate workflow. The purpose of the AI module is to accelerate the process for the expert segmenter by providing an initialization.

    6. Standalone (Algorithm Only) Performance

    Yes, a form of standalone performance was assessed as part of the study: "Comparison of unprocessed outputs of the new automated software module results with the GT to establish the internal validation of the tool performance". However, the specific metrics for this direct AI output (without expert correction) are not separately provided in the summary table. The provided table (under "Performance testing results") shows the metrics after expert review and revision based on AI results.

    7. Type of Ground Truth Used

    The ground truth (GT) was established through expert consensus/approval of segmentations previously obtained using the predicate device. This is referred to as "Approval by external experts (radiologists) of segmentations previously obtained with the predicate device, that will constitute the Ground Truth (GT)".

    8. Sample Size for the Training Set

    • Training Dataset: 796 cases (3D CT original images and segmentation masks).

    9. How the Ground Truth for the Training Set Was Established

    The ground truth for the training set consisted of "previously manually performed segmentations". It is implied these manual segmentations were performed by qualified personnel, similar to how the predicate device was used to establish the test set GT. The document does not specify the number of experts or the exact process (e.g., consensus) for generating the training set ground truth, but it strongly suggests expert-generated manual segmentations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220727

    Validate with FDA (Live)

    Device Name
    inHEART MODELS
    Manufacturer
    Date Cleared
    2022-05-10

    (57 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    inHEART MODELS comprises a suite of medical imaging software modules that are intended to provide qualified medical professionals with tools to aid them in reading, interpreting, and treatment planning, inHEART MODELS accepts DICOM compliant medical images acquired from a variety of imaging devices, including CT and MR. The software is designed to be used by qualified medical professionals (including physicians, cardiologists, radiologists, and technicians) and the users are solely responsible for making all final patient management decisions.

    Device Description

    inHEART MODELS is a suite of medical image processing software tools that enables 3D visualization and analysis of anatomical structures. Specifically, the software modules read DICOM compatible anonymized pre-operative CT and MR images acquired by commercially available imaging devices. These images are then processed to generate 3D models of the anatomy to allow qualified medical professionals to display, review, analyze, annotate, interpret, export, and plan therapeutic interventions. inHEART MODELS includes two non-device Medical Device Data Systems (MDDS) modules that are only intended to transfer, store and convert formats.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the inHEART MODELS device, based on the provided FDA 510(k) summary:

    Overview:
    The inHEART MODELS software is intended to provide tools for qualified medical professionals to aid in reading, interpreting, reporting, and treatment planning using DICOM compliant CT and MR images. The performance study aimed to demonstrate the accuracy of segmentations produced by inHEART MODELS compared to ground truth and previously cleared comparable software tools, establishing substantial equivalence to the predicate device (K200973 - Synapse 3D Cardiac Tools).


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document doesn't detail specific quantitative acceptance criteria (e.g., minimum dice score, maximum Hausdorff distance) or the precise numerical performance metrics achieved by the inHEART MODELS. Instead, it states a qualitative outcome for the primary performance evaluation.

    Acceptance Criterion (Implicit)Reported Device Performance
    Accuracy of segmentations compared to ground truthAll validation criteria were met. The performance evaluation study demonstrated that output segmentations for inHEART MODELS were substantially equivalent to previously cleared legally marketed software devices.
    Accuracy of measurements based on segmentationsAll validation criteria were met. The performance evaluation study demonstrated that measurements on these segmentations for inHEART MODELS were substantially equivalent to previously cleared legally marketed software devices.
    Absence of new safety or efficacy issues"No new safety or efficacy issues were introduced by inHEART MODELS compared to the predicate device."
    Functional equivalence to predicate device"Performance data demonstrate that the functionality, output and clinical usage of inHEART MODELS is substantially equivalent to the predicate device."
    Software V&V according to FDA Guidance"Software verification and validation testing were conducted on all inHEART MODELS software modules and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.'" (The device was considered a "moderate" level of concern, implying V&V was performed to the appropriate standard for this level).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not explicitly state the number of images or cases in the test set. It mentions "CT and MR images were selected for the performance study."
    • Data Provenance: Not specified in the provided text (e.g., country of origin, retrospective or prospective collection).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not specified. The document states, "Radiologists were designated as ground-truth annotators." This implies more than one, but the exact number isn't provided.
    • Qualifications of Experts: "Radiologists." Specific experience levels (e.g., "10 years of experience") are not detailed.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not specified. It only states that "Radiologists were designated as ground-truth annotators," but doesn't explain if this involved consensus, majority vote, or individual expert review.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study Done: No. The study described compares the device's segmentations and measurements to ground truth and other cleared software tools, not explicitly studying the effect of AI assistance on human readers' performance. The focus is on the software's standalone output being substantially equivalent.
    • Effect Size of Human Improvement with AI: Not applicable, as an MRMC comparative effectiveness study involving human readers' performance with and without AI was not described.

    6. Standalone (Algorithm Only) Performance Study

    • Standalone Study Done: Yes. The performance study directly assessed the "accuracy of segmentations for inHEART MODELS" compared to ground truth and other software tools, indicating a standalone evaluation of the algorithm's output. The statement "output segmentations and measurements on these segmentations for inHEART MODELS were substantially equivalent" confirms this.

    7. Type of Ground Truth Used

    • Type of Ground Truth: Expert consensus / Expert annotation. The document states that "Radiologists were designated as ground-truth annotators" who generated 3D heart models. This indicates manual review and annotation by medical experts as the basis for ground truth.

    8. Sample Size for the Training Set

    • Training Set Sample Size: Not specified in the provided document. The document focuses on the validation or test set performance.

    9. How Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set Establishment: Not specified in the provided document. This information would typically be relevant for machine learning models but is not detailed for inHEART MODELS in this summary. The process for generating ground truth for the test set is described (radiologists as annotators), but not for any potential training set.