Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K223794

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-01-17

    (29 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    hyperion X9 pro, NewTom GiANO HR, X-RADiUS TRiO PLUS is digital panoramic, cephalometric and tomographic extra-oral system, intended to:

    (1) produce orthopanoramic images of the maxillofacial region and carry out diagnostic examination on teeth, dental arches and other structures in the oral cavity;

    (II) produce radiographs of jaws, parts of the skull and carpus for the purpose of cephalometric examination, if equipped with tele-radiographic arm (CEPH);

    (III) the production of tomographic images of the head, including the ear, nose and throat (ENT), of the dentomaxillofacial complex, teeth, mandible and maxilla, temporomandibular-articular joint (TMD), other areas of the human skull and neck with sections of the cervical spine for use in diagnostic support, if equipped with the CBCT option. The device is operated and used by physicians, dentists, x-ray technologists and other legally qualified professionals.

    Device Description

    The Proposed device is a panoramic, cephalometric and tomographic radiological system developed and manufactured by CEFLA S.C. The proposed device is a change of the predicate device: hyperion X9 pro, NewTom GIANO HR, X-RADIUS TRIO PLUS (K214084). Like the predicate device the proposed device can be sold under three different proprietary product name and brands for commercial needs, without changing any of the basic safety, essential performances and functional features:
    1 hyperion X9 pro myray CEFLA S.C.
    2 NewTom GIANO HR NewTom CEFLA S.C.
    3 X-RADIUS TRIO PLUS Castellini CEFLA S.C.
    Wherever the Proposed device is mentioned the device with its three different trade/ proprietary names: hyperion X9 pro, New Tom GiANO HR and X-RADiUS TRiO PLUS.
    Like the predicate device hyperion X9 pro, NewTom GIANO HR, X-RADIUS TRIO PLUS (K214084) the proposed device is equipped with X-ray tube generator and X-ray sensors (Solid State X-ray Imaging Detectors) for dental panoramic (PAN), cephalometric radiography (CEPH) and computed tomography (CBCT). The proposed device permits to acquire radiological images (panoramic images, cephalometric images and 3D volumes) at varying radiographic angles by rotating around the patient following different trajectories depending on the selected examination. The exposed area can be adapted to a specific region of interest to keep the radiation dose as low as possible for the patient. This is achieved by collimating the x-ray beam and the adjustment of starting and ending points of the x-ray source and sensor movement. Furthermore, the radiation dose can be adapted by various parameters such as examination types and exposure technique factors. Class I lasers pointers are utilized to define reference lines for the patient position. The patient stabilized through adjustable patient supports, can sit or stand. Control panel allows user actions as: height adjustment, selection of examination, and exposure parameters and delivers information about the unit status. The obtained digital image data are processed to provide a reconstructed image. The images are transferred to a computer, in real time or later depending on the needs and choice of the operator. The software used to manage the images, essential for CBCT acquisitions, is NNT/iRYS, a radiological imaging software developed by CEFLA S.C.

    AI/ML Overview

    This document describes a 510(k) premarket notification for a dental X-ray system. The key aspect of this submission is that the Proposed Device is a modification of an existing, legally marketed Predicate Device. The modifications primarily involve the availability of alternative 2D X-ray detectors. Therefore, the study presented here focuses on demonstrating substantial equivalence to the predicate device, rather than proving novel performance against a clinical ground truth.

    1. Acceptance Criteria and Reported Device Performance

    Since this is a submission demonstrating substantial equivalence to an existing device, the "acceptance criteria" are predominantly the adherence to various medical device standards and the maintenance of comparable performance to the predicate device. The "reported device performance" is largely framed in terms of "no significant difference" or "comparable" to the predicate, as the goal is to show the new version is as safe and effective as the old.

    Acceptance Criteria (Demonstrated Equivalence)Reported Device Performance
    Device Name remains the samehyperion X9 pro, NewTom GiANO HR, X-RADIUS TRIO PLUS (Identical)
    Manufacturer remains the sameCEFLA S.C. (Identical)
    External appearances and materials are identical"The external appearances and materials between Proposed device and Predicate device are identical."
    Regulation Number (21 CFR 892.1750) is the same"No difference." (Identical)
    Regulatory Class (Class II) is the same"No difference." (Identical)
    Classification Product Codes (OAS, MUH) are the same"No difference." (Identical)
    Classification Name (Computed Tomography X-ray System) is the same"No difference." (Identical)
    Indications for Use are identical"No difference." (Identical to predicate: orthopanoramic images, cephalometric examination, tomographic images of head/ENT/dentomaxillofacial/TMJ, etc.)
    Performance specifications (Panoramic, Computed tomography, Cephalometric) are the same"No difference." (Identical)
    Patient population (Adult, Pediatric) is the same"No difference." (Identical)
    Exposure selectable modes (2D: PAN, BTW, DENT, SIN, TMJ, CEPH) are the same"No difference." (Identical)
    Rated input is the same"No difference." (Identical: 20A @ 115V~, 12A @ 240V~, 50/60 Hz)
    X-Ray emission parameters (Tube voltage, Tube current range, Exposure Time range, Shape of X-Ray Beam, Focal spot size, Anode Inclination, Collimator) are the same."No difference." (Identical across all listed parameters)
    FOV (3D) is the same"No difference." (Identical: Max: 16x18 cm, min: 4x4 cm)
    Total filtration for scansions is the same"No difference." (Identical: 2D > 2,5 mm Al @85kV, 3D 6.5 mm Al @ 90 kV)
    2D operating modes MIN and MAX Dose Area Product (DAP) are comparable"No significant difference. The Proposed Device showed comparable measured DAP values than Predicate Device for the same selected exams." (Slight variations, e.g., 12 mGycm2 vs 11 mGycm2 for CEPH Lat Short, 137 mGycm2 vs 136 mGycm2 for TMJ Lat)
    Image X-ray sensors Technology for 2D and 3D imaging are comparable or identical"No significant difference." (Both use CMOS with scintillator and Direct conversion CMOS for 2D, and Amorphous Silicon Flat Panel for 3D.)
    Image X-ray sensors dimensions for 2D are documented and justified if different; 3D dimensions are identical.For 2D CMOS with scintillator, new alternative sensors are slightly higher/wider (e.g., PAN: 152mm x 6.7mm vs 148mm x 6mm). Justified as "The new alternative X-ray sensors with scintillator available with Proposed Device are higher and wider than the X-ray sensors with scintillator available with the Predicate Device, however the Proposed Device uses the identical beam limiting system used by the Predicate Device. The correspondence between X-ray field and effective image reception area is conforming the same recognized consensus standard IEC 60601-2- 63 applied to both Proposed Device and Predicate Device thus it doesn't involve in different safety considerations." For Direct conversion CMOS and 3D, identical dimensions.
    Image X-ray sensors Pixel size for 2D are comparable or better; 3D are identical."No significant difference between Proposed Device and Predicate Device. The 2D sensor pixel sizes are comparable between Proposed Device and Predicate Device. The pixel sizes of new alternative CMOS detectors with scintillator is slight better than the pixel size of the CMOS detectors with scintillator already available with Predicate Device because smaller pixel size can theoretically allow to obtain higher resolution." For 3D, identical pixel size (127x127 μm).
    Source to image X-ray sensor distance (SID) is the same"No difference between Proposed Device and Predicate Device. The two devices share the same mechanical structure." (Identical)
    Laser pointers optical class is the same"No difference." (Identical: Class 1 according to IEC 60825-1:2014)
    Number of fixing points of craniostat and cephalostat are the same"No difference." (Identical: 6 adjustable for craniostat, 3 adjustable for cephalostat)
    Control software (Firmware) changes properly managed and validated"The firmware on board has been updated to manage also the new alternative 2D X-ray sensors models. The changes have been managed according to the same recognized consensus IEC 62304 and FDA Guidance on Medical Device Software."
    Graphical User Interface (GUI) is comparable"No significant differences." (Identical: VKB)
    Viewing & Reconstruction software changes properly managed and validated"The addition of alternative 2D X-ray sensors doesn't require significant change of Viewing & Reconstruction software however specific configuration data has been added to manage the new 2D X-ray Scintillator sensor models. Software changes have been managed according to the same recognized consensus standard IEC 62304 and FDA Guidance on Medical Device Software."
    Software validation according to IEC 62304 + FDA Guidance on MD SW"No difference." (Identical)
    Electrical safety complies with IEC 60601-1: 2012"No difference." (Complies with IEC 60601-1: 2012)
    Electromagnetic compatibility complies with IEC 60601-1-2:2014"No difference." (Complies with IEC 60601-1-2:2014)
    Other relevant standards are met (e.g., IEC 60825-1, IEC 60601-1-3, IEC 62366, IEC 60601-2-63, IEC 60601-1-6)Performance Tests included in this premarket notification verify the conformity of the proposed device with the requirements of these standards.
    Image quality is confirmed to be equivalent to the predicate device."Verification activities for confirmation of the image quality of the proposed device has been performed. The results of the image quality review have demonstrated that the device is substantially equivalent to the predicate device."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a traditional "test set" sample size in terms of clinical cases or patient images. This submission relies on a comparison against the predicate device's established performance and adherence to engineering and safety standards. The testing described is primarily non-clinical performance testing to verify conformity with various IEC standards (electrical safety, EMC, radiation protection, software lifecycle, usability, dental x-ray specific requirements).

    • Sample Size: Not specified in terms of clinical images or patient data, as the study is not based on a new clinical performance claim but on equivalence to a predicate. It indicates "Verification activities for confirmation of the image quality of the proposed device has been performed." This implies technical tests on samples, phantoms, or test patterns, rather than a large human-read image dataset.
    • Data Provenance: Not explicitly stated for specific test data, but the manufacturer is CEFLA S.C. based in Imola, Italy. The testing would have been conducted as part of their device development and validation processes, likely at their facilities or certified labs. The study is a non-clinical validation comparing the proposed device's technical specifications and test results to those of the predicate device and relevant standards. It is retrospective in the sense that it relies on existing data/standards for the predicate and engineering tests for the new device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Experts

    This type of submission (510(k) for substantial equivalence based on non-clinical performance and direct comparison to a predicate) typically does not involve establishing ground truth through multiple expert clinical readers for a test set. The "ground truth" here is the established safety and performance of the predicate device and the adherence to relevant international standards.

    If image quality was assessed, it would likely be through technical metrics (e.g., MTF, contrast, noise) rather than subjective clinical readings by multiple radiologists. The document references "image quality review," which could involve internal engineering and quality experts. No specific number or qualifications of experts for image review are provided in this summary.

    4. Adjudication Method for the Test Set

    Not applicable for this type of submission. There is no multi-reader, multi-case study or a need for clinical adjudication to establish ground truth for a diagnostic accuracy claim. The comparison is based on engineering specifications and adherence to standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, a multi-reader, multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states: "Given the differences from the predicate device, no human clinical studies have been considered necessary to support substantial equivalence." This means the submission relies on non-clinical performance data and a direct comparison of features and technical specifications to the legally marketed predicate device.

    Therefore, there is no effect size of how much human readers improve with AI vs. without AI assistance, as AI assistance is not the subject of this submission, nor is a human-in-the-loop performance study.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This device is an X-ray imaging system, not an AI software algorithm. Therefore, "standalone" algorithm performance testing is not applicable in this context. The core function is image acquisition, not algorithmic analysis of images.

    7. The Type of Ground Truth Used

    The "ground truth" for this submission is implicitly the already established and FDA-cleared performance and safety of the predicate device (K214084), combined with compliance with relevant international and national standards (e.g., IEC 60601 series, IEC 62304). The study aims to demonstrate that the small changes introduced do not alter this established 'ground truth' of safety and fundamental performance. It is not based on expert consensus for disease detection, pathology, or outcomes data.

    8. The Sample Size for the Training Set

    Not applicable. This is a hardware device (X-ray system), not an AI algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no AI algorithm training set involved.

    Ask a Question

    Ask a specific question about this device

    K Number
    K214084

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-01-20

    (24 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    hyperion X9 pro, NewTom GiANO HR, X-RADiUS TRiO PLUS is digital panoramic, cephalometric and tomographic extra-oral system, intended to:

    (1) produce orthopanoramic images of the maxillofacial region and carry out diagnostic examination on teeth, dental arches and other structures in the oral cavity;

    (II) produce radiographs of jaws, parts of the skull and carpus for the purpose of cephalometric examination, if equipped with tele-radiographic arm (CEPH);

    (III) the production of tomographic images of the head, including the ear, nose and throat (ENT), of the dentomaxillofacial complex, teeth, mandible and maxilla, temporomandibular-articular joint (TMI), other areas of the human skull and neck with sections of the cervical spine for use in diagnostic support, if equipped with the CBCT option. The device is operated and used by physicians, dentists, x-ray technologists and other legally qualified professionals.

    Device Description

    The Proposed device is a panoramic, cephalometric andiological system developed and manufactured by CEFLA S.C. The proposed device is a change of the predicate device: hyperion X9 pro (K190496). Like the predicate device the proposed device can be sold under three different product name and brands for commercial needs, without changing any of the basic safety, essential performances and functional features:

    Like the predicate device hyperion X9 pro (K190496) the proposed device is equipped with X-ray tube generator and X-ray sensors (solid state X-ray imaging detectors) for dental panoramic (PAN), cephalometric radiography (CEPH) and cone beam computed tomography (CBCT). The proposed device permits to acquire radiological images (panoramic images, cephalometric images and 3D volumes) at varying radiographic angles by rotating around the patient following different trajectories depending on the selected examination. The exposed area can be adapted to a specific region of interest to keep the radiation dose as low as possible for the patient. This is achieved by collimating the x-ray beam and the adjustment of starting and ending points of the x-ray source and sensor movement. Furthermore, the radiation dose can be adapted by various parameters such as examination types and exposure technique factors. Class I lasers pointers are utilized to define reference lines for the patient position. The patient, stabilized through adjustable patient supports, can sit or stand. Control panel allows user actions as: height adjustment, selection of examination, and exposure parameters and delivers information about the unit status. The obtained digital image data are processed to provide a reconstructed image. The images are transferred to a computer, in real time or later depending on the needs and choice of the operator. The software used to manage the images, essential for CBCT acquisitions, is NNT/iRYS, a radiological imaging software developed by CEFLA S.C.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a dental X-ray system, comparing a proposed device (hyperion X9 pro, NewTom GiANO HR, X-RADiUS TRiO PLUS) to a predicate device (hyperion X9 pro, K190496). The submission focuses on demonstrating substantial equivalence, particularly related to the introduction of new 2D X-ray sensors.

    Based on the provided text, the device in question is not an AI-powered device; it is a general dental X-ray imaging system. Therefore, the detailed acceptance criteria and study design elements typically associated with AI/ML-based medical devices (such as MRMC studies, sample sizes for test/training sets, expert adjudication, or ground truth establishment for AI performance) are not applicable or reported in this document.

    The "acceptance criteria" discussed in this document relate to the performance and safety standards that the medical device must meet to demonstrate substantial equivalence to a legally marketed predicate device. The text does not detail specific performance thresholds or how the device as an AI would be evaluated. Instead, it confirms compliance with general medical device standards.

    Here's a breakdown of what is available and what is not for the requested points, given that this is not an AI device:

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria (as implied by the document for a general medical device): Compliance with recognized consensus standards for medical electrical equipment, radiation protection, and software.
      • IEC 60601-1: Medical electrical equipment (basic safety and essential performance)
      • IEC 60601-1-2: Electromagnetic compatibility
      • IEC 60825-1: Safety of laser products
      • IEC 60601-1-3: Radiation protection in diagnostic X-ray equipment
      • IEC 62366: Application of usability engineering
      • IEC 62304: Medical device software Software lifecycle processes
      • IEC 60601-2-63: Dental extra-oral X-ray equipment (particular requirements)
      • IEC 60601-1-6: Usability
      • Demonstration of equivalent or lower radiation dose (DAP values).
      • Demonstration of equivalent image quality.
    • Reported Device Performance (against these criteria):
      • Radiation Dose: The Proposed Device showed "similar or lower measured DAP values than Predicate Device for the same selected exams."
        • Proposed Device (Sensors with scintillator): 11 mGy*cm2 to 126 mGy*cm2
        • Proposed Device (Direct conversion sensors): 7 mGy*cm2 to 103 mGy*cm2
        • Predicate Device (Sensor with scintillator): 13.01 mGy*cm2 to 229.81 mGy*cm2
      • Mechanical & Technical Features: Largely identical or with justified differences that do not negatively affect safety or performance (e.g., new direct conversion sensors, updated firmware/software to manage new sensors, but conforming to standards).
      • Image Quality: "The results of the image quality review have demonstrated that the device is substantially equivalent to the predicate device." (No specific metrics or quantitative values are provided for image quality beyond this statement).
      • Software Validation: Updated firmware and viewing software for new sensors were managed according to IEC 62304 and FDA Guidance on Medical Device Software.
      • Electrical Safety & EMC: Complies with IEC 60601-1:2012 and IEC 60601-1-2:2014.

    2. Sample size used for the test set and the data provenance:

    • Not applicable / Not specified for AI performance. This document describes non-clinical performance testing (e.g., electrical safety, EMC, radiation output, image quality review) of the X-ray system itself, not a test set for an AI algorithm.
    • The document states "The results of the Non-clinical performance testing support substantial equivalence." No specific "test set" sample size in the context of diagnostic performance (e.g., number of patient images) is mentioned.
    • Data Provenance: Not explicitly stated, but the testing would likely have been conducted by the manufacturer (CEFLA S.C.) in Italy, given their location. The nature of the testing (bench testing, phantom imaging, etc.) suggests it's not patient data for an algorithm.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable. There is no mention of a "test set" requiring expert-established ground truth for diagnostic performance, as this is not an AI device. The "image quality review" is mentioned, but without details on experts or methodology.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. No adjudication method is mentioned as this is not an AI device or a study involving human reader performance evaluation in a diagnostic context.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. The document explicitly states: "Given the differences from the predicate device, no human clinical studies have been considered necessary to support substantial equivalence." Therefore, no MRMC study was performed, and no AI assistance is part of this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device does not feature a standalone algorithm in the context of AI. It is an X-ray imaging system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable in the AI context. Ground truth in this submission pertains to the physical and technical specifications of the device, confirmed through engineering measurements, compliance with standards, and comparison to the predicate device's established performance parameters. For image quality, it's inferred that the review was against expected image characteristics for such a system, rather than diagnostic outcomes.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device, so no training set for an algorithm is mentioned or relevant.

    9. How the ground truth for the training set was established:

    • Not applicable. As above, no training set for an AI/ML algorithm is involved.
    Ask a Question

    Ask a specific question about this device

    K Number
    K190496

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-04-24

    (55 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    hyperion X9 pro, NewTom Giano HR, X-RADiUS TRiO PLUS is a digital panoramic, cephalometric and tomographic extra-oral X-ray system, intended to:

    (1) produce orthopanoramic images of the maxillofacial region and carry out diagnostic examination on teeth, dental arches and other structures in the oral cavity;

    (II) produce radiographs of jaws, parts of the skull and carpus for the purpose of cephalometric examination, when equipped with tele-radiographic arm (CEPH):

    (III) the production of tomographic images of the head, including the ear, nose and throat (ENT), of the dento-maxillofacial complex, teeth, mandible and maxilla, temporomandibular-articular joint (TMI), other areas of the human skull and neck with sections of the cervical spine for use in diagnostic support, if equipped with the CBCT option.

    Device Description

    Not Found

    AI/ML Overview

    This FDA clearance letter does not contain the detailed information required to describe the acceptance criteria and the study that proves the device meets those criteria. The letter primarily states that the device is substantially equivalent to legally marketed predicate devices and outlines the indications for use.

    Specifically, the following information is not provided in the given text:

    • A table of acceptance criteria and the reported device performance: This document does not include any performance metrics, thresholds, or a direct comparison to acceptance criteria.
    • Sample size used for the test set and the data provenance: There is no mention of the number of cases or patients used for testing, nor the origin (country, retrospective/prospective) of any data.
    • Number of experts used to establish the ground truth for the test set and their qualifications: The document does not discuss expert involvement in ground truth establishment.
    • Adjudication method: There is no information regarding how disagreements among experts (if any were used) were resolved.
    • Multi-reader multi-case (MRMC) comparative effectiveness study: The document does not mention whether such a study was performed or any effect size related to AI assistance.
    • Standalone (algorithm-only) performance: While the device is an X-ray system, it's not an AI-driven diagnostic algorithm whose standalone performance would typically be reported.
    • Type of ground truth used: The document does not specify if expert consensus, pathology, or outcomes data was used for validation.
    • Sample size for the training set: As there's no mention of an AI algorithm or a training process, this information is absent.
    • How the ground truth for the training set was established: Similarly, this information is not applicable or provided.

    The document indicates that the devices (hyperion X9 pro, NewTom GiANO HR, X-RADiUS TRiO PLUS) are digital panoramic, cephalometric, and tomographic extra-oral X-ray systems, and it describes their intended uses, such as producing images of the maxillofacial region, jaws, skull, and for cephalometric and tomographic examinations. The substantial equivalence determination is based on the comparison to predicate devices, but the specific performance data for this equivalence is not detailed in this FDA letter.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1