Found 2 results

510(k) Data Aggregation

    K Number
    K252284

    Validate with FDA (Live)

    Device Name
    eMurmur Heart AI
    Manufacturer
    Date Cleared
    2025-12-19

    (150 days)

    Product Code
    Regulation Number
    870.1875
    Age Range
    0 - 150
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The eMurmur Heart AI software is a decision support system for the automated evaluation of recorded patient heart sounds. It identifies the first (S1) and second (S2) heart sounds, the presence of abnormal heart murmurs, and the absence of abnormal heart murmurs, with the latter including cases of no murmurs and innocent murmurs. eMurmur Heart AI also determines the average loudness of individual heart sound components, including S1, systole, S2, and diastole, as well as their respective loudness ratios.

    eMurmur Heart AI is indicated for use in settings where auscultation can be performed properly by a healthcare provider. It is not intended as a sole means of diagnosis. The heart sound interpretations offered by eMurmur Heart AI are only significant when considered in conjunction with healthcare provider over-read and including all other relevant patient data. eMurmur Heart AI is intended for use on pediatric and adult patients.

    Device Description

    To analyze heart sounds using eMurmur Heart AI, a digital recording of a patient's heart sounds is required. Recordings are made using any supported digital stethoscope, connected to a front-end client. The recorded auscultation data are transmitted from the front-end client to the eMurmur backend, which hosts eMurmur Heart AI. After analysis by eMurmur Heart AI, the results are returned to the front-end client where they are displayed to the user. The user can utilize the eMurmur Heart AI results to support their decision-making process regarding the potential presence of an abnormal heart murmur, and to assist them when investigating potential changes among longitudinally recorded heart sounds.

    eMurmur is a multiple function software platform which includes the eMurmur apps and eMurmur web portal. The platform is used to stream, record, display, replay, and store auscultation data, recorded by means of supported digital stethoscopes, i.e., functions that are not subject to regulatory oversight.

    The eMurmur multiple function software platform also has functions subject to FDA premarket review, i.e., eMurmur Heart AI. For this application, FDA assessed those functions only to the extent that they could adversely impact the safety and effectiveness of the functions subject to FDA premarket review.

    AI/ML Overview

    Here is a summary of the acceptance criteria and study details for the eMurmur Heart AI (2.2) device, based on the provided FDA 510(k) clearance letter:

    Acceptance Criteria and Device Performance

    Acceptance Criteria (Non-inferiority to Predicate)eMurmur Heart AI (2.2) Reported Performance (95% CI)eMurmur Heart AI (1.0) (Predicate) Reported Performance (95% CI)
    Non-inferior Sensitivity90.0% (78.8%-95.9%)86.7% (74.9%-93.7%)
    Non-inferior Specificity90.0% (78.8%-95.9%)88.3% (76.8%-94.8%)
    Non-inferior Accuracy90.0% (82.8%-94.5%)87.5% (79.9%-92.6%)

    The study's goal was to demonstrate non-inferiority of eMurmur Heart AI (2.2) compared to its predicate (eMurmur Heart AI (1.0)) for sensitivity, specificity, and accuracy in identifying abnormal heart murmurs. Non-inferiority was established using confidence intervals for paired data with continuity correction for the true difference in sensitivity or specificity. The results shown above indicate that eMurmur Heart AI (2.2) met the non-inferiority criteria relative to the predicate device.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • The exact sample size for the test set is not explicitly stated as a number of recordings. It is described as the "dataset used to validate and obtain clearance for the predicate device augmented with new, previously unseen data."
      • Data provenance: Not explicitly stated (e.g., country of origin). The data included heart sound recordings from both pediatric and adult subjects.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Number of experts: Three
      • Qualifications of experts: Board-certified cardiologists.
    3. Adjudication method for the test set:

      • Method: Majority consensus. The ground truth classification ("abnormal murmur present" or "healthy") was determined by three board-certified cardiologists based on majority consensus.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study involving human readers and AI assistance was mentioned in the provided document for the primary validation of eMurmur Heart AI (2.2). The comparison was strictly between the eMurmur Heart AI (2.2) algorithm and the predicate algorithm (2.1).
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Yes, a standalone (algorithm only) performance study was conducted. The "Performance Data" section directly reports the sensitivity, specificity, and accuracy of "eMurmur Heart AI (2.2)" and "eMurmur Heart AI (1.0)" (predicate) based on their automated analysis against the established ground truth.
    6. The type of ground truth used:

      • Ground truth type: Expert consensus. Specifically, the "ground truth classification was determined by three board-certified cardiologists based on majority consensus."
    7. The sample size for the training set:

      • The sample size for the training set is not explicitly stated in the provided document. The document mentions that the test set "was not used during training of eMurmur Heart AI (2.2)," implying a separate training set was used.
    8. How the ground truth for the training set was established:

      • The method for establishing ground truth for the training set is not explicitly detailed in the provided document. It is only stated that the test set was not used for training. Given the approach for the test set, it is plausible that a similar expert consensus method was used, but this is not confirmed.
    Ask a Question

    Ask a specific question about this device

    K Number
    K220766

    Validate with FDA (Live)

    Device Name
    eMurmur Heart AI
    Manufacturer
    Date Cleared
    2022-05-31

    (76 days)

    Product Code
    Regulation Number
    870.1875
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 'eMurmur Heart AI' software is a decision support system in the evaluation of recorded patient heart sounds. The automated analysis by eMurmur Heart Al identifies specific heart sounds that may be present, including S1, S2, physiological heart murmurs, pathological heart murmurs and the absence of a heart murmur.

    eMurmur Heart AI is indicated for use in a setting where auscultation would typically be performed by a healthcare provider. It is not intended as a sole means of diagnosis. The heart sound interpretations offered by eMurmur Heart AI are only significant when considered in conjunction with healthcare provider over-read and including all other relevant patient data.

    Device Description

    To analyze heart sounds via the eMurmur Heart AI, a digital recording of a patient's heart sounds is required. Recordings are made using a supported digital stethoscope, connected to a front-end client like, e.g., the eMurmur app or the eMurmur web portal. The recorded auscultation data are transmitted from the front-end client to the eMurmur backend, which hosts the eMurmur Heart Al. After analysis by the eMurmur Heart Al, the results of the analysis are returned to the front-end client where they are displayed to the user. The user can utilize the eMurmur Heart AI results to support their decision-making process regarding the potential presence and type of a heart murmur.

    eMurmur is a non-medical device software platform which includes the eMurmur backend, eMurmur apps and eMurmur web portal. The platform is used to stream, record, display. replay, and store auscultation data, recorded by means of supported digital stethoscopes.

    The eMurmur software platform has functions subject to FDA premarket review, i.e., eMurmur Heart AI, as well as functions that are not subject to FDA premarket review. For this application. FDA assessed those functions only to the extent that they could adversely impact the safety and effectiveness of the functions subject to FDA premarket review.

    AI/ML Overview

    The provided text describes the eMurmur Heart AI device, but it mainly focuses on its substantial equivalence to a predicate device (eMurmur ID) and states that no new performance data was required because the core algorithm is the same. Therefore, the information needed to fully answer your request regarding acceptance criteria and a new study proving the device meets those criteria is not explicitly present for eMurmur Heart AI.

    However, based on the information provided, we can infer some details and present the "Clinical Performance" data listed for the predicate device, eMurmur ID, as it's stated that eMurmur Heart AI shares the same heart sound analysis algorithm and thus identical clinical performance.

    Here's the breakdown of the information that can be extracted or reasonably inferred from the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" but lists "Clinical Performance" metrics for the "eMurmur ID" predicate device, which are then declared identical for eMurmur Heart AI. These would serve as the de facto performance metrics considered acceptable for substantial equivalence.

    Performance MetricAcceptance Criteria (from predicate)Reported Device Performance (eMurmur Heart AI, inherited from predicate)
    SensitivityN/A (Inherited from predicate)85.0% (72.9%-92.5%)
    SpecificityN/A (Inherited from predicate)86.7% (74.9%-93.7%)

    Note: The document explicitly states: "eMurmur Heart AI and the predicate, eMurmur ID (K181988), utilize the same heart sound analysis algorithm, hence no new performance data is required." This implies that the performance established for eMurmur ID is directly applicable and "accepted" for eMurmur Heart AI.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not specify the sample size used for the test set for either eMurmur Heart AI or the predicate eMurmur ID, nor does it mention data provenance (country of origin, retrospective/prospective). This information would typically be detailed in the original 510(k) submission for eMurmur ID (K181988).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The provided text does not specify the number of experts used or their qualifications for establishing ground truth for the test set of the predicate eMurmur ID.

    4. Adjudication Method for the Test Set

    The provided text does not specify the adjudication method used for the test set of the predicate eMurmur ID.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The provided text does not mention a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it discuss the effect size of human readers improving with AI assistance.

    6. Standalone Performance

    Yes, a standalone performance study was done based on the predicate device (eMurmur ID). The listed Sensitivity and Specificity (85.0% and 86.7% respectively) represent the algorithm's performance without explicit mention of human-in-the-loop assistance during the reported performance evaluation. The device is described as a "decision support system," emphasizing its role in aiding healthcare providers.

    7. Type of Ground Truth Used

    The type of ground truth used is not explicitly stated in this document. It mentions the algorithm identifies specific heart sounds, including physiological and pathological murmurs. Typically, for such devices, ground truth would be established by expert consensus based on clinical examination, potentially complemented by other diagnostic tests (e.g., echocardiography) for heart murmur validation.

    8. Sample Size for the Training Set

    The provided text does not specify the sample size for the training set.

    9. How the Ground Truth for the Training Set Was Established

    The provided text does not specify how the ground truth for the training set was established.

    Summary of what's missing and why:

    The core reason for much of the missing information is that this document is a 510(k) summary for "eMurmur Heart AI" which leverages the substantial equivalence pathway. It argues that because "eMurmur Heart AI" uses the same heart sound analysis algorithm as the already cleared "eMurmur ID," no new performance data is required. Therefore, the detailed study design, sample sizes, expert qualifications, and ground truth methodologies would have been part of the original 510(k) submission for "eMurmur ID" (K181988), not explicitly reiterated in this document beyond the summary performance metrics.