510(k) Data Aggregation

    K Number
    K192803

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-04-29

    (212 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System, Model IS4000 and da Vinci X Surgical System, Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors. electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    The modifications to the da Vinci Xi and X Surgical Systems adds a hardware component called the Advanced Processor (AP4000) as well software components, both on the Advanced Processor and a separate iOS app, to the surgical system.

    These modifications enable the following functionality: Video recording/Image capture, WiFi pairing of iOS devices with the Advanced Processor, an iOS app to view live endoscopic video once the iOS device has been paired and data logging on the AP4000 hard disk (stereo endoscopic video, kinematics and real time events). This allows surgeons to initiate and control video recordings as well as capture endoscopic images. These modifications also allow OR staff and others present in the OR to pair their iOS mobile devices with the Advanced Processor in order to view the live endoscopic video.

    AI/ML Overview

    This FDA 510(k) summary describes modifications to the da Vinci Xi and X Surgical Systems, not a new AI/ML medical device. Therefore, much of the requested information regarding AI/ML device studies and acceptance criteria is not applicable to this document. The document focuses on regulatory clearance for hardware and software modifications that enhance existing functionalities of a robotic surgical system, rather than an AI-driven diagnostic or treatment recommendation system requiring clinical performance studies against a ground truth.

    Specifically, the modifications enable:

    • Video recording/Image capture.
    • WiFi pairing of iOS devices with an Advanced Processor.
    • An iOS app to view live endoscopic video.
    • Data logging (stereo endoscopic video, kinematics, real-time events).

    The submission asserts that these modifications do not affect the substantial equivalence of the device to its predicate because "verification and validation testing have established there are no new issues of safety or effectiveness."

    Here's an attempt to answer the questions based on the provided document, highlighting what is not applicable due to the nature of the submission:

    1. A table of acceptance criteria and the reported device performance

    Since this is a submission for modifications to an existing device, and not a new diagnostic or prognostic AI/ML device, there are no specific clinical performance acceptance criteria (e.g., sensitivity, specificity, accuracy) reported in this document. The "acceptance criteria" for these modifications appear to be successful completion of verification and validation testing, and compliance with relevant standards.

    Acceptance Criteria (Implied)Reported Device Performance
    All required specifications met"The subject device met all required specifications and functioned as intended."
    Functioned as intended"The modifications enable the following functionality: Video recording/Image capture, WiFi pairing of iOS devices with the Advanced Processor, an iOS app to view live endoscopic video once the iOS device has been paired and data logging on the AP4000 hard disk (stereo endoscopic video, kinematics and real time events)." and "These modifications do not affect the substantial equivalence of the subject device to the predicate as verification and validation testing have established there are no new issues of safety or effectiveness."
    Safety and performance evaluated/verified"Safety and performance of the da Vinci Xi/X Surgical System has been evaluated and verified in accordance with software specifications and applicable performance standards through software verification and validation testing."
    Compliance with IEC 62304:2006/AC:2015"software validation activities were performed in accordance with IEC 62304:2006/AC: 2015- Medical device software - Software life cycle processes"
    Compliance with FDA Guidance for Software in Medical Devices"in addition to the FDA Guidance documents, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submission for Management of Cybersecurity in Medical Devices.""
    Compliance with EMC, Electrical Safety, and Wireless Coexistence"electromagnetic compatibility testing (in accordance with ISO 60601-1-2:2015 / IEC 60601-1-2:2014), electrical safety testing (in accordance with IEC 60601-1: 2012) and wireless coexistence testing were performed."
    No new questions of safety or effectiveness"The da Vinci Xi/X Surgical System raises no new questions of safety or effectiveness."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable as the document describes verification and validation testing of software and hardware modifications for a robotic surgical system, not a clinical study involving a test set of patient data for an AI/ML diagnostic or prognostic algorithm. The testing would involve engineering and software validation tests, not clinical data sets in the typical sense.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. Ground truth establishment by experts is relevant for AI/ML algorithms performing tasks like diagnosis or detection. The current submission focuses on verifying the functionality and safety of video recording, streaming, and data logging features.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable for the same reasons as #3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable. The device provides tools for surgeons (video recording, streaming), but it is not an AI assistant meant to improve human reader performance in interpreting medical images or data.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The modifications are for a human-controlled surgical system, enhancing its capabilities rather than introducing a standalone AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not applicable. The "ground truth" for the verification and validation of these modifications would be the functional specifications and safety requirements met through engineering and software testing.

    8. The sample size for the training set

    This information is not applicable. This document does not pertain to the development or validation of an AI/ML algorithm that would require a training set.

    9. How the ground truth for the training set was established

    This information is not applicable. As no AI/ML training set is mentioned or implied, the method for establishing its ground truth is irrelevant to this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191529

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-02-06

    (241 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System, Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    The subject of this 510(k) submission are software modifications to the da Vinci Xi Surgical System (Model IS4000) and da Vinci X Surgical System (Model IS4200). These modifications allow Intuitive Surgical to offer cloud-based features to the IS4000 and IS4200 systems such as Surgeon Cloud Accounts, Procedure List, and Central Configuration.

    • . Surgeon Cloud Accounts – Enable surgeons to log in to multiple IS4000 systems and access their personal IS4000 system settings (vision, ergonomic, motion scaling, and energy) by storing user preferences in a cloud-based central Intuitive data server.
    • . Procedure List – Enable surgeons/OR staff to select a surgical procedure they intend to perform from the surgeon console touchscreen before the start of surgery.
    • . Central Configuration – Enable authorized Intuitive Surgical employees to remotely configure features of one or more IS4000 systems simultaneously.
    AI/ML Overview

    The provided text describes a 510(k) premarket notification for software modifications to the da Vinci Xi Surgical System and da Vinci X Surgical System. The modifications allow for cloud-based features such as Surgeon Cloud Accounts, Procedure List, and Central Configuration.

    However, the document does not contain any information regarding clinical performance studies, acceptance criteria, sample sizes for test or training sets, expert review processes for ground truth establishment, or multi-reader multi-case (MRMC) studies.

    The "Performance Data" section explicitly states:

    "Using an ISO 14971 compliant risk management process, a comprehensive hazard analysis of the software was performed, and impacted risks assessed, mitigations were evaluated and tested to be acceptable through verification and validation testing. The software development process based on ISO 62304 ensured that the software changes were appropriately documented, risk assessed and tested. Design verification tests were performed to verify the proposed software modifications performed as intended. The testing also verified the modifications to the da Vinci system software and server infrastructure did not raise new questions of safety and effectiveness. In addition, the pre and postmarket FDA guidance documents regarding cybersecurity were followed to document, risk assess and verify that the cybersecurity controls performed as expected in the event of an exploitation of a cybersecurity vulnerability. Performance test data demonstrate that the subject device is substantially equivalent to the predicate device and that the design output meets the design input requirements. None of the software modifications impact the intended use or indications for use of the da Vinci X/Xi Systems."

    This indicates that the performance data for this 510(k) submission primarily consists of design verification and validation testing, hazard analysis, and cybersecurity assessment, rather than a clinical study measuring diagnostic or treatment outcome performance against specific acceptance criteria. The claim of "substantial equivalence" is based on these engineering and software system evaluations, and the fact that the modifications do not alter the intended use of the device.

    Therefore, I cannot populate the requested table or answer most of the questions regarding acceptance criteria, study design, and ground truth, as this information is not present in the provided text.

    However, I can extract the following relevant non-clinical information:

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria: Based on the text, the acceptance criteria are implicitly tied to the successful completion of design verification and validation testing, hazard analysis per ISO 14971, and cybersecurity control verification per FDA guidance. The overall goal is to demonstrate that the software modifications perform as intended and do not raise new questions of safety and effectiveness, thus maintaining substantial equivalence to the predicate device.
    • Reported Device Performance:
      • Software modifications performed as intended.
      • Modifications did not raise new questions of safety and effectiveness.
      • Cybersecurity controls performed as expected.
      • Design output meets design input requirements.
      • No impact on the intended use or indications for use.

    2. Sample sized used for the test set and the data provenance:

    • Not applicable/Not provided. The "test set" here refers to engineering and software validation, not a clinical data set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable/Not provided. Ground truth in a clinical sense (e.g., diagnosis, pathology) was not established for this type of submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable/Not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was done. This submission is for software modifications that add cloud-based functionality, not for AI assistance in diagnosis or treatment.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable/Not provided. The device is a surgical system requiring human operation. The software modifications are infrastructure and user preference-related, not a standalone diagnostic algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not applicable. The "ground truth" for this submission would be defined by engineering specifications and functional requirements, against which the software was verified and validated.

    8. The sample size for the training set:

    • Not applicable/Not provided. No machine learning training set is mentioned in the context of this 510(k) for software modifications.

    9. How the ground truth for the training set was established:

    • Not applicable/Not provided.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182140

    Validate with FDA (Live)

    Date Cleared
    2018-10-24

    (78 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    This 510(k) is for a labeling modification only, to include the following additional representative, specific procedure under the cleared "general laparoscopic surgical procedures" Indication for Use of the da Vinci Xi Surgical System, Model IS4000 (K131861) and the da Vinci X Surgical System, Model IS4200 (K171294) and the associated labeling claims: Gastrectomy. There are no changes to the technological characteristics of the cleared da Vinci Xi or X Surgical Systems (Models IS4000 and IS4200) proposed in this submission. The da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200, are softwarecontrolled, electro-mechanical systems designed for surgeons to perform minimally invasive surgery. The Model IS4000 and Model IS4200 Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

    AI/ML Overview

    The provided document describes a 510(k) submission for a labeling modification of the da Vinci Xi and X Surgical Systems (Models IS4000 and IS4200) to include "Gastrectomy" as a recognized procedure under the existing "general laparoscopic surgical procedures" indication for use. This is explicitly stated as a labeling modification only, with no changes to the technological characteristics of the cleared devices. As such, the "acceptance criteria" and "device performance" in a traditional sense (e.g., diagnostic accuracy, sensitivity, specificity) for a novel device or AI algorithm are not applicable here.

    Instead, the study aims to demonstrate that adding "Gastrectomy" to the labeling does not introduce new safety or effectiveness concerns compared to the already cleared general laparoscopic indications, and that the da Vinci systems are substantially equivalent for this new labeled use.

    Here's a breakdown of the requested information based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a labeling modification and not a new device or AI algorithm with specific performance metrics like accuracy, sensitivity, or specificity, there are no predefined numerical acceptance criteria or typical "device performance" values.

    Instead, the "acceptance criteria" can be inferred as demonstration of comparable safety and effectiveness outcomes for Gastrectomy procedures performed with the da Vinci system when compared to laparoscopic and/or open procedures, based on literature review.

    Metric (Comparison Group)Acceptance Criteria (Inferred)Reported Device Performance Statement for da Vinci (vs. Laparoscopic)Reported Device Performance Statement for da Vinci (vs. Open)
    MortalityComparable mortality ratesComparable mortality ratesComparable mortality rates
    Estimated Blood Loss (EBL)Comparable or lower EBL volumesComparable or lower EBL volumesLower EBL volumes
    Blood Transfusion RatesComparable blood transfusion ratesComparable blood transfusion ratesComparable blood transfusion rates
    Lengths of Hospital Stay (LOS)Comparable or shorter lengths of hospital stayComparable or shorter lengths of hospital stayComparable or shorter lengths of hospital stay
    Intraoperative Complication RatesComparable intraoperative complication ratesComparable intraoperative complication ratesComparable intraoperative complication rates
    Postoperative Complication RatesComparable postoperative complication ratesComparable postoperative complication ratesComparable postoperative complication rates
    Conversion RatesComparable conversion ratesComparable conversion ratesNot explicitly stated for open, but relevant for comparing da Vinci to other minimally invasive approaches.
    Reoperation RatesComparable reoperation ratesComparable reoperation ratesComparable reoperation rates
    Readmission RatesComparable readmission ratesComparable readmission ratesComparable readmission rates
    Anastomotic Leak RatesComparable anastomotic leak ratesComparable anastomotic leak ratesComparable anastomotic leak rates
    Anastomotic Stenosis RatesComparable anastomotic stenosis ratesComparable anastomotic stenosis ratesComparable anastomotic stenosis rates
    Operative TimeComparable or longer operative times, without an increase in mortality or complication ratesComparable or longer operative times; this increase was not associated with an increase in mortality or complication rates.Longer operative times; this increase was not associated with an increase in mortality or complication rates.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size for the Test Set:
      The study is a literature review, not a primary clinical trial with a "test set" in the traditional sense of a prospectively collected cohort. The sample sizes are derived from the N values reported in the individual studies included in the meta-analysis/systematic review. For example, some studies reviewed had sample sizes for Robotic Gastrectomy ranging from 53 to 1830 patients, for Laparoscopic Gastrectomy from 61 to 4123 patients, and for Open Gastrectomy from 145 to 8585 patients.
    • Data Provenance: The data provenance is stated as "Published clinical data from literature." The document extensively cites 16 publications. No specific countries of origin are mentioned, but based on typical medical literature, these would likely be international. The studies cited include:
      • Two (2) randomized controlled trials (level 1b evidence)
      • Ten (10) systematic reviews/meta-analysis (level 2a evidence)
      • Four (4) database studies (level 2c evidence)
        The nature of these studies (e.g., retrospective database analyses, prospective RCTs within the included literature) would vary.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of study does not involve "experts establishing ground truth" for individual cases in a test set. The "ground truth" here is the aggregated clinical outcomes reported in published studies, which are considered evidence-based medical conclusions. The experts involved would be the authors of the 16 published clinical studies (surgeons, statisticians, researchers, etc.) who conducted the original research and established their findings. Their specific qualifications are not detailed in this 510(k) summary, but are implicitly assumed based on the peer-reviewed nature of the published literature.


    4. Adjudication Method for the Test Set

    No adjudication method is applicable here as this is a review of published literature, not a study involving individual case assessments by reviewers. The "adjudication" is inherent in the peer-review process of the published articles and the methodology of systematic reviews and meta-analyses.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This submission is for a surgical system, not an AI diagnostic/interpretive tool requiring human reader performance evaluation.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No standalone algorithm performance study was done. The da Vinci Xi and X Surgical Systems are electro-mechanical systems intended to assist surgeons (human-in-the-loop), not standalone algorithms. The submission explicitly states "The Intuitive Surgical Endoscopic Instrument Control System...is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments...It is intended to be used by trained physicians in an operating room environment..."


    7. The Type of Ground Truth Used

    The ground truth used is clinical outcomes data (e.g., mortality, EBL, LOS, complication rates, reoperation rates, etc.) collected and analyzed in previously published clinical studies (Randomized Controlled Trials, Systematic Reviews/Meta-analyses, and Database Studies).


    8. The Sample Size for the Training Set

    This submission does not involve a "training set" in the context of an AI/ML algorithm. The "training data" for making the determination about substantial equivalence for the new indication are the 16 published clinical studies. The sum of patients across all studies' da Vinci cohorts, laparoscopic cohorts, and open cohorts constitutes the entirety of the evidence base reviewed. These numbers are substantial, with individual studies reporting hundreds to thousands of patients (e.g., Robotic 1830, Lap 4123, Open 8585 in some studies).


    9. How the Ground Truth for the Training Set Was Established

    Since there is no "training set" in the AI/ML sense, the question refers to how the evidence for the labeling modification was established. The "ground truth" (i.e., the clinical outcomes and conclusions) in the selected 16 publications was established through:

    • Clinical Research Methodologies: Randomized controlled trials, systematic reviews, meta-analyses, and database studies.
    • Data Collection: Gathering patient data on various clinical outcomes (e.g., EBL, LOS, complications, mortality) from surgical procedures.
    • Statistical Analysis: Rigorous statistical methods applied by the authors of the original publications to analyze and interpret the collected data, leading to the reported conclusions regarding comparability or differences in outcomes between da Vinci, laparoscopic, and open surgical approaches.
    • Peer Review: The scientific rigor of these publications being accepted into peer-reviewed journals.
    Ask a Question

    Ask a specific question about this device

    K Number
    K173585

    Validate with FDA (Live)

    Date Cleared
    2018-07-19

    (241 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    This 510(k) is for a labeling modification only, to include "Ventral Hernia Repair" (VHR) procedures under the cleared "general laparoscopic surgical procedures" Indication for Use of the da Vinci Xi Surgical System, Model IS4000 and the da Vinci X Surgical System, Model IS4200. There are no changes to the technological characteristics of the cleared da Vinci Xi or X Surgical Systems (Models IS4000 and IS4200) proposed in this submission. The da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200, are software-controlled, electro-mechanical systems designed for surgeons to perform minimally invasive surgery. The Model IS4000 and Model IS4200 Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

    AI/ML Overview

    The provided document, K173585, describes the a 510(k) premarket notification for a labeling modification to include "Ventral Hernia Repair" (VHR) procedures under the cleared "general laparoscopic surgical procedures" Indication for Use of the da Vinci Xi Surgical System (Model IS4000) and the da Vinci X Surgical System (Model IS4200). The document does not describe the acceptance criteria and study proving a device's performance against acceptance criteria in the manner typically seen for AI/ML-enabled devices. Instead, it focuses on demonstrating the substantial equivalence of adding VHR procedures to the existing robotic surgical systems by comparing the outcomes of robotic-assisted VHR to traditional open and laparoscopic VHR.

    Therefore, many of the specific questions about acceptance criteria for an AI/ML device's performance, sample sizes for test sets, establishment of ground truth by experts, MRMC studies, and standalone algorithm performance, as requested in the prompt, are not applicable or directly addressed by this document. This document is about a surgical system, not an AI/ML diagnostic or prognostic tool.

    However, I can extract information related to the clinical study data used to support the labeling modification, which serves a similar purpose to a performance study for a new indication.

    Here's the information as it relates to the provided document:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't define explicit numerical "acceptance criteria" in the way one would for an AI/ML device (e.g., "sensitivity must be >X%"). Instead, it demonstrates comparability and, in some cases, superiority of the robotic-assisted VHR to existing surgical methods across several clinical outcomes and measures. The "acceptance" for this 510(k) is based on demonstrating substantial equivalence and safety and effectiveness for the new indication compared to existing methods.

    The tables provided (Tables 1, 2, 3, and 4a) present the "reported device performance" in terms of clinical outcomes for robotic-assisted VHR procedures compared to open and laparoscopic VHR. Below is a summary of the key findings, which serve as the "performance" data used to support the new indication.

    Summary of Device Performance (Robotic-Assisted VHR vs. Comparators)

    Outcome/MeasureRobotic-Assisted vs. Open (Non-Complex VHR)Robotic-Assisted vs. Laparoscopic (Non-Complex VHR)Robotic-Assisted vs. Open (Complex VHR)Robotic-Assisted vs. Laparoscopic (Complex VHR) - Unmatched AHSQC Data & Published Literature
    Length of StayComparable (2 ± 7 days vs. 2 ± 14 days)Shorter (2 ± 7 days vs. 4 ± 13 days)Shorter (2 ± 3 days vs. 5 ± 7 days)Shorter in AHSQC data (2 ± 3 days vs. 4 ± 3 days). Published literature varies (1-9.2 days).
    Intraoperative ComplicationsComparableComparableComparableComparable in AHSQC data (2% vs. 11%). Published literature (0-1.5%).
    Transfusions (Intra/Post-op)ComparableComparableComparableComparable (0% across all AHSQC cohorts). Published literature (0 or 51-91.7ml EBL).
    Postoperative Complications (through 30 days)Comparable (9% vs. 11%)Comparable (10% vs. 11%)Comparable (24% vs. 20%)Mixed: Robotic-assisted (24%) vs. AHSQC Laparoscopic (11%). Published literature (0-60%).
    Readmission Rates (through 30 days)Comparable (2% vs. 3%)Comparable (2% vs. 3%)Comparable (7% vs. 4%)Comparable in AHSQC data (7% vs. 7%). Published literature (0-13%).
    Re-encounter Rates (through 30 days)Lower clinic re-encounter for robotic-assisted (29% vs. 41%). Comparable ER.ComparableLower clinic re-encounter for robotic-assisted (16% vs. 27%). Comparable ER.Mixed: Robotic-assisted (17% Clinic, 5% ER) vs. AHSQC Laparoscopic (22% Clinic, 15% ER). Published literature (0).
    Reoperation Rates (through 30 days)Comparable (<1% vs. 1%)Comparable (<1% vs. 1%)Comparable (3% vs. 1%)Mixed: Robotic-assisted (3%) vs. AHSQC Laparoscopic (0%). Published literature (0-20%).
    Recurrence Rates (through 30 days)Comparable (1% vs. 0%)Comparable (1% vs. 1%)Comparable (1% vs. 1%)Comparable in AHSQC data (1% vs. 0%). Published literature "Not Reported" or low.
    Mortality (through 30 days)Comparable (<1% vs. 0%)Comparable (<1% vs. 0%)Comparable (<1% vs. 0%)Comparable in AHSQC data (<1% vs. 0%). Published literature (0-1.9%).
    Operative TimeLonger for robotic-assisted (e.g., 45% between 60-119 min vs. 35%).Longer for robotic-assisted (e.g., 43% between 60-119 min vs. 50%).Longer for robotic-assisted (e.g., 45% 240+ min vs. 17%).Longer for robotic-assisted (46% 240+ min vs. 19%). Published literature (125-372 min).
    Conversion Raten/an/an/aRobotic-assisted (4%) vs. AHSQC Laparoscopic (70%). Published literature (0%).

    Note: The document explicitly states that longer operative times for robotic-assisted cases were "not associated with increases in the complication, readmission, reoperation or mortality rates." This addresses a potential negative finding.

    2. Sample sizes used for the test set and the data provenance:

    • Test Set (Clinical Data): Real-world evidence (RWE) from the AHSQC registry database.
      • Non-Complex VHR Procedures:
        • Robotic-Assisted vs. Open: Propensity score matched cohorts of N=871 for each group.
        • Robotic-Assisted vs. Laparoscopic: Propensity score matched cohorts of N=615 for each group.
        • Data Provenance: The registry data for these procedures occurred between July 7, 2013, and January 1, 2017. The document does not specify the country of origin, but AHSQC (America's Hernia Society Quality Collaborative) suggests it is primarily from the United States. The data is retrospective as it's from a registry of past procedures.
      • Complex VHR Procedures:
        • Robotic-Assisted vs. Open: Propensity score matched cohorts of N=297 for each group.
        • Robotic-Assisted vs. Laparoscopic: Unmatched data from the AHSQC Registry Database (Robotic-Assisted N=305, Laparoscopic N=27) and a summary of published literature of laparoscopic VHR (N=3-53 per study).
        • Data Provenance: Similar to non-complex, July 7, 2013, and January 1, 2017, primarily United States, retrospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable in the context of this document. The "ground truth" for the performance data is derived directly from clinical outcomes recorded in a surgical registry (AHSQC) and published clinical literature. These are factual clinical occurrences (e.g., length of stay, complications, readmissions, mortality) rather than subjective interpretations by experts, as would be the case for image-based diagnostic AI/ML.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. As noted above, the data are clinical outcomes from a registry, not expert-labeled data requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is not an AI-assisted diagnostic or interpretation device that would involve human "readers." The "device" (da Vinci Xi/X Surgical System) is a surgical tool. The study compares surgical approaches (robotic, open, laparoscopic), not the performance of human readers with or without AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. The da Vinci system is a human-controlled robotic surgical system; it does not operate "stand-alone" as an algorithm would. Performance is always human-in-the-loop (i.e., the surgeon uses the system).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The ground truth for the clinical performance data (test set) is Real-World Evidence (RWE) from a clinical registry (AHSQC). This constitutes outcomes data (e.g., length of stay, complication rates, readmission rates, recurrence rates, mortality rates) directly recorded from patient care. For certain comparisons (Complex VHR Robotic-Assisted vs. Laparoscopic), it also incorporates published literature on laparoscopic VHR outcomes.

    8. The sample size for the training set:

    The document does not describe a "training set" in the context of machine learning. The da Vinci Xi and X Surgical Systems are existing electromechanical systems; this submission is for an expanded indication. The "training" for the system itself has already occurred and is inherent in its design and previous clearances. The data used in this submission serves as clinical validation data for the expanded Indication for Use.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no "training set" in the AI/ML sense described.

    In summary, this 510(k) submission demonstrates the substantial equivalence and safety/effectiveness of adding a new surgical indication to an already cleared robotic surgical system, not the performance of a new AI/ML device against specific quantitative acceptance criteria. The "study" here is a retrospective analysis of clinical outcomes from a large surgical registry and published literature.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173842

    Validate with FDA (Live)

    Date Cleared
    2018-04-23

    (126 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuitive Surgical Endoscopic Instrument Control System Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    The Intuitive Surgical Endoscopic Instrument Control X Surgical System Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

    Device Description

    This 510(k) is for a labeling modification only, to include the following additional representative, specific procedures under the cleared "general laparoscopic surgical procedures" Indication for Use of the da Vinci Xi Surgical System, Model IS4000 (K131861) and the da Vinci X Surgical System, Model IS4200 (K171294): Gastric Bypass (Roux-en-Y) and Gastric Sleeve. There are no changes to the technological characteristics of the cleared da Vinci Xi or X Surgical Systems (Models IS4000 and IS4200) proposed in this submission. The da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200, are softwarecontrolled, electro-mechanical systems designed for surgeons to perform minimally invasive surgery. The Model IS4000 and Model IS4200 Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

    AI/ML Overview

    This document is a 510(k) summary for a labeling modification of the Intuitive Surgical da Vinci Xi and X Surgical Systems (Models IS4000 and IS4200). It focuses on adding specific gastric procedures (Gastric Bypass and Gastric Sleeve) to the "general laparoscopic surgical procedures" indication for use. Therefore, the acceptance criteria and supporting studies are related to the safety and effectiveness of the device for these additional procedures, primarily against existing surgical methods (open and laparoscopic).

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly state formal "acceptance criteria" with specific numerical thresholds for performance metrics. Instead, it demonstrates substantial equivalence to predicate devices and established surgical practices (open and laparoscopic) by showing comparable or improved outcomes in various metrics. The acceptance is implicitly based on these comparisons showing non-inferiority or superiority in critical safety and effectiveness parameters.

    Metric (Acceptance Criteria implicitly: comparable or better than predicate/laparoscopic/open surgery)Reported Device Performance (da Vinci-assisted vs. Open)Reported Device Performance (da Vinci-assisted vs. Laparoscopic)
    Mortality Rate (in-hospital or 30 days)ComparableComparable
    Lengths of Hospital StayShorterComparable or Shorter
    Intraoperative Complication RatesComparableComparable
    Postoperative Complication RatesComparable or LowerComparable or Lower
    Anastomotic Leak RateComparableComparable or Lower (depending on anastomosis technique)
    Anastomotic Stricture RateComparableComparable or Lower (depending on anastomosis technique)
    Reoperation RateComparable or LowerComparable or Lower
    Readmission RateNot reported for vs. OpenComparable
    Operative TimeIncreased (but not associated with increased mortality/complication)Increased (in N=7 publications) or Comparable/Shorter (in N=3 publications), not associated with increased mortality/complication
    Estimated Blood Loss (EBL)Not reported for vs. OpenComparable
    Conversion RatesNot reported for vs. OpenComparable or Lower

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Description: The "test set" in this context refers to the published clinical data analyzed for Gastric Bypass (Roux-en-Y), which served as the "umbrella procedure" to support both Gastric Bypass and Gastric Sleeve.
    • Sample Size (for Gastric Bypass umbrella procedure): The clinical data summarized in Table 1A involved 10 publications. The total number of patients included across these studies is:
      • da Vinci (Robotic) Cohort: 172 + 61 + 100 + 100 + 100 + 125 + 100 + 65 + 388 + 143 = 1,354 patients
      • Laparoscopic (Lap) Cohort: 173 + 46 + 100 + 100 + 100 + 121 + 100 + 54 + 389 + 323 = 1,506 patients
      • Open Cohort: 95 + 524 = 619 patients
    • Data Provenance: The data are from published clinical studies. The specific countries of origin are not mentioned, but clinical literature is generally international. The studies include both non-randomized, controlled, prospective studies (2 studies) and retrospective cohort studies (8 studies).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • This information is not provided in the given text. The "ground truth" for these clinical outcome studies is inherent in the reported patient outcomes and surgical metrics gathered and published by the original study authors. There is no mention of independent experts establishing a ground truth for a test set for the purpose of this 510(k) submission. The FDA's review process inherently involves medical experts, but they are assessing the submitted evidence, not establishing a new ground truth.

    4. Adjudication Method for the Test Set

    • This information is not provided. The text describes a review of published clinical data. Adjudication methods (like 2+1 or 3+1 consensus) are typically associated with image-based diagnostic studies or clinical trials where expert review of individual cases is part of the study design. For a meta-analysis or review of published literature on surgical outcomes, the "adjudication" is primarily how each original study defined and reported its outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done. MRMC studies are typically used to evaluate the diagnostic performance of a device (often imaging-based AI) by comparing how multiple readers perform with and without the device. This submission is for a surgical system and its labeled indications, and the performance evaluation relies on real-world surgical outcomes and peer-reviewed literature, not diagnostic reader performance.
    • Effect size of human readers improving with AI vs. without AI assistance: Not applicable, as it's not an MRMC study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • No, a standalone (algorithm-only) performance study was not done. The da Vinci system is a surgical assistance system, inherently involving a human surgeon (human-in-the-loop). Its "performance" is evaluated by the outcomes of surgeries performed with the system assisting the surgeon, compared to traditional methods.

    7. The Type of Ground Truth Used

    • The "ground truth" for the clinical performance evaluation is based on patient outcomes and surgical metrics reported in peer-reviewed clinical literature (e.g., mortality, complication rates, length of stay, operative time, reoperation rate, etc.). This is essentially outcomes data collected and documented in clinical studies.

    8. The Sample Size for the Training Set

    • This information is not applicable in the conventional sense for this type of submission. The da Vinci system is an electro-mechanical surgical system, not an AI/ML algorithm that undergoes a distinct "training" phase on a dataset of cases to learn to perform a task. The "experience" or "learning" of the system happens through engineering design, software development, preclinical testing, and extensive clinical validation over many years.
    • However, if one were to consider the continuous development and refinement of the system's design and software, it would be based on an enormous "dataset" of engineering principles, surgical knowledge, user feedback, and prior clinical experience.

    9. How the Ground Truth for the Training Set Was Established

    • As explained above, a "training set" with established ground truth in the context of machine learning is not directly applicable here. The development of the surgical system relies on established surgical practices, engineering standards, and medical knowledge. Pre-clinical animal studies (mentioned as 6 evaluations in 24 animals) contribute to the developmental understanding and validation of the system's design, but this isn't a "training set" in the AI sense.
    • The "ground truth" for surgical device development is rooted in the safety and effectiveness principles derived from surgical anatomy, physiology, surgical techniques, and clinical outcomes, primarily established by expert surgeons and medical researchers over decades.