Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K160571

    Validate with FDA (Live)

    Date Cleared
    2016-12-19

    (294 days)

    Product Code
    Regulation Number
    862.1373
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The cobas c 513 clinical chemistry analyzer is a fully automated, standalone clinical chemistry analyzer intended for the in-vitro quantitative determination of analytes in body fluids.

    The Tina-quant Hemoglobin A1cDx Gen.3 assay is intended for use as an aid in diagnosis of diabetes and as an aid in identifying patients who may be at risk for developing diabetes. It is an in vitro diagnostics reagent system intended for quantitative determination of mmol/mol hemoglobin A1c (IFCC) and % hemoglobin A1c (DCCT/NGSP) in hemolysate or whole blood on the Roche/Hitachi cobas c 513 clinical chemistry analyzer. HbA1c determinations are useful for monitoring of long-term blood glucose control in individuals with diabetes mellitus.

    Device Description

    The cobas c 513 clinical chemistry analyzer is a fully automated, software controlled analyzer system for in vitro quantitative determination of analytes in human body fluids. The cobas c 513 analyzer system includes a control unit and an analyzer with a closed tube sampling functionality.

    The cobas c 513 control unit includes a computer (PC) located in the sampler unit, a touchscreen monitor, soft-keyboard, a mouse (optional use), and a printer. The control unit is used to perform tasks on the analyzer and the PC runs the software that controls the analyzer.

    The software manages all instrument functions, all system functions, and all information related to orders and results. The software offers a graphical user interface (GUI) to control all functions by the operator. The control unit contains System Control software including interfaces to a Medical Device Data System (cobas Link) and to the customer Laboratory Information System (LIS).

    The analyzer is composed of the sampler unit and the analytical unit. The sampler unit is composed of the rack loading/unloading areas, a barcode reader, a rack rotor, a STAT port, and conveyor lines. It is used to load and unload racks, power on the system, access the PC, and manage the conveyance of samples to the analytical unit.

    The analytical unit is comprised of the reagent area, the sample area, and the reaction disk. Samples are conveyed to the analytical unit from the sample unit for photometric analysis before being returned to the sample unit.

    Anticoagulated whole blood is hemolyzed either manually or automatically prior to determination of HbAlc by a turbidimetric inhibition immunoassay. Liberated hemoglobin (Hb) in the hemolyzed sample is converted to a derivative having a characteristic absorption spectrum and measured bichromatically. The instrument calculates the % HbAlc from the HbAlc/Hb ratio according to a user selected protocol, either IFCC or NGSP protocols.

    The assay offers separate applications that are specific to the sample types whole blood and hemolysate. The Whole Blood Application differs from the Hemolysate Application in the hemolyzing step. For the Whole Blood Application, whole blood samples are placed on the analyzer. Hemolysis occurs onboard the analyzer. For the Hemolysate Application, hemolyzed samples are placed on the analyzer. Hemolysis occurs manually before placing the samples onboard the analyzer. The two applications yield the same results.

    AI/ML Overview

    This document, a 510(k) premarket notification, describes the performance evaluation of the cobas c 513 Analyzer and cobas c 513 Tina-quant HbA1cDx Gen.3 Assay. It does not describe an AI/ML powered device, but rather an in-vitro diagnostic system for measuring Hemoglobin A1c. Therefore, many of the typical acceptance criteria and study designs relevant to AI/ML based devices (e.g., expert consensus, MRMC studies, training set details) are not applicable or detailed in the same way.

    However, I can extract and present the information as per your request based on the provided document, interpreting "acceptance criteria" as the performance criteria the device was tested against and "study" as the non-clinical performance evaluation conducted.

    Device Name: cobas c 513 Tina-quant HbA1cDx Gen.3 Assay

    Device Type: In-vitro diagnostic reagent system for quantitative determination of HbA1c.


    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) for an in-vitro diagnostic (IVD) device, the "acceptance criteria" are not explicitly listed in a single table with pass/fail remarks as might be seen for an AI/ML algorithm. Instead, the document details various performance studies (e.g., precision, linearity, method comparison, interference) and their results, implicitly demonstrating that the device meets an acceptable level of performance for its intended use based on standard IVD guidelines.

    The table below summarizes the key performance metrics and their reported values. The "Acceptance Criteria" column reflects typical expected performance ranges or goals for such IVDs based on general industry standards and the context provided by the results themselves. The document states "All acceptance criteria for method comparison were met" and "All acceptance criteria for drug interferences were met," indicating that internal predefined criteria were satisfied.

    Performance MetricAcceptance Criteria (Implicit/Inferred from Context)Reported Device Performance (cobas c 513 Tina-quant HbA1cDx Gen.3 Assay)
    Limit of Blank (LoB)Expected to be very low, indicating minimal signal in the absence of analyte.Hb: 0.50 g/dL (0.31 mmol/L) HbA1c: 0.19 g/dL (0.12 mmol/L)
    Limit of Detection (LoD)Expected to be low, indicating the lowest concentration reliably detected.Hb: 1.00 g/dL (0.62 mmol/L) HbA1c: 0.29 g/dL (0.18 mmol/L)
    Precision (Total %CV) (Hemolysate)Expected to be low across the measuring range (e.g., typically < 5-10% depending on concentration).Range across human samples (4.63% to 13.8% HbA1c): 0.9% - 1.3%
    Precision (Total %CV) (Whole Blood)Expected to be low across the measuring range (e.g., typically < 5-10% depending on concentration).Range across human samples (4.60% to 13.9% HbA1c): 0.9% - 1.6%
    Method Comparison (Mean Bias vs. NGSP TOSOH)Expected to be close to zero, with narrow confidence intervals, implying good agreement with the reference method.Whole Blood: -0.016 (95% CI: -0.260 to 0.227) Hemolysate: -0.138 (95% CI: -0.429 to 0.153)
    Method Comparison (% Relative Bias)Expected to be low (e.g., within a few percentage points of the reference).At various %HbA1c levels: Whole Blood: -0.8% to 0.1% Hemolysate: -2.8% to -0.8%
    Total Error (TE)Expected to be within clinically acceptable limits for HbA1c measurement.Whole Blood: 2.1% - 3.9% Hemolysate: 3.7% - 5.2%
    Linearity (Pearson's r)Expected to be very close to 1, indicating a strong linear relationship between measured and theoretical values.Hb: 0.9999 HbA1c: 0.9990
    Measuring Range (Claimed)Established range should be supported by linearity and other performance data.Hemoglobin: 4.0 - 40 g/dL (2.48-24.8 mmol/L) HbA1c: 0.3 - 1.93 g/dL (0.186-1.20 mmol/L) This corresponds to 4.2-15.5 % HbA1c (DCCT/NGSP)
    Matrix ComparisonMean differences and 95% CIs should be clinically insignificant across different anticoagulants and fill volumes.Mean Differences: -0.0559 to 0.0393 (within 95% CIs of approx. +/-0.5)
    Endogenous InterferenceNo significant interference (defined as >7% deviation in HbA1c) at specified maximum concentrations.Met criteria; specific maximum concentrations without interference provided for: Lipemia, Bilirubin, Ditaurobilirubin, Glucose, Rheumatoid Factor
    Drug InterferencesNo significant interference (defined as > ±7% deviation) at specified drug concentrations.Met criteria for all 16 tested drugs.
    Cross-ReactivityNo significant interference due to common hemoglobin variants and other related substances.No significant interference with tested Hb variants (HbS, HbC, HbE, HbD, HbA2) unless otherwise specified. HbF > 7% may yield lower than expected values.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Limit of Blank (LoB): One analyte-free sample measured 10-fold with three lots, over six runs distributed over >3 days on one c 513 analyzer. Total 60 measurements per lot.
    • Limit of Detection (LoD): Five samples with low-analyte concentration measured 2-fold with three lots, over six runs distributed over >3 days on one c 513 analyzer. Total 60 measurements per lot.
    • Precision: Eight samples (two controls, seven human samples with varying HbA1c concentrations) measured in two aliquots per sample, once each in two runs per day for 21 days, on three cobas c 513 analyzers using 3 reagent lots per system.
    • Method Comparison: 155 samples (Hemolysate) and 154 samples (Whole Blood) from the secondary NGSP reference laboratory. Samples were tested over a 3-day period with one lot of reagent on one cobas c 513 analyzer.
    • Linearity: Two separate dilution series consisting of at least 11 levels, using human hemolysate sample pools. Samples measured in triplicate.
    • Matrix Comparison: At least 40 samples of each sample type (anticoagulant and fill volume) and at least 40 half-filled tubes of each sample type. Overall sample count not explicitly summed but implies a substantial number.
    • Endogenous Interference: 12 spiked samples (2 HbA1c levels x 6 interferents). Each treated pool (spiked and control) had a dilution series of at least 10 levels. Dilution series tested 10-fold.
    • Drug Interferences: Native patient samples at 2 different HbA1c levels (approx. 6% and 8% HbA1c) were spiked with 16 common drugs at two defined concentrations. Samples measured in 10-fold.
    • Cross-reactivity: Multiple experiments for each cross-reactant (HbA0, HbA1a+b, Acetylated Hb, Carbamylated Hb, Glycated Albumin, Labile HbA1c). Each involved a series of dilutions or concentration variations, with 10 replicates per sample.
    • Hemoglobin Variants: 20 samples each for HbS, HbC, HbE, HbD, Elevated F. 13 samples for HbA2. Each tested once in at least one run.

    Data Provenance: The document does not explicitly state the country of origin for the human samples used in these studies. The studies were non-clinical performance evaluations (laboratory-based testing). The samples for method comparison were obtained from a secondary NGSP (National Glycohemoglobin Standardization Program) reference laboratory. This implies a standardized and presumably well-characterized dataset. The studies are prospective in the sense that they are designed performance evaluations conducted specifically for device submission, rather than analysis of existing stored data.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This document describes a laboratory diagnostic device, not an AI/ML algorithm that interprets medical images or other complex data requiring human expert consensus for ground truth.

    • Ground truth for HbA1c values: For the method comparison study, the ground truth was established by a secondary NGSP reference laboratory using a Tosoh HPLC system. This is a recognized and standardized reference method for measuring HbA1c. The document mentions "Traceability: This method has been standardized against the approved IFCC reference method for the measurement of HbA1c in human blood and can be transferred to results traceable to DCCT/NGSP by calculation." This indicates that the ground truth is established via an internationally recognized and highly accurate analytical method, rather than through expert human assessment.

    Therefore, the concept of "number of experts" and their "qualifications" for establishing ground truth in the context of an AI/ML diagnostic for image interpretation is not directly applicable here.


    4. Adjudication Method for the Test Set

    Not applicable. As described above, the ground truth for this IVD device is established through a laboratory reference method (Tosoh HPLC system) traceable to international standards (IFCC, NGSP), not through human expert assessment needing adjudication.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No. This document describes the performance of an in-vitro diagnostic instrument and assay, not an AI/ML application intended to assist human readers (e.g., radiologists) in interpreting medical images. Therefore, an MRMC study and the concept of "how much human readers improve with AI vs. without AI assistance" are not relevant to this device.


    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    Yes, in essence. The performance studies detailed in this document (precision, method comparison, linearity, interference, etc.) directly evaluate the cobas c 513 Tina-quant HbA1cDx Gen.3 Assay and cobas c 513 Analyzer as a standalone system for quantitative determination of HbA1c in patient samples. The results are generated entirely by the instrument and reagents, independent of human interpretive input beyond standard laboratory procedures for operating the device and reporting results. There is no "human-in-the-loop" interpretive component that would require a separate "standalone" versus "human-in-the-loop" performance evaluation as might be seen for an AI algorithm.


    7. The Type of Ground Truth Used

    The ground truth for the HbA1c values, particularly in the method comparison study, was established using a Tosoh HPLC system operated by a secondary NGSP (National Glycohemoglobin Standardization Program) reference laboratory. This method is traceable to the approved IFCC (International Federation of Clinical Chemistry and Laboratory Medicine) reference method and linked to DCCT/NGSP standards. This represents a highly standardized and analytically robust reference method/pathology/analytical outcome data rather than expert consensus on subjective findings.


    8. The Sample Size for the Training Set

    This document does not describe the development of an AI/ML algorithm that would have separate "training" and "test" sets in the conventional sense. This is an IVD device where performance characteristics are assessed on "test sets" (i.e., various sample groups used in the performance studies) to demonstrate its analytical capability. Therefore, the concept of a "training set" is not applicable here.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1