510(k) Data Aggregation

    K Number
    K203220

    Validate with FDA (Live)

    Device Name
    cobas BKV
    Date Cleared
    2021-01-29

    (88 days)

    Product Code
    Regulation Number
    866.3183
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    cobas® BKV is an in vitro nucleic acid amplification test for the quantitation of BKV) DNA in human EDTA plasma and urine stabilized in cobas® PCR media on the cobas® 6800/8800 Systems.

    In EDTA plasma, cobas® BKV is intended for use as an aid in the management of BKV in transplant patients. In patients undergoing monitoring of BKV in EDTA plasma, serial DNA measurements can be used to indicate the need for potential treatment changes and to assess viral response to treatment.

    In urine stabilized in cobas® PCR Media, cobas® BKV is intended for use as an aid in the management of BKV in transplant patients.

    The results from cobas® BKV are intended to be read and analyzed by a qualified licensed healthcare professional in conjunction with clinical signs and symptoms and relevant laboratory findings. Test results must not be the sole basis for patient management decisions .

    cobas® BKV is not intended for use as a screening test for blood products or human cells, tissues, and cellular and tissue-based products (HCT/Ps).

    Device Description

    cobas® BKV (Figure 1) is based on fully automated sample preparation (nucleic acid extraction and purification) followed by PCR amplification and detection. The cobas® 6800/8800 Systems consist of the sample supply module, the transfer module, the processing module, and the analytic module. Automated data management is performed by the cobas® 6800/8800 software which assigns test results for all tests as either target not detected, BKV DNA detected < LLoQ (lower limit of quantitation), BKV DNA detected > ULoQ (upper limit of quantitation), or a value in the linear range LLoQ < x < ULoQ. Results can be reviewed directly on the system screen, exported, or printed as a report.

    Nucleic acid from patient samples and added lambda DNA-OS molecules is simultaneously extracted. In summary, viral nucleic acid is released by addition of proteinase and lysis reagent to the sample. The released nucleic acid binds to the silica surface of the added magnetic glass particles. Unbound substances and impurities, such as denatured protein, cellular debris and potential PCR inhibitors are removed with subsequent wash reagent steps and purified nucleic acid is eluted from the glass particles with elution buffer at elevated temperature.

    Selective amplification of target nucleic acid from the sample is achieved by the use of a dual target virus specific approach from highly-conserved regions of the BKV located in the BKV small t-antigen region and the BKV VP2 region. Selective amplification of DNA-QS is achieved by the use of sequence-specific forward and reverse primers which are selected to have no homology with the BKV genome. A thermostable DNA polymerase enzyme is used for amplification. The target and DNA-QS sequences are amplified simultaneously utilizing a universal PCR amplification profile with predefined temperature steps and number of cycles. The master mix includes deoxyuridine triphosphate (dUTP), instead of deoxythimidine triphosphate (dTTP), which is incorporated into the newly synthesized DNA (amplicon).13 Any contaminating amplicon from previous PCR runs is eliminated by the AmpErase enzyme, which is included in the PCR mix, when heated in the first thermal cycling step. However, newly formed amplicons are not eliminated since the AmpErase enzyme is inactivated once exposed to temperatures above 55°C.

    The cobas® BKV master mix contains two detection probes specific for BKV target sequences and one for the DNA-QS. The probes are labeled with target-specific fluorescent reporter dyes allowing simultaneous detection of BKV target and DNA-QS in two different target channels.45 The fluorescent signal of the intact probes is suppressed by the quencher dye. During the PCR amplification step, hybridization of the probe to the specific single-stranded DNA templates results in cleavage by the 5'-to-3' nuclease activity of the DNA polymerase resulting in separation of the reporter and quencher dyes and the generation of a fluorescent signal. With each PCR cycle, increasing amounts of cleaved probes are generated and the cumulative signal of the reporter dye is concomitantly increased. Real-time detection and discrimination of PCR products are accomplished by measuring the fluorescence of the released reporter dyes for the viral targets and DNA-OS.

    AI/ML Overview

    This document describes the acceptance criteria and supporting studies for the cobas® BKV device for the quantitation of BK virus (BKV) DNA in human urine. The information is extracted from a 510(k) summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance (cobas® BKV in Urine)
    Limit of Detection (LoD): 95% hit rate for BKV DNA.12.2 IU/mL (WHO International Standard, 95% confidence range: 9.2-18.3 IU/mL). Achieved ≥95% hit rate for subgroups Ia, Ic, and subtypes II, III, IV at 12.2 IU/mL.
    Linear Range: Accuracy within ± 0.2 log10.7.41E+01 IU/mL to 7.41E+08 IU/mL. Maximum deviation of linear regression from better fitting non-linear regression was ≤ ± 0.2 log10 for all tested genotypes within the linear range.
    Lower Limit of Quantitation (LLoQ): Mean deviation between observed and assigned log10 titer ≤ ± 0.3 log10 (based on upper 95% CI of worst performing lot). Total Analytical Error (TAE) ≤ 1 log10.200 IU/mL. Mean deviation between observed and assigned log10 titer was ≤ 0.3 log10. TAE was ≤ 0.44 for all lots and concentrations (Table 7).
    Precision (Within-Laboratory): High precision across the concentration range.Demonstrated high precision across a concentration range of 7.41E+02 IU/mL to 7.41E+05 IU/mL (Table 8, Table 9). Total %CV ranged from 7% (highest concentration) to 23% (lowest concentration).
    Analytical Specificity: No interference from listed microorganisms; mean log10 titer of positive BKV samples with interfering organisms within ± 0.5 log10 of spike control.None of the tested non-BKV pathogens (bacteria, yeast, viruses in Table 10) interfered. Mean log10 titer of positive BKV samples was within ± 0.5 log10 of the spike control.
    Interfering Substances: No interference from listed endogenous substances and drug compounds, with mean log10 titer of positive BKV samples with interfering substances within ± 0.5 log10 of spike control.All listed endogenous interferences and drug compounds (except talcum powder at >0.05%) did not interfere. Mean log10 titer of positive BKV samples was within ± 0.5 log10 of the spike control.
    Cross Contamination: Zero cross-contamination rate with a low upper 95% confidence interval.0.0% (upper one-sided 95% CI 1.24%) with 240 replicates of negative samples.
    Clinical Reproducibility: Acceptable reproducibility; 100% detection of 3 x LLoQ samples; 95% CI for difference between 2 measurements within ± 0.20 log10 copies/mL.Acceptable clinical reproducibility. 100% of 3 x LLoQ samples detected. All estimated 95% CLs for the difference between 2 measurements from the same subject were within ± 0.20 log10 copies/mL.
    Negative Percent Agreement (NPA) (Clinical): High negative percent agreement.100% (95% Exact CI of 94.1% to 100%) for 61 valid negative samples.
    Clinical Concordance with LDT: High agreement at various thresholds and strong correlation.Concordance analysis with comparator LDT showed high agreement (e.g., 93.9% at Target Not Detected threshold, 99.5% at LLoQ threshold). Deming linear regression showed a strong correlation with CI for intercept within ±0.5 log10 IU/mL.

    2. Sample Sizes Used for the Test Set and Data Provenance

    The studies focused on analytical performance (Limit of Detection, Linearity, Precision, Specificity, Interference, Cross-contamination) and clinical performance (Reproducibility, Clinical Concordance).

    • Limit of Detection (LoD):
      • WHO International Standard: 63 replicates per concentration level (total 7 levels, x3 lots = 1323 replicates).
      • Subgroups/Subtypes Verification: 63 replicates per concentration level for each genotype (total 5 genotypes, 3 levels each, x3 lots = 2835 replicates).
    • Linear Range:
      • Main linearity: 36 replicates per panel member (10 panel members, x3 lots = 1080 replicates).
      • Genotype linearity: 12 replicates per level for each genotype (8 panel members each, 5 genotypes, x3 lots = 1440 replicates).
    • Lower Limit of Quantitation (LLoQ): Data from the Linearity study at 100, 200, and 300 IU/mL concentrations.
    • Precision (Within-Laboratory): 72 replicates for each of 5 dilution levels (x3 lots = 1080 replicates).
    • Analytical Specificity: 3 replicates for each of the microorganisms listed in Table 10, both in BKV-negative and BKV-positive urine (number of microorganisms not explicitly totalled, but substantial).
    • Interfering Substances: Replicates for each substance in presence and absence of BKV DNA (number of replicates not explicitly stated, but implies multiple for each substance in Table 11).
    • Cross Contamination: 240 replicates of BKV-negative matrix samples, 225 replicates of high titer BKV DNA urine samples.
    • Clinical Reproducibility: 270 tests per concentration (5 concentrations, total 1350 tests, not including controls).
    • Clinical Performance / Concordance: 308 neat urine samples from 84 transplant subjects (for concordance analysis). 61 negative samples for NPA. 153 BKV positive urine samples from 55 transplant subjects (for correlation analysis).

    Data Provenance: The document implies that the non-clinical studies were conducted internally or at authorized labs. The clinical performance evaluation was conducted at 3 testing sites, suggesting multi-site prospective data collection. The data samples were human EDTA plasma and urine. The origin of the samples (country) is not explicitly stated in the provided text. The clinical study used a retrospective cohort of samples from transplant patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The ground truth for the analytical studies was established based on known concentrations of BKV international standards or armored DNA. For clinical performance, the comparator was a "well-established laboratory developed nucleic acid test (LDT)".

    The document does not mention the use of experts to establish ground truth for the test sets in the typical sense of human readers for image-based diagnostics. The "ground truth" for this diagnostic device study is based on the highly controlled properties of the spiked samples (known concentrations, genotypes) for analytical performance, and the results from a comparator LDT for clinical concordance.

    4. Adjudication Method for the Test Set

    Not applicable in the context of this in vitro diagnostic device, as the "ground truth" is based on quantitative measurements and known concentrations, not subjective expert assessment requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an in vitro nucleic acid amplification test (NAAT) for quantitative measurement of BKV DNA. It is not an AI-assisted diagnostic device requiring human reader input or interpretation in the way an imaging diagnostic device would. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant to this submission.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the studies described are for the standalone performance of the cobas® BKV system, which is an automated molecular diagnostic assay. The system performs "fully automated sample preparation (nucleic acid extraction and purification) followed by PCR amplification and detection." The results are "assigned... by the cobas® 6800/8800 software." While "results are intended to be read and analyzed by a qualified licensed healthcare professional in conjunction with clinical signs and symptoms and relevant laboratory findings," the primary performance metrics are based on the direct output of the automated system.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    • Analytical Studies: The ground truth for analytical studies (LoD, Linearity, Precision, Specificity, Interference) was established using known concentrations of BKV DNA, including the WHO International Standard (NIBSC 14/212), BKV armored DNA, and clinical specimens diluted to specific concentrations. Samples were spiked into BKV-negative urine.
    • Clinical Studies: For clinical concordance, the ground truth was based on the results from a "well-established laboratory developed nucleic acid test (LDT) (comparator BKV LDT)" on actual clinical urine samples. DNA sequencing was also used in some cases to confirm BKV presence in discordant results.

    8. The Sample Size for the Training Set

    The document describes performance evaluation studies (validation and verification) rather than a machine learning model development process that typically involves distinct training and test sets.
    Therefore, a separate "training set" sample size for a machine learning algorithm is not applicable in the context of this in vitro diagnostic device, which is based on established molecular biological techniques (PCR).

    9. How the Ground Truth for the Training Set Was Established

    As this is not an AI/ML-based device with a "training set," this question is not applicable. The ground truth for the evaluation of the device was established through known concentrations of viral standards and comparison to a comparator LDT, as described in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202215

    Validate with FDA (Live)

    Date Cleared
    2020-09-02

    (27 days)

    Product Code
    Regulation Number
    866.3183
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    cobas® BKV is an in vitro nucleic acid amplification test for the quantitation of BK virus (BKV) DNA in human EDTA plasma on the cobas® 6800/8800 Systems.

    cobas® BKV is intended for use as an aid in the management of BKV in transplant patients. In patients undergoing monitoring of BKV, serial DNA measurements can be used to indicate the need for potential treatment changes and to assess viral response to treatment.

    The results from cobas® BKV are intended to be read and analyzed by a qualified licensed healthcare professional in conjunction with clinical signs and symptoms and relevant laboratory findings. Test results must not be the sole basis for patient management decisions.

    cobas® BKV is not intended for use as a screening test for blood products or human cells, tissues, and cellular and tissue-based products (HCT/Ps).

    Device Description

    cobas® BKV is based on fully automated sample preparation (nucleic acid extraction and purification) followed by PCR amplification and detection. The cobas® 6800/8800 Systems consist of the sample supply module, the transfer module, the processing module, and the analytic module. Automated data management is performed by the cobas® 6800/8800 software which assigns test results for all tests as either target not detected, BKV DNA detected < LLoQ (lower limit of quantitation), BKV DNA detected > ULoQ (upper limit of quantitation), or a value in the linear range LLoQ < x < ULoQ. Results can be reviewed directly on the system screen, exported, or printed as a report.

    Nucleic acid from patient samples and added lambda DNA-QS molecules is simultaneously extracted. In summary, viral nucleic acid is released by addition of proteinase and lysis reagent to the sample. The released nucleic acid binds to the silica surface of the added magnetic glass particles. Unbound substances and impurities, such as denatured protein, cellular debris and potential PCR inhibitors are removed with subsequent wash reagent steps and purified nucleic acid is eluted from the glass particles with elution buffer at elevated temperature.

    Selective amplification of target nucleic acid from the sample is achieved by the use of a dual target virus specific approach from highly-conserved regions of the BKV located in the BKV small t-antigen region and the BKV VP2 region. Selective amplification of DNA-QS is achieved by the use of sequence-specific forward and reverse primers which are selected to have no homology with the BKV genome. A thermostable DNA polymerase enzyme is used for amplification. The target and DNA-QS sequences are amplified simultaneously utilizing a universal PCR amplification profile with predefined temperature steps and number of cycles. The master mix includes deoxyuridine triphosphate (dUTP), instead of deoxythimidine triphosphate (dTTP), which is incorporated into the newly synthesized DNA (amplicon). Any contaminating amplicon from previous PCR runs is eliminated by the AmpErase enzyme, which is included in the PCR mix, when heated in the first thermal cycling step. However, newly formed amplicons are not eliminated since the AmpErase enzyme is inactivated once exposed to temperatures above 55°C.

    The cobas® BKV master mix contains two detection probes specific for BKV target sequences and one for the DNA-QS. The probes are labeled with target-specific fluorescent reporter dyes allowing simultaneous detection of BKV target and DNA-QS in two different target channels. The fluorescent signal of the intact probes is suppressed by the quencher dye. During the PCR amplification step, hybridization of the specific single-stranded DNA templates results in cleavage by the 5'-to-3' nuclease activity of the DNA polymerase resulting in separation of the reporter and quencher dyes and the generation of a fluorescent signal. With each PCR cycle, increasing amounts of cleaved probes are generated and the cumulative signal of the reporter dye is concomitantly increased. Real-time detection and discrimination of PCR products are accomplished by measuring the fluorescence of the released reporter dyes for the viral targets and DNA-QS.

    AI/ML Overview

    The provided document is a 510(k) Summary for the cobas® BKV test for use on the cobas® 6800/8800 Systems. This document focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance evaluation, rather than an AI/ML device requiring a comparative effectiveness study with human readers. Therefore, several of the requested sections (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, ground truth for test set, training set details) are not directly applicable or explicitly stated in the context of this In Vitro Diagnostic (IVD) device submission for a quantitative nucleic acid amplification test.

    However, I will extract relevant information to address the applicable criteria based on the provided text.

    Acceptance Criteria and Device Performance for cobas® BKV

    The acceptance criteria for this diagnostic device are primarily defined by various performance characteristics required for quantitative viral nucleic acid tests. The study proves the device meets these criteria through a series of non-clinical performance evaluations.

    Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Implicit from Study Design/Context)Reported Device Performance and Study Details
    Limit of Detection (LoD)Determine the concentration at which 95% hit rate is achieved for the WHO International Standard and all subgroups/subtypes, across multiple lots.- WHO International Standard (NIBSC 14/212): Determined as 21.5 IU/mL (PROBIT, 95% hit rate) with a 95% CI of 16.3 – 32.4 IU/mL in EDTA plasma, using the least sensitive lot. - Subgroups Ia, Ic and Subtypes II, III and IV: Verified detection at 21.5 IU/mL with a ≥ 95% hit rate for all tested genotypes (e.g., Subgroup Ia at 21.5 IU/mL: 100.0% hit rate, 63/63 positives; Subgroup Ic at 21.5 IU/mL: 100.0% hit rate, 62/62 positives).
    TraceabilityDemonstrate proportionality and correlation to the 1st WHO International Standard for BKV DNA.- Calibration and standardization process provides quantitation values for panels and standards similar to expected values. - Maximum deviation from expected values was not more than 0.19 log10 IU/mL. - Deming regression for BKV WHO Standard showed Y = 0.951x + 0.208 with R² = 0.973.
    Linear RangeDemonstrate linearity of quantification within a specified range with acceptable deviation and accuracy.- Demonstrated linear from 1.01E+01 to 1.97E+08 IU/mL. - Absolute deviation from non-linear regression ≤ ± 0.1 log10 in human EDTA plasma. - Accuracy within ± 0.2 log10 across the linear range. - Verified for subgroups Ia, Ic and subtypes II, III and IV: maximum deviation between linear and higher order non-linear regression ≤ ± 0.2 log10.
    Lower Limit of Quantitation (LLoQ)Determine the lowest titer meeting acceptance criteria for Total Analytical Error (TAE ≤ 1.0 log10 IU/mL) and difference between two measurements.- Established as 21.5 IU/mL. - At 19.0 IU/mL (nominal concentration, lowest tested for LLoQ), all three lots combined showed TAE of 0.69 and difference between measurements (SD) of 0.73, both within 1.0 log10 IU/mL.
    Precision – Within LaboratoryDemonstrate high precision across specified concentration ranges, instruments, operators, and days.- High precision shown across 5.90E+01 IU/mL to 9.83E+05 IU/mL. - Total %CV ranged from 8% to 36% across the concentrations tested (e.g., 8% at 9.83E+05 IU/mL, 36% at 5.90E+01 IU/mL). - Results represent all aspects of the test procedure.
    Analytical SpecificityNo interference or cross-reactivity with common microorganisms and endogenous/exogenous substances.- Microorganisms: None of 17 viruses, 13 bacteria, and 3 yeast species interfered at tested concentrations (1.00E+05 to 1.00E+06 units/mL). Negative results for BKV-negative samples, positive for BKV-spiked samples. Titer within ± 0.5 log10 of control. - Interfering Substances: Elevated triglycerides, bilirubin (conjugated/unconjugated), albumin, hemoglobin, human DNA, and 17 drug compounds (including antimicrobials and immune suppressants) did not interfere. Titer within ± 0.5 log10 of control.
    Cross-ContaminationDemonstrate a low or zero cross-contamination rate.- 0% cross-contamination rate (0/240 negative replicates) with an upper one-sided 95% CI of 1.24%.
    ReproducibilityDemonstrate consistent performance across different reagent lots, test sites, batches, and testing days.- Evaluated at 3 testing sites using 3 reagent lots per site by 2 operators over 5 days. - Total Precision Standard Deviation ranged from 0.068 to 0.304 log10 IU/mL. - Lognormal CV(%) for Total Precision ranged from 15.74% to 79.43%. - 100% detection of 3 x LLoQ samples. - Equivalence shown between cobas® 6800 and 8800 systems. - Negative percent agreement (NPA) for reproducibility study: 100% (270/270 samples negative), 95% Exact CI: 98.6% to 100%.
    Clinical ConcordanceDemonstrate agreement with a well-established laboratory-developed test (LDT).- Total of 550 valid samples (217 neat, 303 diluted clinical, 30 contrived) from 129 transplant subjects were evaluable. - Agreement with Comparator BKV LDT (IU LDT): High concordance shown across different concentration ranges. - Negative Percent Agreement (NPA): 100% (43/43 samples negative) with 95% Exact CI of 91.8% to 100%. - At thresholds, percent agreement was high (e.g., ≥ threshold 2.3 Log10 IU/mL: 87.7%; < threshold 4.0 Log10 IU/mL: 100.0%). - Correlation: Deming Linear Regression Y = -0.429 + 1.019X with R-value of 0.96 (N=313 samples where both assays were BKV positive). - Mean bias was -0.357 log10 IU/mL, indicating a systematic difference but within acceptable analytical precision limits of the comparator LDT.

    Study Details:

    1. Sample Size and Data Provenance:

      • Test Set (Clinical Concordance): 550 valid samples (217 neat clinical, 303 diluted clinical derived from 129 transplant subjects, and 30 contrived samples).
      • Provenance: Clinical samples were from BKV infected and non-infected patients. Although not explicitly stated, the context of an FDA submission for an IVD test likely implies data collected under controlled conditions at testing sites, potentially from multiple geographical locations if a multi-site study was conducted (the Reproducibility study specified 3 testing sites). The samples included both retrospective (stored clinical samples) and prospective (contrived samples) elements.
      • Non-Clinical Studies: Sample sizes vary by test as detailed in the table (e.g., 63 replicates per level for LoD, 36 replicates per panel member for linear range, 72 replicates per level for precision, 240/225 replicates for cross-contamination, 270 valid tests per concentration for reproducibility).
    2. Number of Experts and Qualifications: Not applicable. This is an IVD device for quantitative viral DNA measurement, not an AI/ML device for image interpretation or diagnosis by experts. Ground truth is established analytically (e.g., known concentrations of viral standards) or through comparison to a well-established laboratory method.

    3. Adjudication Method: Not applicable. For IVD tests like this, objective analytical measurements are compared rather than subjective interpretations needing adjudication.

    4. MRMC Comparative Effectiveness Study: No. This is an In Vitro Diagnostic (IVD) quantitative nucleic acid amplification test, not an AI-assisted diagnostic imaging device that involves human readers. The standard of comparison is against a well-established laboratory method (LDT) and analytical performance targets, not human reader improvement.

    5. Standalone Performance: Yes, the entire document describes the standalone performance of the cobas® BKV assay. The device measures BKV DNA concentration directly without human-in-the-loop assistance for the measurement itself, though a qualified healthcare professional interprets the results in a clinical context.

    6. Type of Ground Truth Used:

      • Analytical/Reference Standards: For LoD, linearity, traceability, precision, and specificity, the ground truth was established using known concentrations of the 1st WHO International Standard for BK Virus DNA (NIBSC 14/212), armored DNA for specific subgroups/subtypes, and well-characterized high titer BKV DNA (subgroup Ib).
      • Comparative Method: For clinical concordance, the ground truth was established by comparing results from the cobas® BKV test to those from a well-established laboratory developed nucleic acid test (LDT) from Indiana University (referred to as "comparator BKV LDT"). This LDT results served as the reference for comparison and agreement.
    7. Sample Size for Training Set: Not applicable. This is a molecular diagnostic assay characterized through analytical and clinical performance studies, not an AI/ML model that undergoes a distinct training phase with a dedicated training set. The "training" of the assay involves its development and optimization based on scientific principles of nucleic acid amplification and detection, and then validated through the studies described.

    8. How Ground Truth for Training Set was Established: Not applicable, as there isn't a "training set" in the AI/ML sense. The development of the assay (analogous to training) would rely on molecular biology knowledge, optimization experiments, and characterization against internally developed standards and reference materials, ultimately ensuring performance against defined analytical targets and external standards (like the WHO standard).

    Ask a Question

    Ask a specific question about this device