Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K142572

    Validate with FDA (Live)

    Date Cleared
    2015-01-21

    (131 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Zimmer Zfx Abutment for Zimmer 3.1mmD Implant System, Titanium is used as a terminal or intermediate abutment for a cemented prosthesis. The abutment can be used for a single or multiple-unit restoration

    Device Description

    The Zimmer Zfx Abutment for Zimmer 3.1mmD Implant System is designed for use with Zimmer 3.1mmD Dental Implants to support single or multi tooth restorations. The new abutment will be offered in a 2.9mm implant platform.

    The Zimmer Zfx Abutment for Zimmer 3.1mmD Implant System is a patient specific dental implant abutment. The purpose of a Patient-Specific abutment is to satisfy customer needs that are otherwise difficult to meet with off-the-shelf abutments. They can be manufactured in multiple sizes, shapes, and angles within the limits established in this submission. They frequently incorporate the modifications typically done at a dental laboratory or "chair-side" by a dentist.

    Traditional methodologies require the customer (dentist/laboratory technician) to begin with a "stock" abutment and use manual subtractive techniques to remove material from this original "stock" design. However, a Patient-Specific abutment will incorporate these same modifications desired by the customer (dentist/laboratory technician) at the time of fabrication at the manufacturing facility.

    The engineering drawings list ranges in areas (attributes) of the abutment that may be modified depending upon patient-specific needs.

    The abutment is composed of Titanium alloy (Ti6Al4V ELI), and secured to the implant with a separate Titanium alloy screw for retention.

    AI/ML Overview

    This 510(k) premarket notification describes the Zimmer Zfx Abutment for Zimmer 3.1mmD Implant System, a custom dental implant abutment made from Titanium alloy (Ti6Al4V ELI). The submission aims to demonstrate substantial equivalence to a predicate device, the Zimmer Patient-Specific Abutment, Internal Hex, Titanium (K071439).

    Here's an analysis of the provided information concerning acceptance criteria and the supporting study:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly present a table of acceptance criteria with corresponding performance metrics in a pass/fail format. Instead, it refers to compliance with an FDA guidance document and states that the new device demonstrated improvements over the predicate in certain non-clinical tests.

    Acceptance Criterion (Implicit)Reported Device Performance
    Mechanical Strength: Withstand anticipated forces (Fatigue and Compression)"Non-clinical testing consisted of performance of fatigue and compression testing in accordance with the FDA guidance Class II Special Controls Guidance Document: Root-form Dental Implants and Endosseous Dental Implant Abutments. The testing indicates that the new device is strong enough to withstand the anticipated forces and demonstrated improvements over the predicate device(s)." (This indicates the device met or exceeded the requirements outlined in the guidance and performed better than the predicate in these tests.)
    Sterilization Efficacy: Achieve minimum sterility assurance level (SAL)."The sterilization procedures listed in the Instruction For Use were validated to provide a minimum sterility assurance level of 10-9." (This is a clear statement of meeting the SAL of 10^-9.)
    MRI Compatibility: Pose no additional restrictions on MRI compared to patient."Additionally, Zimmer Dental implant systems were evaluated for interactions with magnetic fields during Magnetic Resonance Imaging (MRI) in accordance with the FDA Guidance: Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment. This was done to determine that the presence of the abutment poses no additional restrictions on MRI beyond those that would otherwise occur for the patient." (This confirms the evaluation was done, and the outcome indicates no additional restrictions, implying compatibility within the guidance for passive implants.)
    Material Equivalence: (Implicit from comparison table)The device is made of "Titanium 6Al-4V ELI," which is the same as the predicate device.
    Manufacturing Process & General Design Features Equivalence: (Implicit from comparison table)Stated that "The new abutment device is substantially equivalent to the predicate relative to material, manufacturing process and general design features." While specific metrics aren't provided here, the comparison table details design features (Implant Interface, Emergence, Margin, Platform Diameter, Cuff Width/Diameter, Minimum Height, Cone Angle), indicating that the new device's design falls within acceptable ranges or is comparable to the predicate, with modifications to platform diameter and cuff width/diameter being highlighted as specific differences allowed by the "patient-specific" nature.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: The document does not specify the exact sample sizes (number of units tested) for the fatigue, compression, or sterilization validation tests.
    • Data Provenance: The tests were "non-clinical" and performed by Zimmer Dental, Inc. The country of origin for the data is not explicitly stated, but given the company's address in Carlsbad, CA, it's presumed to be the USA. The data is from a prospective testing regimen designed to support the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable as the study was non-clinical (mechanical, sterilization, and MRI compatibility testing). There was no "ground truth" in the context of expert diagnosis or interpretation of data in the way it's typically understood for AI/medical imaging devices. The "ground truth" for these tests would be the measured physical properties and performance against established engineering and regulatory thresholds.

    4. Adjudication Method for the Test Set:

    This information is not applicable as the study was non-clinical. Adjudication methods are relevant for studies involving human interpretation (e.g., radiologists reviewing images).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:

    No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "No clinical testing was performed. Non-clinical testing was used to support the decision of safety and effectiveness." This device is an implantable medical device, not an AI diagnostic tool that assists human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    This information is not applicable. The device is a physical dental implant abutment, not an algorithm or AI system.

    7. The Type of Ground Truth Used:

    The "ground truth" for this non-clinical study relied on:

    • Engineering Standards and Regulatory Guidance: Performance was evaluated against the "FDA guidance Class II Special Controls Guidance Document: Root-form Dental Implants and Endosseous Dental Implant Abutments" for mechanical properties and the "FDA Guidance: Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment" for MRI compatibility.
    • Analytical Measurement Outcomes: The results of the fatigue, compression, sterilization, and MRI compatibility tests serve as the "ground truth" for whether the device meets the specified performance criteria derived from these standards. For example, a measured force to failure during compression testing provides a numerical ground truth for mechanical strength.

    8. The Sample Size for the Training Set:

    This information is not applicable. The device is a physical product and does not involve AI or machine learning models that require training sets.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1