Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K173709

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-01-26

    (53 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Zenith Support Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.

    Device Description

    The InNeuroCo Zenith Support product consists of a catheter, a hemostasis valve, and a split sheath introducer. The Zenith Support catheter is a single lumen, variable stiffness catheter that has an outer polymer shaft and is reinforced with a Stainless Steel cross coil. The Zenith Support catheter has a radiopaque Platinum/Iridium marker band on the distal end. The Zenith Support is available with an internal diameter of 0.074 inches and a nominal outer diameter of 0.086 inches. There are three working lengths available: 95 cm, 105 cm, and 115 cm in either diameter. The Zenith Support catheter has a PTFE-lined lumen throughout the catheter shaft. The catheter is flexible and has a hydrophilic coating. The Zenith Support catheter is inserted through a guide catheter or vascular sheath, provides access to the target site, and once in place, provides a reinforcing conduit for other intravascular devices. Accessories included with the device are a Tuohy-Borst hemostasis valve with an extension luer and a split sheath introducer. The Zenith product line is supplied sterile, non-pyrogenic, and intended for single use only.

    AI/ML Overview

    This request concerns the InNeuroCo Zenith Support Catheter, a percutaneous catheter used for introducing interventional devices into the peripheral and neuro vasculature. The provided document is a 510(k) premarket notification summary, which focuses on demonstrating substantial equivalence to a predicate device rather than a comprehensive study for novel device acceptance criteria.

    The information primarily revolves around non-clinical data and justifications for not re-testing certain parameters due to the device's similarity to its predicate. Therefore, detailed information on ground truth establishment, expert qualifications, adjudication methods, or MRMC studies is not available as this type of submission typically relies on a comparison to an already approved device.

    Here's an analysis based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a "SUMMARY OF NON-CLINICAL DATA" (page 8, "Table 1: Non-Clinical Testing Analysis") which serves this purpose. Many tests were not re-tested for the Zenith Support Catheter because no material changes were made from the predicate device (Zenith 074) and manufacturing processes were identical or within the predicate's tolerance.

    TestAcceptance CriteriaReported Device Performance (Analysis/Outcome)
    BiocompatibilityNo re-testing for Zenith Support Catheter: "There are no material changes from the predicate device. Manufacturing process is identical to the predicate. Only the segment lengths were changed and pitch adjustments were within the predicate device's tolerance." This applies to all biocompatibility tests below.
    - Material Mediated PyrogenThe test article extracts must not cause a febrile reaction greater than 0.5°C in any individual subject.Not re-tested.
    - Cytotoxicity MEM ElutionThe cultures treated with the test article must not have a reactivity grade greater than 2.Not re-tested.
    - Hemolysis (ASTM Method, extract human blood)The hemolytic index above the negative control article must be less than 5%.Not re-tested.
    - Hemolysis (ASTM method, direct contact human blood)The hemolytic index above the negative control article must be less than 5%.Not re-tested.
    - Unactivated Partial Thromboplastin TimeThere must be no statistical decrease between the UPTT of plasma exposed to the test article and to the negative or untreated control.Not re-tested.
    - Complement ActivationThere must be no statistical increase between either the C3a or SC5b-9 concentrations in plasma exposed to the test article as compared to the negative and untreated controls.Not re-tested.
    - Dog ThromboresistanceThe test articles must receive a thrombus formation score less than or equal to that of the control.Not re-tested.
    - Maximization SensitizationThe test article must elicit a positive response in less than 10% of the test animals.Not re-tested.
    - Intracutaneous Toxicity/ReactivityThe test article extracts must not induce a significantly greater biological reaction than the control.Not re-tested.
    - Acute Systemic Toxicity TestThe test article extracts must not induce a significantly greater biological reaction than the control.Not re-tested.
    Animal Testing - Angiographic EvaluationNo units tested can cause vessel injury.Not re-tested. "The tips, materials, and transitions are identical to the predicate, and the initial evaluation for the predicate included the use of the device without the Scout Introducer."
    Chemical CompatibilityChemicals have no negative effect on the catheter or accessories.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    Radiographic DetectabilityProduct shall be visible under fluoro imaging.Not re-tested. "The marker bands are identical to the predicate."
    Visual InspectionCatheter shall appear free from damage, including a rounded tip and smooth transition points.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    Working LengthTest samples should be within existing working length specification.Not re-tested. "The segment lengths do not impact the cut to length step in addition the working length test compares the output to the target length."
    Hub compatibilityHub shall meet existing Luer specifications.Not re-tested. "The hub is identical to the predicate device."
    Outside DiameterTest samples should be within existing outside diameter specifications.Not re-tested. "OD is unaffected by the working length change."
    Leak – AirTest samples should be within existing Air - Leak specifications.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    Leak – LiquidTest samples should be within existing Leak - Liquid specifications.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    ParticulatesTest samples should be within existing Particulate specifications.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    Simulated Use - BenchTest samples must meet predetermined user needs.Re-tested Yes. "Zenith Support Catheter test samples met the acceptance criteria for Simulated Use-Bench to demonstrate that it is substantially equivalent to the predicate device."
    Catheter BurstTest sample burst pressures must meet or exceed existing minimum burst pressure specification.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    TensileTest sample ultimate tensile strength must meet or exceed existing tensile strength specifications.Not re-tested. "The bonds and joints are the same as the predicate Zenith 074 device. Any pitch adjustments are within the tolerance of the predicate device."
    CorrosionTest samples shall exhibit no evidence of corrosion.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    Packaging - Dye LeakTest sample shall not exhibit any visual leaks or channels.Not re-tested. "The packaging is identical to the cleared reference SDA device." (Note: The reference here is to an "SDA device," not the Zenith 074 predicate, indicating identical packaging to another cleared device.)
    Packaging - PeelTest sample tensile strength must meet or exceed existing tensile strength specifications.Not re-tested. "The packaging is identical to the cleared reference SDA device." (Again, referring to SDA device for packaging.)
    Kink ResistanceTest sample kink resistance must meet or exceed existing Kink Resistance specifications.Not re-tested. "All transitions are the same, only the location of the transition changes, in addition, the pitch adjustments are within the Zenith 074 tolerance."
    TorqueTest sample torque results must meet or exceed existing torque specifications.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    PTFE Liner InspectionTest sample liner adhesion must meet or exceed existing PTFE Liner inspection specifications.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    Hydrophilic Coating IntegrityTest sample results must meet or exceed existing Hydrophilic Coating Integrity specifications.Not re-tested. "There are no material changes from the predicate device, and pitch adjustments are within the tolerance of the predicate device."
    Labeling LegibilityTest samples shall demonstrate text legibility.Not re-tested. "The label is almost identical to the predicate device, except that the Scout Introducer is removed from the labeling."
    BarcodeTest samples shall demonstrate readily readable barcodes.Not re-tested. "Barcodes are similar to the predicate."
    SterilizationSterilization load shall pose an equal or lesser challenge to sterilize than existing sterile product loads.Not re-tested. "The Zenith Support Catheter was adopted into the Zenith Family Sterilization Cycle."
    Shelf LifeAged test samples must meet or exceed existing specifications.Not re-tested. "There are no material changes from the predicate device, therefore shelf life testing conducted for the Zenith 074 still applies."

    2. Sample Size Used for the Test Set and the Data Provenance

    Due to the nature of a 510(k) summary relying on substantial equivalence, specific sample sizes for re-tested parameters (like Simulated Use - Bench) are not explicitly stated. For most tests, the data provenance is essentially inherited from the predicate device (Zenith 074) and another reference device (InNeuroCo SDA) because the new device's materials and manufacturing are identical or within tolerance. The document indicates these are non-clinical studies (benchtop and animal models for predicate devices). The geographical origin of the data is not specified beyond the manufacturer being based in Florida, USA. All studies are retrospective in relation to the current device's testing, as they refer to data from predicate devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided. For non-clinical tests, "ground truth" is typically established by laboratory procedures and measurements against defined standards (e.g., ISO, ASTM). For the "Simulated Use - Bench" test, it states it was conducted "by a physician," implying at least one expert, but qualifications and number are not detailed.

    4. Adjudication Method for the Test Set

    Not applicable/not provided. This type of detail (e.g., 2+1, 3+1 consensus) is typical for clinical studies, especially those involving expert interpretation of medical images, which is not the focus of this non-clinical submission.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (catheter), not an AI/Software as a Medical Device (SaMD) product. Therefore, MRMC studies and AI-assisted performance metrics are not relevant or performed for this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a physical catheter, not an algorithm.

    7. The Type of Ground Truth Used

    For the majority of the non-clinical tests not re-performed, the "ground truth" (or accepted performance) is based on:

    • Established performance of the predicate device (Zenith 074): This implies that the predicate device previously met these standards.
    • International Standards (ISO 10993, ISO 10555-1, ISO 594-1, ISO 594-2, ISO 11135) and ASTM standards (ASTM F1929-12, ASTM F88-09, ASTM F1980): These standards define methodologies and acceptable ranges for material properties, biocompatibility, and device functionality.
    • USP 788: For particulates.

    For the re-tested "Simulated Use - Bench" test, the ground truth was "predetermined user needs," likely established through design validation or clinical input.

    8. The Sample Size for the Training Set

    Not applicable. The Zenith Support Catheter is a physical medical device. The concept of "training set" is relevant for AI/machine learning models, which this device is not.

    9. How the Ground Truth for the Training Set was Established

    Not applicable as there is no "training set" for this physical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1