Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K210996

    Validate with FDA (Live)

    Date Cleared
    2021-04-23

    (21 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ZOOM Reperfusion Catheters, with the ZOOM Aspiration Tubing and ZOOM Aspiration Pump (or equivalent vacuum pump), are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    The ZOOM Aspiration Tubing is intended to connect the ZOOM Reperfusion Catheter of the ZOOM Aspiration Pump and to allow the user to control the fluid flow.

    Device Description

    The ZOOM™ Reperfusion Catheter is a single lumen, braid and coil reinforced, variable stiffness catheter that facilitates removal of thrombus/clot from the neurovasculature when connected to a vacuum source, such as the ZOOM Aspiration Pump, using the ZOOM Aspiration Tubing.

    The ZOOM Reperfusion Catheter is offered in various working lengths and nominal inner diameters (ID) and outer diameters (OD) as shown in Table 1.

    The ZOOM Reperfusion Catheter is comprised of a hollow cylindrical tube which is bonded to a standard luer fitting. The wall of the tube is constructed using a combination of metal coils/braids and medical grade polymers.

    The distal section of the ZOOM Reperfusion Catheter has a hydrophilic coating to enhance tracking through the vasculature. The beveled distal tip allows for atraumatic tracking past vessel branches during insertion. A radiopaque marker provides the user with visual confirmation of the distal tip location under fluoroscopy.

    The ZOOM Reperfusion Catheter is packaged with an accessory Rotating Hemostasis Valve (RHV). The RHV is designed to be attached to the proximal luer of the catheter and helps the user maintain hemostasis.

    The ZOOM Aspiration Tubing is offered in one model with the features indicated in Table 2.

    The ZOOM Aspiration Tubing is comprised of a hollow cylindrical tube which is bonded to a standard luer fitting that connects to the ZOOM Reperfusion Catheter and a slip fit connector that connects to the canister on the aspiration pump. The ZOOM Aspiration Tubing is made of common medical grade polymers.

    AI/ML Overview

    The provided document describes the acceptance criteria and a study conducted for the ZOOM Reperfusion Catheters and ZOOM Aspiration Tubing (K210996).

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    Test AttributeAcceptance Criteria (Specification)Reported Device Performance
    VisualFree of kinks, breaks, separation or particulate (greater than $0.25mm^2$). No exposed metal.The test results were reviewed and found to demonstrate that the differences between the subject ZOOM Reperfusion Catheters and predicate ZOOM Reperfusion Catheters do not significantly impact any performance parameters that would negatively affect the safety or effectiveness. (Implies performance met this and other specifications by not negatively impacting safety/effectiveness).
    Dimensional (Effective Length)All defined catheter dimensions are within the specified tolerances.The test results were reviewed and found to demonstrate that the differences between the subject ZOOM Reperfusion Catheters and predicate ZOOM Reperfusion Catheters do not significantly impact any performance parameters that would negatively affect the safety or effectiveness.
    Catheter Bond StrengthThe catheter shall have sufficient bond strengths to remain intact throughout a procedure.The test results were reviewed and found to demonstrate that the differences between the subject ZOOM Reperfusion Catheters and predicate ZOOM Reperfusion Catheters do not significantly impact any performance parameters that would negatively affect the safety or effectiveness.
    Flowrate - Positive (Forward) PressureThe catheter lumen shall allow for a minimum flowrate comparable to competitive products.The test results were reviewed and found to demonstrate that the differences between the subject ZOOM Reperfusion Catheters and predicate ZOOM Reperfusion Catheters do not significantly impact any performance parameters that would negatively affect the safety or effectiveness.
    Flowrate – Vacuum PressureThe flowrate under a vacuum shall be similar to or greater than competitive devices.The test results were reviewed and found to demonstrate that the differences between the subject ZOOM Reperfusion Catheters and predicate ZOOM Reperfusion Catheters do not significantly impact any performance parameters that would negatively affect the safety or effectiveness.
    Freedom from Leakage – Positive PressureNo liquid leakage from the hub or catheter shaft at 46psi for 30 seconds.The test results were reviewed and found to demonstrate that the differences between the subject ZOOM Reperfusion Catheters and predicate ZOOM Reperfusion Catheters do not significantly impact any performance parameters that would negatively affect the safety or effectiveness.
    Freedom from Leakage – Negative PressureNo air leakage into a 20cc syringe when vacuum pulled for 15 seconds.The test results were reviewed and found to demonstrate that the differences between the subject ZOOM Reperfusion Catheters and predicate ZOOM Reperfusion Catheters do not significantly impact any performance parameters that would negatively affect the safety or effectiveness.
    Dynamic Burst PressureCatheter does not burst under pressures that could be seen when performing contrast injections with a standard 10cc syringe.The test results were reviewed and found to demonstrate that the differences between the subject ZOOM Reperfusion Catheters and predicate ZOOM Reperfusion Catheters do not significantly impact any performance parameters that would negatively affect the safety or effectiveness.
    Tip FlexibilityThe flexibility of the catheter tip shall be comparable to competitive products and allow for easily tracking the device to the desired target anatomy.The test results were reviewed and found to demonstrate that the differences between the subject ZOOM Reperfusion Catheters and predicate ZOOM Reperfusion Catheters do not significantly impact any performance parameters that would negatively affect the safety or effectiveness.
    Corrosion ResistanceNo visible corrosion present on devices after saline immersion followed by 30 minutes in boiling water followed by 48 hours in 37°C water bath.The test results were reviewed and found to demonstrate that the differences between the subject ZOOM Reperfusion Catheters and predicate ZOOM Reperfusion Catheters do not significantly impact any performance parameters that would negatively affect the safety or effectiveness.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily describes bench testing to evaluate differences between the subject and predicate devices. No information about sample sizes for a clinical test set is provided in this submission summary, nor details about data provenance (country of origin, retrospective/prospective). This type of submission (510(k) Special) largely relies on demonstrating substantial equivalence through engineering and performance testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. The study described is bench testing, not a clinical study involving experts to establish ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This was bench testing, not a clinical study requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. No MRMC study or AI component is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This submission is for medical devices (catheters and tubing), not an algorithm or AI.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the bench testing, the "ground truth" was the predefined performance specifications and test methods based primarily on catheter performance standard ISO 10555-1. The devices were tested against these engineering and functional standards.

    8. The sample size for the training set

    Not applicable. This is a medical device submission based on bench testing of physical products, not an AI or machine learning model that would require a training set.

    9. How the ground truth for the training set was established

    Not applicable. As above, this is not an AI/ML submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202182

    Validate with FDA (Live)

    Date Cleared
    2020-09-02

    (29 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ZOOM Reperfusion Catheters, with the ZOOM Aspiration Tubing and ZOOM Aspiration Pump (or equivalent vacuum pump), are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

    The ZOOM Aspiration Tubing is intended to connect the ZOOM Reperfusion Catheter of the ZOOM Aspiration Pump and to allow the user to control the fluid flow.

    Device Description

    The ZOOM Reperfusion Catheter is a single lumen, braid and coil reinforced, variable stiffness catheter that facilitates removal of thrombus/clot from the neurovasculature when connected to a vacuum source, such as the ZOOM Aspiration Pump, using the ZOOM Aspiration Tubing.

    The ZOOM Reperfusion Catheter is offered in various working lengths and nominal inner diameters (ID) and outer diameters (OD) as shown in Table 1 below.

    The ZOOM Reperfusion Catheter is comprised of a hollow cylindrical tube which is bonded to a standard luer fitting. The wall of the tube is constructed using a combination of metal coils/braids and medical grade polymers.

    The distal section of the ZOOM Reperfusion Catheter has a hydrophilic coating to enhance tracking through the vasculature. The beveled distal tip allows for atraumatic tracking past vessel branches during insertion. A radiopaque marker provides the user with visual confirmation of the distal tip location under fluoroscopy.

    The ZOOM Reperfusion Catheter is packaged with an accessory Rotating Hemostasis Valve (RHV). The RHV is designed to be attached to the proximal luer of the catheter and helps the user maintain hemostasis.

    The ZOOM Aspiration Tubing is offered in one model with the features indicated in Table 2.

    The ZOOM Reperfusion Tubing is comprised of a hollow cylindrical tube which is bonded to a standard luer fitting that connects to the ZOOM Reperfusion Catheter and a slip fit connector that connects to the canister on the aspiration pump. The ZOOM Aspiration tubing is made of common medical grade polymers.

    AI/ML Overview

    The provided text describes the performance data for the ZOOM Reperfusion Catheters and ZOOM Aspiration Tubing (Gen 2), comparing it to a predicate device (Gen 1). However, it does not include information about AI/algorithm-specific acceptance criteria or studies. The documentation focuses on demonstrating substantial equivalence to a previous version of the same medical device through bench and lab testing, biocompatibility, sterilization, and shelf-life testing.

    Therefore, many of the requested sections related to AI/algorithm performance (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC studies, standalone performance, training set details) are not applicable to this document.

    Here's the information that can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally qualitative "Pass" or "Comparable to predicate/competitive products" based on the described tests. The reported performance is summarized as "Pass" for all validated attributes, indicating that the device meets its internal specifications and is comparable to the predicate.

    ZOOM Reperfusion Catheter

    Test AttributeSpecificationReported Device Performance
    Delivery, Compatibility, & Retraction (Trackability)The catheter shall be able to be delivered, deployed, and retracted per the IFU within a simulated neurological model without incurring any damage to the catheter.Pass
    Flexibility and Kink ResistanceThere shall be no kinking of shaft (permanent deformation) after simulated use.Pass
    Compatibility with other devices (external)The catheters shall be able to be delivered through the minimum introducer sheath or guide catheter size indicated in the product labeling.Pass
    Guidewire compatibilityThe catheters shall be able to be delivered over the guidewire size indicated in the product labeling.Pass
    Interventional device compatibility (internal)The catheters shall be able to accommodate other interventional devices (e.g., support catheter, diagnostic catheter) up to the maximum size indicated in the product labeling.Pass
    Luer compatibilityDevices and accessories shall be compatible with standard syringe luer fittings per ISO 80369-7.Pass
    Accessory compatibilityDevices shall be compatible with the accessory RHV.Pass
    VisualFree of kinks, breaks, separation or particulate (greater than 0.25mm^2). No exposed metal.Pass
    DimensionalAll defined catheter dimensions are within the specified tolerances.Pass
    Catheter Bond StrengthThe catheter shall have sufficient bond strengths to remain intact throughout a procedure.Pass
    Flowrate - positive (forward) pressureThe catheter lumen shall allow for a minimum flowrate comparable to competitive products.Pass (Comparable)
    Flowrate – vacuum pressureThe flowrate under a vacuum shall be similar to or greater than competitive devices.Pass (Similar/Greater)
    Freedom from Leakage – positive pressureNo liquid leakage from the hub or catheter shaft at 46psi for 30 secondsPass
    Freedom from Leakage – negative pressureNo air leakage into a 20cc syringe when vacuum pulled for 15 seconds.Pass
    Burst PressureCatheter does not burst under pressures that could be seen when performing contrast injections with a standard 10cc syringe.Pass
    Catheter Torque StrengthNo separation of any portion of the catheter when rotated at least two (2) full rotations (720 degrees).Pass
    Lumen IntegrityThe catheter lumen shall not collapse under vacuum after multiple passes.Pass
    Kink ResistanceThere shall be no kinking of the catheter shaft (permanent deformation) at anatomically relevant bend radii.Pass
    FlexibilityThe flexibility of the catheter tip shall be comparable to competitive products and allow for easily tracking the device to the desired target anatomy.Pass (Comparable)
    Coating - ParticulateThe amount of particulate matter that comes off the hydrophilic-coated shaft during simulated use testing shall be determined and compared to competitive products and techniques.Pass (Comparable)

    ZOOM Aspiration Tubing

    Test AttributeSpecificationReported Device Performance
    Vacuum Force at Catheter TipThe vacuum force delivered by the aspiration tubing to the tip of the catheter should be comparable to the vacuum force delivered by the predicate aspiration tubing.Pass (Comparable)
    Connector CompatibilityThe aspiration tubing connectors shall securely connect to the pump canister lid and standard luer fittings.Pass
    Lumen Collapse TestThe tubing lumen shall not collapse under vacuum.Pass
    Flow Control FunctionalityThe flow control mechanism shall allow users to start and stop flow multiple times when the connected pump is running at maximum vacuum.Pass
    Freedom From LeakageThe vacuum pressure delivered at the tip of the aspiration tubing shall be consistent with the pressure generated by the pump.Pass
    Tensile StrengthThe bonds between the tubing and connectors shall be sufficiently strong to ensure the tubing remains intact during use.Pass

    Biocompatibility Testing (ZOOM Reperfusion Catheter)

    TestTest MethodAcceptance CriteriaResults
    Cytotoxicity: ISO MEM ElutionISO 10993-5Sample extracts must yield cell lysis grade 2 or lower.Pass, Non-cytotoxic
    Cytotoxicity: ISO MTT AssayISO 10993-5The percentage of cells exhibiting lysis should be similar for all test devices. No Significant DifferencesPass, No Significant Differences
    Hemocompatibility: Hemolysis (Extract Method)ASTM F 756 ISO 10993-4Sample extracts must be non-hemolytic (≤ 2% hemolytic index).Pass, Non-hemolytic

    Sterilization Validation (ZOOM Reperfusion Catheter & ZOOM Aspiration Tubing)

    Requirement/Acceptance CriteriaResultsSummary
    Positive BI controls must be positive in fractional, half and full cycles.PASSAll positive controls in each cycle read positive.
    Fractional cycle requirements: 1) IPCDs and EPCDs must be more difficult to sterilize than the natural product.PASSIPCDs and EPCDs showed more growth than natural product.
    Fractional cycle requirements: 2) Product Sterility testing must show no growth in all samples.PASSThe devices did not show any growth when tested per the methods in ISO 11737-2.
    Fractional cycle requirements: 3) Bacteriostasis/Fungistasis testing must show that the product is not inhibitory for growth.PASSBacteriostasis/Fungistasis Testing showed product is not inhibitory to growth when tested per the methods in ISO 11737-2.
    Half cycle requirements: 1) IPCDs and EPCDs should show no growth for all samples.PASSIPCDs and EPCDs showed no growth.
    Full cycle requirements: 1) IPCDs and EPCDs must show no growth for all samples.PASSIPCDs and EPCDs showed no growth.
    Full cycle requirements: 2) The results for Ethylene Oxide (EO) and Ethylene Chlorohydrin (ECH) shall meet the requirements of ISO 10993-7.PASSSamples passed EO residual testing.

    Packaging Validation (ZOOM Reperfusion Catheter & ZOOM Aspiration Tubing)

    TestTest MethodT0 Results (Pass/Fail)Shelf Life Results (Pass/Fail)
    Packaging Visual InspectionASTM F1886 / Imperative Care InternalPassPass
    Pouch Integrity Test - Gross Leak DetectionASTM F2096PassPass
    Pouch Seal Strength – Peel StrengthASTM F88PassPass
    Label IntegrityImperative Care InternalPassPass

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document states that "The largest and smallest diameter catheters were tested for both the subject and predicate devices." Exact numerical sample sizes for each test are not specified, but the tests were performed on representative samples of the devices.
    • Data Provenance: The studies were bench and lab testing. This means the tests were conducted in a laboratory setting on the physical devices themselves, not on patient data. Therefore, "country of origin of the data" and "retrospective or prospective" are not applicable in the clinical data sense. The testing was conducted by Imperative Care Inc.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The reported studies are bench and lab testing of physical device performance, biocompatibility, sterilization, and packaging. There is no "ground truth" in the diagnostic sense or involvement of medical experts for interpreting diagnostic results. The acceptance criteria are based on engineering specifications and recognized consensus standards (e.g., ISO, ASTM).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as this is related to clinical interpretation or AI model output validation, which is not covered in this document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This document describes the performance of a physical medical device (catheter and aspiration tubing), not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No. This document does not describe an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The "ground truth" for these engineering and biological tests are the established specifications, standards (e.g., ISO, ASTM), and comparison to the predicate device's performance.

    8. The sample size for the training set

    Not applicable. There is no AI model or training set described in this document.

    9. How the ground truth for the training set was established

    Not applicable. There is no AI model or training set described in this document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1