Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K072923

    Validate with FDA (Live)

    Date Cleared
    2008-04-24

    (192 days)

    Product Code
    Regulation Number
    870.5310
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Defibrillator Function: The E Series products contain a dc defibrillator capable of delivering up to 200 joules of energy. It may be used in synchronized mode to perform synchronized cardioversion by using the R-wave of the patient's ECG as a timing reference. The unit uses paddles or disposable, pre-gelled, MFE Pads for defibrillation. The E Series products must be prescribed for use by a physician or medical advisor of an emergency response team. Intended Use - Manual Operation: Use of the E Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by these three conditions: Unconsciousness, Absence of breathing, and Absence of pulse. This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats. In manual mode, the E Series unit may also be used for synchronized cardioversion to terminate atrial fibrillation (AF) or ventricular tachycardias (VT) by using the R-wave of the patient's ECG as a timing reference. A qualified physician must decide when synchronized cardioversion is appropriate. The advisory function should be used to confirm ventricular fibrillation and wide complex ventricular tachycardia (greater than 150 beats per minute) in patients meeting the three conditions indicating lack of circulation (previously listed). Intended Use — Semiautomatic Operation (AED): The E Series AED unit is designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient. They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol. Use of the device in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation.

    External Pacemaker (Pacer Version Only): Intended Use - Pacemaker: This product may be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation. The purposes of pacing include: Resuscitation from standstill or bradycardia of any etiology, As a standby when standstill or bradycardia might be expected, Suppression of tachycardia. Pediatric Pacing: Pacing can be performed on pediatric patients weighing 33lbs / 15kg or less using special ZOLL. pediatric MFE Pads.

    Monitor: Intended Use - Multi-parameter Monitoring: This product may be used for monitoring various patient vital signs, including: electrocardiogram (ECG), Pulse Oximetry (SpO2), End Tidal CO2, 12-Lead ECG, and Non-Invasive Blood Pressure (NIBP). ECG monitoring is performed by connecting the patient to the 3 or 5 lead patient cable, MFE Pads, or through the paddles. SpO2 monitoring is indicated for detecting arterial oxygen saturation of blood and pulse rate for adult, pediatric and neonatal patients who are well or poorly perfusing, during both no motion and patient motion conditions. EtCO2 monitoring is indicated for the continuous measurement of end tidal carbon dioxide (EtCO2) and respiration rate for adult, pediatric and neonatal patients. 12 Lead ECG analysis is indicated for the diagnosis and treatment of adult and pediatric patients with acute myocardial infarction or other cardiac arrhythmias. NIBP monitoring is indicated for the measurement of arterial blood pressure for resting adult, pediatric, and neonatal patients.

    Intended Use -- CPR Monitoring: The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 1.5 to 2 inches (3.8 to 5.0 cm) for adult patients.

    Device Description

    The ZOLL E Series External Defibrillator is indicated for the defibrillation, Noninyasive Transcutaneous Pacing and multi-parameter monitoring of patient vital signs, including: ECG Monitoring, Pulse Oximetry, end tidal CO2, 12-Lead ECG monitoring, non-invasive blood pressure measurement, CPR performance and data printing and recording for resting patients in critical care and transport. The device is light weight and easy to carry for transport. The device is capable of providing Basic Life-Saving (BLS) personnel the option of analyzing a patient's ECG signal via the Advisory feature on the device. The Advisory Algorithm will determine if the acquired heart rhythm is shockable or non-shockable and will prompt the end-user to provide therapy in the event of a shock advised determination. The user will be prompted to perform Cardiopulmonary Resuscitation in the event that the Analyze feature determines that CPR compressions are more beneficial to the victim.

    AI/ML Overview

    The provided document does not contain explicit acceptance criteria tables or detailed study results for the Advisory Algorithm within the ZOLL E Series Defibrillator. While it states that "Performance and safety testing of the ZOLL E Series Defibrillator demonstrates that its features, functions and incorporated interpretive algorithm are that of the indicated commercially distributed predicate devices with regard to performance, safety and effectiveness," it relies on substantial equivalence to the Philips Medical Systems HeartStart FR2+ AED (K051632) rather than presenting a standalone study with specific metrics for the Advisory Algorithm's performance.

    Therefore, many of the requested details about acceptance criteria and study particulars cannot be directly extracted from the provided text.

    However, based on the context of an Automated External Defibrillator (AED) and its advisory function, we can infer common performance metrics and the type of acceptance criteria that would typically be expected, even if not explicitly stated with numerical targets.

    Inferred Acceptance Criteria and Reported Device Performance (Based on AED Standards and Substantial Equivalence):

    Given that the device's Advisory Algorithm determines if a rhythm is shockable or non-shockable, the key performance metrics would revolve around its accuracy in this determination.

    Acceptance Criterion (Inferred)Reported Device Performance (Inferred from Substantial Equivalence and General Statements)
    Sensitivity (Shockable Rhythm Detection)Ability to correctly identify shockable rhythms (e.g., ventricular fibrillation, rapid ventricular tachycardia) and advise a shock."Performance and safety testing... demonstrates that its features, functions and incorporated interpretive algorithm are that of the indicated commercially distributed predicate devices with regard to performance, safety and effectiveness." (Implies meeting or comparing favorably to predicate's known sensitivity)
    Specificity (Non-Shockable Rhythm Discrimination)Ability to correctly identify non-shockable rhythms and advise no shock."Performance and safety testing... demonstrates that its features, functions and incorporated interpretive algorithm are that of the indicated commercially distributed predicate devices with regard to performance, safety and effectiveness." (Implies meeting or comparing favorably to predicate's known specificity)
    Time to Analysis/Shock Advisory (If applicable)Not explicitly stated.
    CPR prompt accuracy/timing (for advisory on CPR)Not explicitly stated for the advisory algorithm's accuracy, but the CPR monitoring function prompts based on AHA/ERC guidelines.

    Detailed Study Information (Based on Available Text):

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not specified.
      • Data Provenance: Not specified. Given the context of a 510(k) submission, it's likely that a dataset of previously recorded ECGs (retrospective) would have been used for algorithm testing, but this is not explicitly stated.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not specified. For AED rhythm analysis, ground truth is typically established by expert cardiologists or emergency physicians, but details are not provided.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not specified.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not specified. The document focuses on the AED's standalone advisory function, not an AI-assisted human reading scenario. The device prompts the end-user based on its analysis, but there's no mention of a study comparing human performance with and without this specific AI's assistance.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, implicitly. The "Advisory Algorithm will determine if the acquired heart rhythm is shockable or non-shockable and will prompt the end-user." This describes the algorithm's standalone function (without immediate human intervention in the analysis process itself, though a human acts on the advice). The "Performance Testing" section states this testing was done.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not explicitly stated. For AED rhythm analysis, the ground truth for ECG interpretation (e.g., whether a rhythm is Ventricular Fibrillation, asystole, etc.) is typically established by expert consensus of cardiologists/electrophysiologists analyzing the ECG waveforms.
    7. The sample size for the training set:

      • Not specified. The document describes the "Advisory Algorithm" but does not detail its development or training process.
    8. How the ground truth for the training set was established:

      • Not specified. Assuming a similar methodology to the test set, it would likely involve expert interpretation of ECGs, but this is not confirmed.
    Ask a Question

    Ask a specific question about this device

    K Number
    K042007

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-04-07

    (255 days)

    Product Code
    Regulation Number
    870.5310
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ZOLL E Series External Defibrillator is indicated for the defibrillation, Noninvasive Transcutaneous Pacing, and multi-parameter monitoring of patient vital signs, including: ECG Monitoring, Pulse Oximetry, End Tidal CO2, 12-Lead ECG Monitoring, Non-Invasive Blood Pressure measurement and data printing and recording for resting patients in critical care and transport. The ZOLL E Series intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability to monitor and assess the physiological characteristics of the indicated patients in a critical care environment.

    Defibrillation: Use of the E Series products in the manual mode for defibrillation is indicated for converting/terminating ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats. Use of the device in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by: Unconsciousness, Absence of breathing, Absence of pulse.

    External Pacemaker (Pacer Version Only): This product may be used for temporary external, demand or non-demand pacing, as an alternative to endocardial stimulation. External cardiac pacing is indicated for use on conscious or unconscious patients in asystole, profound bradycardia or any other conditions determined by a clinician to require external pacing.

    ECG Monitoring: The device is indicated for monitoring a patient's electrocardiogram (ECG), via the 3 or 5 lead patient cable, MFE Pads, or through the paddles, for the purposes of identifying and diagnosing cardiac rhythms and dysrhythmias and calculating heart rate.

    SpO2 Option (if equipped): The SpO2 Option with Masimo Set Technology is indicated for the continuous noninvasive monitoring of arterial oxygen saturation (SpO2) and pulse rate for adult, pediatric and neonatal patients, during both no motion and patient motion conditions, and for patients who are well or poorly perfusing in the hospital or pre-hospital environments.

    EtCO2 Option (if equipped): The EtCO2 Option with Respironics Novametrix Technology and Capnostat® is indicated for the continuous noninvasive monitoring of end tidal carbon dioxide (EtCO2) and respiration rate in adult, pediatric and neonatal patient's requiring ventilatory support, transport and anesthesia.

    12SL Analysis Option (if equipped): The 12SL™ Analysis Option using the GE/Marquette 12SL™ Algorithm is useful in the diagnosis of patients with acute myocardial infarction (AMI) and is useful in the diagnosis of patients with other transient cardiac arrhythmias. The interpretation and 12-Lead ECG Analysis is indicated for the recording and analysis of ECG signals acquired from adult and pediatric patients in the supine, resting position.

    NIBP Option (if equipped): The NIBP Option with SunTech Medical Systems, Inc. Technology is indicated for the noninvasive measurement of arterial blood pressure for resting patients in critical care and transport. The NIBP Option on M Series units is designed to measure blood pressure for adult, pediatric, and neonatal patients.

    Device Description

    The ZOLL E Series External Defibrillator is indicated for the defibrillation, Noninvasive Transcutaneous Pacing, and multi-parameter monitoring of patient vital signs, including: ECG Monitoring, Pulse Oximetry, End Tidal CO2, 12-Lead ECG Monitoring, Non-Invasive Blood Pressure measurement and data printing and recording for resting patients in critical care and transport. The ZOLL E Series intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability to monitor and assess the physiological characteristics of the indicated patients in a critical care environment.

    The E Series products contain a DC defibrillator capable of delivering up to 200 joules of energy. It may be used in synchronized mode to perform synchronized cardioversion by using the R-wave of the patient's ECG. The E Series uses either standard paddles or disposable, pre-gelled, MFE Pads for defibrillation.

    AI/ML Overview

    The provided document is a 510(k) summary for the ZOLL E Series defibrillator and does not contain information regarding objective performance acceptance criteria or a study designed to prove the device meets such criteria for AI/ML-driven features.

    The document discusses the substantial equivalence of the ZOLL E Series to predicate devices based on:

    • Technological Characteristics: The design characteristics are the same as indicated predicate devices, and the technology is very similar to the ZOLL M Series. ECG analysis algorithms are identical to those in the ZOLL AED Plus defibrillator.
    • Performance Testing: "Extensive performance testing ensures that the ZOLL E Series Defibrillator meets all of its functional requirements and performance specifications."
    • Safety Testing: "Safety testing assures the device complies with applicable sections of recognized industry and safety standards."

    However, it does not detail specific acceptance criteria for the mentioned "ECG Analysis Algorithms" or "12SL Analysis Option," nor does it describe a study with a test set, ground truth, or statistical endpoints typically associated with AI/ML device performance evaluation as per your request. The document describes a traditional medical device clearance process focused on equivalence to existing devices, functional requirements, and safety standards.

    Therefore, I cannot provide the requested information in the format specified.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1