    K Number
    K964219

    Validate with FDA (Live)

    Date Cleared
    1997-08-11

    (293 days)

    Product Code
    Regulation Number
    876.5250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    1. For Penile Attachment Only
    2. Non-indwelling Urinary Continence Aid
    3. Management of Urinary Incontinence
    4. Urine Collection Device
    Device Description

    The ZIP™ Condom Catheter has four basic elements: 1) the condom/sheath with integral urine collection compartment; 2) the Velcro-like band and support strap; 3) the Application ring; and 4) the specially designed male underpants brief.

    The condom/sheath with integral urine collection compartment forms a bulbous tube approximately 10 ½ inches long. The tube is divided into two sections, the condom/sheath and the urine collection compartment. Non-return flow valves separate the condom/sheath portion from the urine collection compartment. The distal end of the urine collection compartment is affixed with a twist-open twistclosed stop-cock drain valve. The urine collection capacity is approximately 450 ml. Typical human urine output thus yields a time related capacity of about 4 to 6 hours.

    A series of 4 sizes differing in sheath diameter and length will be offered. Specific name designators for the sizes are vet to be determined.

    Design: The ZIP™ Condom Catheter expands upon existing male urine continence devices by first approaching the device as a "total" or "system" concept rather than just a collection of items originally developed for other urine collection purposes.

    Only the essential elements of the ideal male urine continence system were considered. First, the system had to be easy to apply and use. Second, it should allow use of traditional male urination methods, i.e., the stand-up urinals found in public restrooms. Third, the system should be small enough to go unnoticed during the typical activities of daily living. Fourth, the system should have adequate capacity so that urine disposal intervals are not unreasonably short. Lastly, the system should be reusable and easily cleaned and sanitized.

    This is a natural latex containing product. The ZIP™ Condom Catheter is made from the exact same natural latex formulation as the Everyday - Self-Adhesive Urinary External Catheter made by Hollister. Inc. The use of pure latex to construct the ZIP™ Condom Catheter requires the inclusion of certain FDA required cautionary statements. Latex containing devices must legibly bear an appropriate allergic reaction statement on the device labeling. In conformance with this FDA requirement, Goulter Medical, Inc. attaches a label containing the following cautionary statement to the device. "This product is made from natural rubber latex which may cause allergic reactions in some individuals".

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the ZIP™ Condom Catheter. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through clinical trials with specific acceptance criteria and performance metrics. Therefore, the document does not contain the information requested in your prompt regarding acceptance criteria, reported device performance, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance.

    The submission aims to show that the new device is as safe and effective as existing ones by comparing its materials, intended use, design, and method of operation.

    Here's an analysis based on the available information:

    1. A table of acceptance criteria and the reported device performance

    • Not Applicable. The 510(k) process for this device relies on demonstrating substantial equivalence to predicate devices, not on meeting specific, quantifiable performance acceptance criteria from a clinical trial in the way a novel diagnostic would. The "performance" being evaluated is primarily the functional equivalence to existing devices.

    2. Sample sized used for the test set and the data provenance

    • Not Applicable. There is no mention of a formal "test set" or clinical study with patient data used to directly evaluate the ZIP™ Condom Catheter's performance. The comparison is against predicate devices based on their established characteristics.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. No experts were used to establish ground truth for a test set as no such study was conducted or detailed in this submission.

    4. Adjudication method for the test set

    • Not Applicable. No test set or adjudication method is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This device is a physical medical device (condom catheter), not an AI diagnostic or assistance system. Therefore, no MRMC study or AI-related effectiveness is relevant or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used

    • Not Applicable / Implied Equivalence. For a 510(k) for this type of device, the "ground truth" is that the predicate devices are considered safe and effective for their intended use. The new device demonstrates substantial equivalence to these established predicate devices.

    8. The sample size for the training set

    • Not Applicable. There is no mention of a "training set" as this is not a machine learning or AI device.

    9. How the ground truth for the training set was established

    • Not Applicable. There is no training set involved.

    Summary of Device Comparison (from Section 10. Feature Comparison Table):

    The submission provides a feature comparison table to demonstrate substantial equivalence to predicate devices. This table serves as the primary "proof" for the 510(k) process.

    FEATUREZIP™ Condom CatheterMentor Freedom Cath®Everyday External CatheterUri-Drain®Substantially Equivalent?
    Materials:Medical Grade LatexLatexIdentical Latex FormulationLatexYes
    Intended Use(s):Non-Indwelling Male Urinary Continence Aid Management of Male Urinary Incontinence Male Bladder Training Aid to Improve Urinary ContinenceSame(Not explicitly stated for predicate in table, but implied by "Same" for Mentor)(Not explicitly stated for predicate in table, but implied by "Same" for Mentor)Yes
    Design:Single Piece UnitCondom OnlyCondom OnlyCondom OnlyYes
    Method of Operation:One Way Flow to Collection BagSameSameSameYes
    Bag:Integral to Condom SheathRequires Secondary Leg BagRequires Secondary Leg BagRequires Secondary Leg BagYes
    Bag Capacity:Approx. 450 mlAbout 1 LiterAbout 1 LiterAbout 1 LiterYes
    Underpant:Male Brief W/ PouchNANANANo
    How Secured:Velcro Band & StrapAdhesive SheathSpiral Foam BandSpiral Foam BandYes
    Sizes:4433Yes
    Additional Items Needed to Use the Device:None - Sold CompleteLeg Bag Connectors Tubing Adhesive tube HolderLeg Bag Connectors Tubing Tube clampLeg Bag Connectors Tubing Adhesive tube HolderNo
    Manufacturer:Goulter MedicalMentorHollisterSherwood MedicalYes
    Product Code:78EXJ78EXJ78EXJ78EXJYes
    K - NumberK964219PreamendmentPreamendmentPreamendment

    The "Yes" in the "SE?" column (which stands for Substantially Equivalent) indicates that, for that specific feature, the ZIP™ Condom Catheter is considered substantially equivalent to the predicate devices. The two "No" entries highlight differences (Underpant, Additional Items Needed to Use the Device) which the FDA presumably deemed not significant enough to prevent a finding of substantial equivalence overall, especially since the ZIP™ device offers a more integrated solution (integral bag, complete kit) which could be seen as an improvement in convenience rather than a safety/effectiveness concern.