Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K121338

    Validate with FDA (Live)

    Device Name
    ZEOBI
    Date Cleared
    2012-07-27

    (85 days)

    Product Code
    Regulation Number
    890.5290
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Adjunctive use in the palliative treatment of post-operative pain and edema in superficial soft tissue.

    Device Description

    The Zeobi is a portable battery powered non-invasive therapy device which applies to the body electromagnetic energy at a radio frequency (RF) of 27.12 MHz for the treatment of medical conditions by means other than the generation of deep heat within body tissues, i.e., by athermal means. The Zeobi was designed to deposit mean radio frequency energy in tissue which is equivalent to that of the Torino II. The Zeobi delivers the RF signal to the tissue target via inductive coupling with an applicator coil. The device is portable and treatment can occur directly through dressings, clothing, casts, compression garments or supports.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Zeobi device, based on the provided text:

    Acceptance Criteria and Device Performance

    The core of the "acceptance criteria" in this context is the concept of "substantial equivalence" to a predicate device. For the Zeobi, the predicate device is the SofPulse Torino II (K070541). The study's purpose is to demonstrate that the Zeobi is substantially equivalent to the Torino II regarding its indications for use, technical specifications, operating performance features, and general design.

    Acceptance Criteria (from Predicate Device - Torino II)Reported Device Performance (Zeobi)
    Indications for Use:
    Adjunctive use in the palliative treatment of post-operative pain and edema in superficial soft tissueAdjunctive use in the palliative treatment of post-operative pain and edema in superficial soft tissue
    Technical Specifications:
    Carrier Frequency: 27.12 MHz27.12 MHz
    Burst Duration: 2 msec2 msec
    Burst Repetition: 2 Hz2 Hz
    Energy Density: $0.13 \pm 0.02 \mu$ Ws/cm³$0.13 \pm 0.02 \mu$ Ws/cm³
    Electrical Safety: Conforms with IEC 60601-1Conforms with IEC 60601-1
    Electromagnetic Safety: Conforms with IEC 60601-1-2Conforms with IEC 60601-1-2
    Power Supply: In-Circuit Battery Source: Primary Lithium Coin Cell Batteries (2)Detachable Battery Pack: Primary Lithium Coin Cell Batteries (2)
    User Display: LED Display with two green lightsMulti-Function LCD Display
    Treatment Modes: • Automatic mode: 6 treatments per for three days, 3 treatments per day for 3 days and 2 treatments per day until shut off or battery depletes • Manual mode: user activated• Automatic mode: 12 treatments every two hours until the unit is shut off or battery depletes • Manual mode: user activated
    Treatment Time: Maximum of 30 minutesMaximum of 15 minutes

    Study Details for Zeobi (K121338)

    1. Sample size used for the test set and the data provenance:

      • No clinical test set with human subjects was used. The study relied on bench testing of the device's technical specifications.
      • The document implies that the "test set" for performance evaluation consisted of the Zeobi device itself undergoing various physical and electrical measurements.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the document. The ground truth for the bench testing would be established by the specifications of the predicate device (Torino II) and relevant industry standards (IEC 60601-1, IEC 60601-1-2). The experts would be the engineers and technicians performing the tests, but their specific qualifications are not detailed.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • None specified, as no clinical test set requiring adjudication was used. The evaluation was based on direct measurement comparisons to established performance parameters.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC study was done, as this is a physical therapy device and not an imaging-based diagnostic tool involving "human readers" or AI assistance in that context.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • This is a physical therapy device; therefore, the concept of "standalone algorithm performance" as typically applied to AI/ML devices is not applicable. The device's performance is intrinsically linked to its physical output and application by a user for therapeutic purposes.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth for the technical specifications (carrier frequency, burst duration, repetition, energy density, safety standards) was based on bench testing against the specifications of the predicate device (SofPulse Torino II K070541) and adherence to recognized international standards (IEC 60601-1, IEC 60601-1-2).
      • The "Indications for Use" ground truth was established by being identical to the predicate device.
    7. The sample size for the training set:

      • This document describes a 510(k) submission for a medical device focusing on substantial equivalence through technical comparison, not an AI/ML device requiring a "training set." Therefore, a training set is not applicable.
    8. How the ground truth for the training set was established:

      • As there is no training set mentioned or implied for this device, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1